Accuracy of the HVAD Pump Flow Estimation Algorithm

Reyes, Carlos; Voskoboynikov, Neil; Chorpenning, Katherine; LaRose, Jeffrey A.; Brown, Michael C.; Nunez, Nathalie J.; Burkhoff, Daniel; Tamez, Daniel

doi: 10.1097/MAT.0000000000000295
Adult Circulatory Support

Controller algorithms are an important feature for assessment of ventricular assist device performance. Flow estimation is one algorithm implemented in the HeartWare continuous-flow ventricular assist device pump system. This parameter estimates flow passing through the pump and is calculated using speed, current, and hematocrit. In vitro and in vivo studies were conducted to assess the algorithm accuracy. During in vitro testing, three pumps were tested in four water–glycerol solutions at 37°C with viscosities equivalent to hematocrits of 20, 30, 40, and 50%. By using a linear regression model, a correlation coefficient of >0.94 was observed between measured and estimated flow for all conditions. In vivo studies (n = 9) were conducted in an ovine model where a reference flow probe was placed on the outflow graft and speed was adjusted from 1,800 to 4,000 revolutions per minute. During in vivo experiments, estimated pump flow (mean, minimum, and maximum) was compared with measured pump flow. The best-fit linear regression equation for the data is y = 0.96x + 0.54, r2 = 0.92. In addition, waveform fidelity was high (r2 > 0.96) in normal (i.e., nonsuction) cases where flow pulsatility was >2 L/min. The flow estimation algorithm demonstrated strong agreement with measured flow, both when analyzing average waveform magnitude and fidelity.

From HeartWare, Inc., Miami Lakes, Florida.

Submitted for consideration April 2015; accepted for publication in revised form September 2015.

Disclosure: All authors are employees of HeartWare, Inc.

Correspondence: Daniel Tamez, HeartWare, Inc., 14400 NW 60th Avenue, Building 7C, Miami Lakes, FL 33014. Email: dtamez@heartwareinc.com.

Ventricular assist device (VAD) flow is a key parameter for managing patients undergoing mechanical circulatory support.1 For this reason, some VADs incorporate sensors for direct measurement of pump flow.2 Limitations of implantable sensor technologies are that they increase overall system cost and complexity while reducing long-term reliability.3–5 In lieu of implantable sensors, other devices have opted to estimate pump flow by algorithms based on other parameters of pump function, particularly power consumption.6,7 Several studies were conducted to assess the accuracy of these algorithms, which concluded that the flow estimation algorithms are valuable as trending tools, but not for accurate measurement of absolute flow values.6,8,9 A study by Giridharan et al.10 analyzed flow estimation accuracy of the HeartWare continuous-flow ventricular assist device (HVAD) pump (HeartWare, Inc., Miami Lakes, FL) in a static mock loop with glycerol–water mixtures at three viscosities, reporting flow estimation errors of ± 0.5 L/min at 2.7 cP, ± 0.2 L/min at 3.2 cP, and ± 0.5 L/min at 3.7 cP. It should be noted that no study has yet reported on the accuracy of flow estimation of the HVAD pump in in vivo pulsatile environment or analyzed accuracy of waveform average, minimum, maximum, and fidelity.

The HVAD pump is a continuous-flow, centrifugal VAD implanted in patients with end-stage heart failure. The device is implanted inside the thorax with the inflow cannula inserted into the ventricle and the outflow graft anastomosed to the arterial vasculature. The HVAD pump utilizes a wireless hybrid hydromagnetic impeller suspension system. The rotational speed of the impeller ranges from 1,800 to 4,000 revolutions per minute (RPM) and can support flows of up to 10 L/min. The percutaneous driveline connects the pump to an external controller that serves to power, control, monitor, and log pump operation.11 Further description of the HVAD pump system is provided by LaRose et al.12

Back to Top | Article Outline

HVAD Flow Estimation Algorithm Description

The HVAD flow estimation algorithm has three inputs and two outputs. The input parameters consist of pump rotational speed, pump current, and the patient’s hematocrit. Clinicians are able to adjust two of these inputs, the rotational speed and hematocrit setting, through the system monitor. The output parameters consist of an instantaneous pump flow signal and average pump flow. The waveform is displayed in real time on the monitor along with the average value.

