Development of an Inlet Pressure Sensor for Control in a Left Ventricular Assist Device

Fritz, Bryan; Cysyk, Joshua; Newswanger, Ray; Weiss, William; Rosenberg, Gerson

doi: 10.1097/MAT.0b013e3181d2a56e
Adult Circulatory Support

A Tesla type continuous flow left ventricular assist device (VAD) has been designed by Penn State and Advanced Bionics, Inc. (ABI). When a continuous flow device is used, care must be taken to limit low pressures in the ventricle, which can produce an obstruction to the inlet cannula or trigger arrhythmias. Design of an inexpensive, semiconductor strain gauge inlet pressure sensor to detect suction has been completed. The research and design analysis included finite element modeling of the sensing region. Sensitivity, step-response, temperature dependence, and hysteresis tests have been performed on prototype units. All sensors were able to withstand the maximum expected strain of 82 μm/in at 500 mm Hg internal pressure. Average sensitivity was 0.52 ± 0.24 μV/mm Hg with 0.5 V excitation (n = 5 units). Step-response time for a 0- to 90-mm Hg step change averaged 22 msec. Hysteresis was measured by applying and holding 75 mm Hg internal pressure for 4 hours, followed by a zero pressure measurement, and ranged from −15 to 4.1 mm Hg (n = 3 units). Offset drift varied between 180 and −140 mm Hg over a 4-week period (n = 2 units). Span temperature sensitivity ranged from 18 to −21 μV/°C (n = 5 units). Gain temperature sensitivity ranged from −7.4 to 4.9 μV/°C (n = 5 units). With the inherent drift, it is currently not possible to use the transducer to measure actual pressures, but it can easily be used to measure pressure changes throughout the cardiac cycle. This signal can then be used in the control system to avoid ventricular suction events.

From the Division of Artificial Organs, Department of Surgery, College of Medicine, Pennsylvania State University, Hershey, Pennsylvania.

Submitted for consideration July 2, 2009; accepted for publication in revised form December 21, 2009.

Reprint Requests: Gerson Rosenberg, PhD, Division of Artificial Organs, The Pennsylvania State University, College of Medicine, 500 University Drive, H151, PO Box 850, Hershey, PA 17033. Email: grosenberg@psu.edu.

Heart failure currently occurs in >500,000 patients per year.1 Although a heart transplant is a viable option for some of these patients, there are currently >40,000 heart transplant eligible candidates and <2,200 hearts available.2 Today, a relatively common treatment includes implantation of a left ventricular assist device (LVAD) as either a bridge to transplant or as end destination therapy. According to the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS), 483 LVADs were implanted at 94 different centers over a 2-year span.3 This number is expected to increase as more devices are approved and more centers are incorporated into the database. Many of these implanted pumps are of a continuous flow scheme. By their design, axial, centrifugal, and mixed continuous flow, pumps can create suction at the pump inlet, and care must be taken to avoid negative pressures during decompression of the left ventricle, since it can produce an obstruction to the inlet cannula. Suction on the inlet to the device can potentially trigger malignant arrhythmias and even cause death.4 An inlet pressure control system with an integrated pressure sensor that can detect potential suction and signal the device controller to take corrective actions is required for these devices to operate at peak efficiencies. Other groups have used other means to estimate pressure including measuring displacement of the rotor5 and investigating pulsatility in the flow.4,6 One study used a similar technique as the one used here with good results.7

To facilitate the design of the control scheme, design of an inexpensive, semiconductor strain gauge inlet pressure sensor to detect suction in the inlet of a Tesla style continuous flow device has been completed. The Tesla style pump is unique when compared with other LVADs in that it is a shear flow pump that uses a series of rapidly rotating disc to impart momentum to the fluid.8–10 The transducer is incorporated into the lumen of the titanium inlet connector of the VAD.

Here, we describe in detail the design and evaluation of the sensor and how it may be used to monitor pressure changes during the cardiac cycle and control the LVAD. Analytical analysis and finite element modeling of the sensing region has been completed. Sensitivity, step-response, temperature dependence, and hysteresis tests have been performed on prototype units. Discussion regarding the end use of the signal will also be introduced.

