Tidal Volume Lower than 6 ml/kg Enhances Lung Protection: Role of Extracorporeal Carbon Dioxide Removal

Terragni, Pier Paolo M.D.*; Del Sorbo, Lorenzo M.D.*; Mascia, Luciana M.D., Ph.D.*; Urbino, Rosario M.D.*; Martin, Erica L. Ph.D.*; Birocco, Alberto M.D.†; Faggiano, Chiara M.D.†; Quintel, Michael M.D.‡; Gattinoni, Luciano M.D.§; Ranieri, V Marco M.D.‖

Free Access
Supplemental Author Material
Article Outline
Collapse Box

Author Information

Collapse Box

Abstract

Background: Tidal hyperinflation may occur in patients with acute respiratory distress syndrome who are ventilated with a tidal volume (VT) of 6 ml/kg of predicted body weight develop a plateau pressure (PPLAT) of 28 ≤ PPLAT ≤ 30 cm H2O. The authors verified whether VT lower than 6 ml/kg may enhance lung protection and that consequent respiratory acidosis may be managed by extracorporeal carbon dioxide removal.
Methods: PPLAT, lung morphology computed tomography, and pulmonary inflammatory cytokines (bronchoalveolar lavage) were assessed in 32 patients ventilated with a VT of 6 ml/kg. Data are provided as mean ± SD or median and interquartile (25th and 75th percentile) range. In patients with 28 ≤ PPLAT ≤ 30 cm H2O (n = 10), VT was reduced from 6.3 ± 0.2 to 4.2 ± 0.3 ml/kg, and PPLAT decreased from 29.1 ± 1.2 to 25.0 ± 1.2 cm H2O (P < 0.001); consequent respiratory acidosis (Paco2 from 48.4 ± 8.7 to 73.6 ± 11.1 mmHg and pH from 7.36 ± 0.03 to 7.20 ± 0.02; P < 0.001) was managed by extracorporeal carbon dioxide removal. Lung function, morphology, and pulmonary inflammatory cytokines were also assessed after 72 h.
Results: Extracorporeal assist normalized Paco2 (50.4 ± 8.2 mmHg) and pH (7.32 ± 0.03) and allowed use of VT lower than 6 ml/kg for 144 (84–168) h. The improvement of morphological markers of lung protection and the reduction of pulmonary cytokines concentration (P < 0.01) were observed after 72 h of ventilation with VT lower than 6 ml/kg. No patient-related complications were observed.
Conclusions: VT lower than 6 ml/Kg enhanced lung protection. Respiratory acidosis consequent to low VT ventilation was safely and efficiently managed by extracorporeal carbon dioxide removal.
LIMITATION of tidal volume (VT) to 6 ml/kg predicted body weight (PBW) and of end-inspiratory plateau pressure (PPLAT) to a maximum of 30 cm H2O represents the standard for mechanical ventilation of patients with acute respiratory distress syndrome (ARDS).1,2 However, recent studies found that (1) tidal hyperinflation may occur in some patients despite limiting VT to 6 ml/kg and PPLAT to 30 cm H2O3, (2) ARDS patients may benefit from VT reduction even if they already have PPLAT < 30 cm H2O.4
Extracorporeal assist separating carbon dioxide removal from oxygen uptake has been proposed by Gattinoni et al.5 With this technique, carbon dioxide is removed by a pump-driven veno-venous bypass, and oxygenation is accomplished by high levels of positive end-expiratory pressure (PEEP) and three to five sighs every minute.5 Although effective,5 negative results of a clinical trial,6 the extensive amount of required resources,7 and the high incidence of side effects8 restricted the use of extracorporeal carbon dioxide removal as “rescue” therapy for the most severe case of ARDS.9
To reduce complexity, expenses, and side effects of extracorporeal lung assistance, Pesenti et al. proposed the concept of removing “only a portion of carbon dioxide production” to allow less traumatic ventilator settings.10 The current study set up to examine the hypothesis that a modified renal replacement circuit incorporating a neonatal membrane lung coupled in series with a hemofilter may safely remove the amount of carbon dioxide sufficient to buffer the respiratory acidosis associated to VT lower than 6 ml/kg and allow more protective ventilatory settings.
Back to Top | Article Outline