The relationship between electrical current drawn by the pump (I) and blood flow through the pump (Q) is the basis for the HVAD flow estimation algorithm. These I-Q curves characterize the pump performance over the physiological range of pump speeds, pump flows, and blood viscosities. The valid range of estimated flow with the HVAD controller is between −2.0 and +10.0 L/min. In addition, HVAD pump I-Q curves are characterized for normal (dual motor stator operation) and also for single stator operation (front motor or rear motor). The latter allows the estimation of pump flow even when one motor stator fails, such as when there is a damage to the pump driveline. I-Q curves are programmed into the controller software in the form of tables that are interpolated to create the real-time flow estimation waveforms as shown schematically in Figure 1.

Figure 1

Figure 1

Back to Top | Article Outline

Methods

In Vitro Testing

In vitro studies were performed in a simple, steady-state flow loop with a water–glycerol solution at 37°C. Four hematocrits: 20, 30, 40, and 50% were simulated through adjustments to fluid viscosity. The conversion between target hematocrit and viscosity was approximated by Guyton et al.13 A Brookfield viscometer (Model LVDV-II+P CP; Brookfield Engineering Laboratories, Middleborough, MA) was used to measure fluid viscosity. In these experiments, simulated hematocrits of 20, 30, 40, and 50% had measured viscosities of 1.9 ± 0.2, 2.3 ± 0.1, 3.4 ± 0.3, and 5.0 ± 0.1 cP, respectively.

The HVAD pump was connected in series to a turbine flow probe (Flow Technology, Tempe, AZ) whose accuracy is independent of fluid temperature and viscosity. Controller hematocrit settings were set to match the test conditions using the system monitor. Pump speed was increased from 1,800 to 4,000 RPM in increments of 200 RPM for each hematocrit. In addition, to provide different hemodynamic operating conditions, the pump outflow was tested in both an unrestricted and a semirestricted state. In the semirestricted state, a side clamp was applied 3.0 cm away from the pump outlet to reduce the standard 10 mm diameter by 20, 30, 40, and 50%. A pin gauge was used to confirm the restricted diameter. This outflow restriction was applied for a distance of 3.5 cm.14 Three trials were performed using a total of three (n = 3) HVAD pumps to assess repeatability. Flow estimation error was calculated by computing the absolute difference between estimated pump flow and measured pump flow at each test condition.

Back to Top | Article Outline

In Vivo Testing

In vivo studies were performed in an acute ovine model to compare HVAD estimated flow with flow measured directly by the flow probe. Testing was conducted in nine (n = 9) healthy Suffolk Cross sheep weighing 68.4 ± 3.1 kg. All animals received humane care in compliance with the “Principles of Laboratory Animal Care” formulated by the National Society for Medical Research and the “Guide for the Care and Use of Laboratory Animals” prepared by the Institute of Laboratory Animal Resources and published by the National Institutes of Health (NIH Publication No. 86-23, revised 1996). The pump was implanted without the use of cardiopulmonary bypass. The inflow cannula for the HVAD pump was inserted into the left ventricular apex, and the 10 mm Dacron outflow graft was anastomosed to the descending aorta in an end-to-side fashion. A transit-time ultrasonic flow probe (Transonic Systems, Ithaca, NY) was placed around the pump outflow graft. Before the experiment, this flow probe had been specifically calibrated by the manufacturer for this graft size and type. Once the pump was implanted, speed was initiated at 1,800 RPM. A blood sample was collected to determine hematocrit at the time of measurements, and the results were input to the HVAD system monitor. Pump speed was then increased from 1,800 to 4,000 RPM in increments of 200 RPM. Each speed was maintained for approximately 2 minutes to obtain a steady-state condition. The speed ramp was discontinued if significant episodes of pump suction were observed. Herein, we define the occurrence of suction as intermittent negative deflections in the measured flow waveform that exceed a 2.0 L/min decrease in magnitude.

Real-time data from the HVAD controller and flow probe were collected using a custom data acquisition (DAQ) system, which was developed on a LabVIEW platform (National Instruments, Austin, TX). The sampling rate for this DAQ system was set to 50 Hz to match controller output.

Data were analyzed using a custom software viewer developed on a LabVIEW platform (National Instruments Corporation, Austin, TX). Estimated pump flow was compared with measured pump flow by examining a 10 second window at each pump speed and comparing the mean, maximum, and minimum of the two signals. Flow estimation error was calculated by computing the absolute difference between the mean estimated pump flow and the mean measured pump flow at each test condition. The average and standard deviation of the absolute mean flow estimation error was then calculated. Waveform fidelity was also examined by calculating the correlation coefficient between the estimated flow and the measured flow. A phase shift was observed between the measured and the estimated flow signals, so this phase shift was adjusted to obtain a maximum correlation coefficient value.