Back to Top | Article Outline

Description

The sensor body is manufactured from 6Al-4V titanium. This material was chosen for its ease of machining, on our prior experience using this material, and on biological response of the body to the material, which has been well documented. Two diaphragm areas located 180 degrees apart, where the sensor gauges are mounted, are produced by wet grinding across the outside of the part, tangential to the inside lumen of the inlet cap to the pump. The diaphragm is flat only on the outside of the sensor body; a nonobstructed circular path remains through the entire inlet to the device, thereby not disturbing the flow field in any manner. A mandrel is placed inside the inlet cap of the device to give mechanical support to the thin region during the manufacturing process. Once grinding has been completed, a thin region in the shape of an oval and tangential to the axis of rotation and inlet of the device is produced. It is intended that this region be no more than 0.001 in thick in an area that lies along the central axis of the pump. Once fabrication of the inlet cap diaphragms is complete, the inside is polished to a mirror-like finish. This eliminates any need for a coating to be applied to alleviate potential biological depositions. A picture of the prototype inlet cap may be seen in Figure 1 with the oval indicating the general shape of the diaphragm region. The long side of the oval is along the axis of flow through the sensor. Two strain gauges are mounted on each side of the inlet of the device oriented 180 degrees apart, one each along and transverse to the central axis of the device. The four gauges on each sensor are cross wired in a Wheatstone bridge configuration to help compensate for bending in the inlet. It can be seen in the Figure that as the inlet of the device is bent, a gauge on one side of the inlet is in compression, whereas the other side will be in tension. When this occurs, one gauge will exhibit an increase in resistance and the corresponding gauge on the opposite side of the device will exhibit a decrease in resistance, thereby canceling one another in the output signal.

Back to Top | Article Outline

Materials and Methods

General Instrumentation

Instrumentation used with the pressure sensors remained fixed for all experiments, except as noted. Each bridge on the inlet pressure sensors was connected to a National Instruments Corporation (Model SC-2345; Austin, TX) data acquisition system (NI-DAQ). The NI-DAQ system had SCC-SG04 modules installed to acquire the output signal from the device. An external variable voltage source (Model 3630A DC; Hewlett-Packard, Palo Alto, CA) was set at 0.50 V to provide excitation to the bridge. Ambient temperature of the surrounding air was monitored using an Omega Chromega/Alomega Type-J thermocouple (Model 5TC-TT-K-30-36; Omega Engineering Incorporated, Stamford, CT) and connected to the NI-DAQ system using a SCC-TC01 thermocouple module. All modules contained a differential amplifier providing a gain of 100 with an error of ±0.8% full scale in the nominal temperature range (23°C ± 2°C). The thermocouple module also includes a dual-pole 2 Hz filter onboard. Unless otherwise noted, all experiments were conducted in the nominal temperature range. For all tests, data were sampled at 1,000 Hz for 1 second using a custom developed LabVIEW program. The lengths of time between data samplings varied for each experiment. Each of the sampling files was then averaged to compile a single data point.

Back to Top | Article Outline

Finite Element Analysis

Finite element analysis (FEA) was run on SolidWorks Cosmos Simulation Software (Dassault Systèmes SolidWorks, Concord, MA). Several different scenarios were run on the solid model of the inlet cap to the LVAD. During all FEA analysis, the mesh for the simulation was tightly controlled in the region of the diaphragm. A close-up view of the mesh control may be seen in Figure 2. For all analyses, the large round cross-sections in each Figure were fixed, and internal pressures of 10, 25, and 500 mm Hg were applied. The rigidly mounted setup is similar to the arrangement that will be seen in the final clinical device. The 10 and 25 mm Hg pressures were intended to simulate physiologic conditions expected to be seen before a suction event, and FEA analysis indicated that negative pressures inside the device yielded identical results. The 500-mm Hg pressure was intended for design purposes only. Fatigue life for the thin diaphragm region was considered using the FEA-generated stress values at the maximum design criteria. The resultant von Mises stress of 1,542 psi is well below the yield strength of the titanium material. As a result, at this stress level, the fatigue life of the diaphragm is assumed infinite.

Back to Top | Article Outline

Sensitivity Testing

All sensitivity data were acquired with five sensors connected in series by silicone tubing (Figure 3). An Argon Medical (Athen, TX) CDXpress disposable pressure transducer was calibrated using a certified digital pressure manometer (Sper Scientific, Scottsdale, AZ). An initial reading of each sensor output with zero pressure applied was acquired. After the initial reading, a small internal pressure (∼5 mm Hg) was applied using a stopcock and syringe setup as seen on the left side of in Figure 3. This process of adding 5-mm Hg increments of pressure to the system was repeated until the overall internal pressure reading was 75 ± 5 mm Hg. These tests were repeated at both room temperature (23°C ± 2°C) and physiological body temperature (37°C ± 1°C) for all five sensor prototypes.

Back to Top | Article Outline

Drift Testing

Two sensor units were placed under a plastic tub to maintain consistent atmospheric conditions, absence of large fluctuations in temperature because of air currents. The sensors were allowed to remain undisturbed for a period of 4 weeks. Temperature was monitored using the Omega Chromega/Alomega Type-J thermocouple, and data were taken for 1 second in half-hour intervals.

Back to Top | Article Outline

Hysteresis Testing

The experimental setup for hysteresis testing was similar to the one that was used during the sensitivity testing described earlier and shown in Figure 3. Three of the prototype sensors were connected in series and run through two types of hysteresis tests. The first test involved holding 75 mm Hg of internal pressure (relative to atmospheric pressure) for a prescribed period of time and then allowing the system to return to zero pressure. The length of time that the system remained pressurized varied for each test. The holding times tested were 1 minute, 15 minutes, 1 hour, and 4 hours. For all test conditions, an initial reading was taken with no pressure; the system was pressurized, and a reading was taken after 15–20 seconds. The system was allowed to remain pressurized for the prescribed time; another reading was captured, and the pressure was released. After another 15–20-second stabilization period, the final reading was taken with zero pressure in the system. For the two longer tests (1 and 4 hour), an intermediate reading was also taken at the half-way point. For the second hysteresis test, the internal pressure applied to the gauges was incrementally increased in 5–7 mm Hg steps until a pressure of 75 mm Hg was reached. Each reading was held for 15–20 seconds, allowed to stabilize, and data were then recorded. For each of the second type tests, at least 10 data points were taken on the way up to 75 mm Hg and during the return to zero pressure.

Back to Top | Article Outline

Step Testing

Testing of pressure gauges typically includes step testing, in which a pressure or load is suddenly applied to or removed from the gauge. A standard type of step test for pressure sensors is the pop test.11 The pressure sensors were connected in series with an Argon pressure transducer and a standard 60-ml Luer lock medical syringe (Figure 3) attached to the system via a 3-way stopcock. A negative pressure of −95 ± 5 mm Hg was pulled onto the system using the syringe. The syringe rubber plunger was “popped” free and removed from its housing allowing ambient pressure to rush into the system, creating the step input.

Back to Top | Article Outline

In Vitro Testing

Response of the sensor to detect potential suction events was also tested in vitro. On a Penn State mock circulatory loop, the Tesla type LVAD (Pleasanton, CA) was plumbed in series with a Thoratec pneumatic VAD (which simulated the native ventricle). The custom pressure sensor was placed between the two devices, as was a fitting to allow the connection of an Argon commercial pressure transducer. Figure 4 shows the loop setup for this test.

Back to Top | Article Outline

In Vivo Testing

A 5-hour in vivo study using the Tesla type VAD with an inlet pressure sensor was conducted on a 105-kg calf. To compensate for baseline drift, the pressure trace was detrended by subtracting a 1-hour moving average of the input signal. The detrended inlet pressure signal was input to a custom control algorithm to automatically vary the pump speed. Suction was detected as a large drop in inlet pressure (−dP/dt). When a suction event was recognized, the control algorithm instructed the pump to reduce the set speed of the device.

Back to Top | Article Outline

Results

Finite element analysis estimated strain for the 10 and 25 mm Hg internal pressure on the inlet cap was 1.6 and 3.5 μm/m, respectively. Estimated strain at the design maxima criterion of 500 mm Hg was 82 μm/m. The FEA-estimated pictorial representations of strain for the operating pressures may be seen in Figure 5, A and B, where the internal pressure was fixed at 10 mm Hg (Figure 5A) and 25 mm Hg (Figure 5B).

Drift tests revealed that offset (zero pressure) drift varied from −180 to 140 mm Hg over a 4-week period (n = 2 units). Average sensitivity for the prototype devices were 0.52 ± 0.24 μV/mm Hg with 0.5 V excitation (n = 5 units). A comparison of the sensitivity tests at two temperatures (23 ± 2°C and 37 ± 1°C) yielded the gain drift for the sensors of −8.1 to 4.2 ± 45.3 μV/°C (n = 5 units).

Pulsatile waveforms were also introduced to the sensors to measure span drift at multiple temperatures, room temperature (23 ± 2°C), and physiological body temperature (37 ± 1°C). The physiological temperature was introduced to the sensor through the use of a heated water bath and warmed fluid circulating through the devices. Comparing the two pulsatile waveforms, it was observed that maximum span drift ranged from 18 to −21 ± 19.7 μV/°C (n = 5 units).

Step-response time for a 0- to 90-mm Hg step change averaged 22 msec (n = 5 units). The resulting unfiltered output signal from the two sensors may be seen in Figure 6. The remaining three sensors were not shown for clarity; however, all remaining sensors exhibited similar response characteristics. As can be seen in Figure 6, the sensor responds almost instantaneously to the step input, similar to the Argon control transducer.