Back to Top | Article Outline

Results

In Vitro Studies

Both in vitro and in vivo testing demonstrated strong agreement between measured pump flow and estimated pump flow using a linear regression model. Results from the flow loop experiments, in which flows ranged between 1.4 and 7.0 L/min, showed >0.94 correlation coefficient between the measured turbine flow and the estimated pump flow for all hematocrit settings tested (20–50%) as well as all outflow restrictions tested (0–50%; Figure 2). Average flow errors for each test condition were <1.0 liter per minute as summarized in Table 1.

Table 1

Table 1

Figure 2

Figure 2

Back to Top | Article Outline

In Vivo Studies

Hematocrit values ranged between 18% and 35% during the nine in vivo studies. Maximum pump speeds achieved averaged 3,354 ± 312 RPM at which point flow averaged 4.1 ± 1.1 L/min. Suction occurred in three of the nine trials (each time at speeds >3,200 RPM).

Signal fidelity between the estimated flow waveform and the measured flow waveform was analyzed to determine the similarity of waveform shapes. A phase shift of 0.02 ± 0.01 seconds was applied to account for differences in processing time between the two flow signals. Examples showing the concordance between simultaneous real-time HVAD flow estimated signals and those provided by high-resolution transit-time flow probe are shown in Figure 3. A typical normal waveform morphology with flow ranging between 2 and 5 L/min (with a pulsatility of ~3 L/min) is shown in Figure 3A. An example with very low pulsatility (as may occur with low ventricular preload and high pump speeds) is shown in Figure 3B. An example showing intermittent suction events is shown in Figure 3C. These curves are characterized by sharp, transient, negative deflections in flow.

Figure 3

Figure 3

To facilitate comparisons, flow waveforms were quantified by their maximum (peak), minimum (trough), and mean values. There was a linear correlation for each of these values between measured and estimated flow signals as shown in Figure 4. This plot includes data from all animals and all speeds investigated and thus included waveforms with characteristics spanning the spectrum summarized in Figure 3. Analysis confirmed a high correlation between direct and estimated flow with linear regression line that did not differ appreciably from the line of identity: y = 0.96x + 0.54, r2 = 0.92. Strong agreement was present for the minimum flow values, the maximum flow values, and the mean flows values. The average absolute flow estimation error for mean flow values was 0.5 ± 0.3 L/min.

Figure 4

Figure 4

Accordingly, there was an equally strong agreement between pulsatility (i.e., maximum − minimum flow) from the flow probe and from the flow estimator (Figure 5): y = 0.93x + 0.25 L/min, r2 = 0.88. In this figure, the suction data points are differentiated from the normal data points. It was observed that measured flow reported larger decreases in pump flow during suction events than what was reported by the flow estimation algorithm. This is summarized quantitatively in Figure 6, which shows the correlation coefficient as a function of flow pulsatility. High correlation coefficients (r2 > 0.96) were observed for normal waveforms with pulsatility >2 L/min. For waveforms with pulsatility <2 L/min, the waveform correlation coefficients were lower (0.51 < r2 < 0.96). Suction data points also showed comparatively lower correlation coefficient values (0.66 < r2 < 0.88). This was the result of filtering by the flow estimation algorithm, which blunts the rapid drop in flow. However, as seen in Figure 3C, the estimated flow waveforms for suction cases are still a good overall representation of measured waveform shape for clinician purposes.

Figure 5

Figure 5

Figure 6

Figure 6

Back to Top | Article Outline

Discussion

The measured and estimated pump flow signal were in strong agreement and provided equivalent values for peak, trough, and mean values during both the in vitro and the in vivo studies. For in vitro studies, the average of the flow estimation algorithm’s absolute error for any test condition was at most 0.5 ± 0.3 L/min. This was observed at a 50% hematocrit and restrictions of 20 and 40%. Similarly, in vivo studies showed that the average of the flow estimation algorithm’s absolute error was 0.5 ± 0.3 L/min for mean values. Although a phase shift was observed that required postprocessing to achieve signal alignment, it is important to note that this did not impact calculations of the waveform mean, minimum, and maximum.

Analysis of signal fidelity demonstrated that estimated pump flow is a good indicator of real-time waveform shape with high correlation coefficients (r2 > 0.96) observed for normal waveforms with pulsatility >2 L/min. However, some rapid, transient events such as suction or low pulsatility (<2 L/min) may be filtered, resulting in a lower correlation coefficient value. It may be possible to improve algorithm accuracy even further by addressing the signal’s higher frequency content.