Hysteresis ranged from −15 to 4.1 mm Hg (n = 3 units). These values were measured from the 4-hour test, during which the three sensors exhibited their most hysteresis. Hysteresis was not statistically significant during the shorter duration tests.

During in vitro testing, suction events were easily identifiable using output of the custom sensor. In Figure 7, the top trace is the custom pressure sensor, while the bottom trace indicates the Argon control. A suction event was determined to be occurring through visual observation of the blood sac in the pneumatic Thoratec device (Figure 4). During a suction event, the blood sac would occlude the inlet to the Advanced Bionics, Inc. (ABI) device and a simultaneous zero flow reading would result. As can be observed, three suction events occur in the first 9 seconds of the experiment and both devices successfully register these events. After ∼10 seconds, the suction events are fully resolved. Both sensors then return to “normal” trace. The negative pressure spikes are the result of rapid flow deceleration during the occlusion of the inlet cannula. As can be seen in Figure 8, the inlet pressure sensor can also adequately detect suction on the inlet cannula during an in vivo experiment.

Back to Top | Article Outline

Discussion

Drift of the magnitude seen during the drift tests precludes the use of the sensor to measure true pressure at the VAD inlet (i.e., gauge pressure relative to intrathoracic pressure). Low drift versions of the strain gauges used in the sensors are available and need to be investigated. Also, the use of a different approach to sensor gauge bonding in which the gauge is molecularly bonded to the diaphragm is currently being explored. Larger values of excitation voltage for the sensor were experimented with; however, this caused self heating of the gauges and led to very erratic drift behavior.

Although the exact cause of the hysteresis was not determined, it was speculated that the problem derives from the use of an epoxy to hold the gauges to the diaphragm. Transient response of the sensor seems to adequately detect impending suction events.

Pressure sensor drift did not occur over time scales <1 hour, and therefore deviations in the pressure trace on a beat-to-beat basis were because of physiologic changes in pressure and not because of baseline drift. The output signal from the inlet pressure sensor will be used in active control of the device. As a suction event is detected, the control algorithm reduces pump speed until the suction event has resolved.

Back to Top | Article Outline

Conclusion

An inexpensive, custom pressure sensor has been developed to detect potential suction events in the inlet to a new type of VAD. Drift currently remains an issue, and alternative methods of gaging the system are being investigated. If drift in the system can be stabilized, a duplicate type of device may be placed on the outlet of the device to help in flow estimation. Although the goal to measure physiological blood pressure may not have been achieved, the output is still indicative of pressure inside the lumen to the inlet of the device. This output signal will be fed into the control algorithm to determine if a suction event is occurring and reduces pump speed as necessary. Although at the present time, the drift of the sensor limits pressure measurements, the signal is quite satisfactory as a control signal to reduce/eliminate ventricular suction events.

Back to Top | Article Outline

Acknowledgment

Supported by NIH Grant no R01HL81119, Development of an Innovatively Suspended TESLA Pump LVAD.

Back to Top | Article Outline

References

1. Cardiovascular disease statistics 2005. Available at: http://www.americanheart.org/presenter.jhtml?identifier=4478. Accessed June 1, 2009.
2. Trends in hospitalization for heart failure by age group, 1979–2004, United States 2006. Available at: http://www.cdc.gov/DHDSP/library/fs_heart_failure.htm. Accessed May 25, 2009.
3. Holman WL, Pae WE, Teutenberg JJ, et al: INTERMACS: interval analysis of registry data. J Am Coll Surg 208: 755–761, 2009.
4. Vollkron M, Voitl P, Ta J, et al: Suction events during left ventricular support and ventricular arrhythmias. J Heart Lung Transplant 26: 819–825, 2007.
5. Hetzer R, Weng Y, Potapov EV, et al: First experiences with a novel magnetically suspended axial flow left ventricular assist device. Eur J Cardiothorac Surg 25: 964–970, 2004.
6. Vollkron M, Schima H, Huber L, et al: Development of a reliable automatic speed control system for rotary blood pumps. J Heart Lung Transplant 24: 1878–1885, 2005.
7. Bullister E, Reich S, D'Entremont P, et al: A blood pressure sensor for long-term implantation. Artif Organs 25: 376–379, 2001.
8. Tesla N: Inventor, Turbine. US patent 1,061,206, 1913.
9. Izraelev V: Inventor. Blood pump having rotor with internal bore for fluid flow. US patent 5,938,412, 1997.
10. Izraelev V, Weiss WJ, Fritz B, et al: A passively suspended Tesla pump left ventricular assist device. ASAIO J 55: 556–561, 2009.
11. Nichols WW, O'Rourke MF. McDonald's blood flow in arteries: theoretic, experimental and clinical principles. London, UK: Hodder Arnold, 1998.