This study had several limitations that should be discussed. As a result of the selected animal model, higher flow rates (>7 L/min) and hematocrit ranges as may be encountered in patients could not be tested in these set of experiments. The highest average pump flow achieved was 6.9 L/min with a waveform peak of 8.7 L/min. As a result of the implantation procedure, the highest hematocrit tested was 35%. Future experiments should test these higher physiologic limits, which are within the expected clinical range. Additional validation in patients would further bolster the conclusions.

Another limitation in this study was the accuracy of the reference flow probe, which reports an absolute accuracy of ±10%. Thus, the inherent error present with the measured flow signal should be taken into consideration. Also, when comparing the two signals, a phase shift was observed because of differing processing times by the flow probe/flowmeter console, controller hardware, and DAQ system. Further characterization of this signal latency may be useful for future experiments.

Back to Top | Article Outline

Conclusion

A noninvasive and accurate pump flow indicator is an important tool for VAD patient management. In these experiments, the HVAD pump system demonstrated excellent agreement between measured pump flow and estimated pump flow, both when analyzing average waveform magnitude (mean, minimum, and maximum) and waveform fidelity. These results lay the foundation for the development and testing of new algorithms that, in turn, rely on the accurate noninvasive estimation of pump flow signals. For example, HVAD pump algorithms to assess ventricular contractility and left ventricular end-diastolic pressures using features of the estimated real-time flow waveform signals have already been proposed and tested.15 Such algorithms offer the possibility to provide clinicians with additional information for guiding management.

Back to Top | Article Outline

References

1. Slaughter MS, Pagani FD, Rogers JG, et al. Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. J Heart Lung Transplant. 2010;29:S1–S39
2. Goldstein DL. Worldwide experience with the MicroMed DeBakey ventricular assist device as a bridge to trasplantation. Circulation. 2003;108:272–277
3. Ayre PJ, Lovell NH, Woodard JC. Non-invasive flow estimation in an implantable rotary blood pump: A study considering non-pulsatile and pulsatile flows. Physiol Meas. 2003;24:179–189
4. Benkowski R, Lynch B, Morello G, Morley D, Noon GP. Real time flow measurement for a continuous flow VAD. ASAIO J.;47:133–2001
5. Ogawa D, Yoshizawa M, Tanaka A, et al. Evaluation of flow rate estimation method for rotary blood pump with chronic animal experiment. Conf Proc IEEE Eng Med Biol Soc. 2005;7:7616–7619
6. Lund LH, Gabrielsen A, Tirén L, Hallberg A, El Karlsson K, Eriksson MJ. Derived and displayed power consumption, flow, and pulsatility over a range of HeartMate II left ventricular assist device settings. ASAIO J. 2012;58:183–190
7. Malagutti N, Karantonis DM, Cloherty SL, et al. Noninvasive average flow estimation for an implantable rotary blood pump: A new algorithm incorporating the role of blood viscosity. Artif Organs. 2007;31:45–52
8. Slaughter MS, Bartoli CR, Sobieski MA, et al. Intraoperative evaluation of the HeartMate II flow estimator. J Heart Lung Transplant. 2009;28:39–43
9. Hasin T, Huebner M, Li Z, et al. Association of HeartMate II left ventricular assist device flow estimate with thermodilution cardiac output. ASAIO J. 2014;60:513–518
10. Giridharan GA, Koenig SC, Soucy KG, et al. Left ventricular volume unloading with axial and centrifugal rotary blood pumps. ASAIO J. 2015;61:292–300
11. Chorpenning K, Brown MC, Voskoboynikov N, Reyes C, Dierlam AE, Tamez D. HeartWare controller logs a diagnostic tool and clinical management aid for the HVAD pump. ASAIO J. 2014;60:115–118
12. Larose JA, Tamez D, Ashenuga M, Reyes C. Design concepts and principle of operation of the HeartWare ventricular assist system. ASAIO J. 2010;56:285–289
13. Guyton AC. Textbook of Medical Physiology. 2006 Philadelphia Elsevier Saunders
14. Krabatsch T, Hennig E, Stepanenko A, et al. Evaluation of the HeartWare HVAD centrifugal pump for right ventricular assistance in an in vitro model. ASAIO J. 2011;57:183–187
15. Granegger M, Moscato F, Casas F, Wieselthaler G, Schima H. Development of a pump flow estimator for rotary blood pumps to enhance monitoring of ventricular function. Artif Organs. 2012;36:691–699
Keywords:

flow estimation; HVAD pump

Copyright © 2016 by the American Society for Artificial Internal Organs