Materials and Methods

Patient Selection
Patients admitted from July 2006 to September 2007 in the intensive care units of the S. Giovanni Battista-Molinette hospital (University of Turin, Turin, Italy) were enrolled. Inclusion criteria were age of at least 18 yr and diagnosis of ARDS.11 Exclusion criteria were more than 3 days since they met ARDS criteria, pulmonary artery occlusion pressure greater than 18 mmHg, history of ventricular fibrillation, tachyarrhythmia, unstable angina, or myocardial infarction within preceding month, chronic obstructive pulmonary disease, chest wall abnormalities, chest tube, abdominal distension, body mass index greater than 30, pregnancy, intracranial abnormality.12
The institutional review board (Comitato Etico Interaziendale, Regione Piemonte, Turin) approved the study. If the patient was incompetent at study entry, consent was delayed, the family was informed of the study (although not required), and the study was performed. Written permission for using collected data were hence obtained from the patient (if competent) or from the family (in case of death or if the patient remained incompetent).13
Back to Top | Article Outline
Study Protocol
All consecutive patients who met inclusion and exclusion criteria were treated for 72 h according to the ARDSNet strategy.2 A detailed description of the protocol can be found in the Supplemental Digital Content (see text file, Supplemental Digital Content 1, http://links.lww.com/ALN/A542).
After 72 h of ventilation according to the ARDSNet strategy,2 PPLAT was recorded for a period of 1 h with 0.5-s inspiratory pause keeping ventilator setting constant and abolishing spontaneous respiratory muscles activity by (1) reaching a Ramsay score of sedation 5 (midazolam up to 0.15 mg · kg−1 · h−1, morphine up to 0.03 mg · kg−1 · h−1, and propofol up to 2 mg · kg−1 · h−1)14; (1) increasing, if required before measurements, doses of midazolam (up to 10 mg/h) and/or propofol (150 mg/h increments every 10 min).15,16
Fig. 1
Fig. 1
Image Tools
Fig. 2
Fig. 2
Image Tools
In the patients who had 25 < PPLAT < 28 cm H2O, the ARDSNet strategy2 was maintained at least for the subsequent 72 h (fig. 1). In patients who had 28 ≤ PPLAT ≤ 30 cm H2O, the following strategy was implemented and maintained for at least the next 72 h (fig. 1): (1) VT was stepwise reduced (1 ml/kg of PBW every 4 h) until 25 < PPLAT < 28 cm H2O; (2) to manage the reabsorption atelectasis that may occur during mechanical ventilation with low VT,17,18 PEEP-Fio2 combination was set according to the “higher PEEP” arm of the ALVEOLI study12; (3) respiratory rate was increased up to 40 breaths/min, and bicarbonate was infused up to 20 mEq/h; (4) if pH was no more than 7.25, extracorporeal carbon dioxide removal was initiated by using a modified continuous veno-venous hemofiltration system equipped with a membrane lung with a total membrane surface of 0.33 m2 (Decap®, Hemodec, Salerno, Italy)19 (Lower ARDSNet/Carbon Dioxide Removal; fig. 2).
Femoral vein was accessed via a double lumen catheter (14 F; Arrow International Inc. Reading, PA) inserted with the Seldinger technique and connected with the extracorporeal circuit. Blood flow was driven through the circuit by a roller nonocclusive low-flow pump (0–500 ml/min) through a membrane lung (Polystan SAFE; Maquet, Rastatt, Germany) that was connected to a fresh gas flow source delivering 100% oxygen at a constant rate of 8 l/min. Exiting the membrane lung, blood was driven to a hemofilter (Medica D200, Medolla, Italy). The resulting plasmatic water was recirculated through the membrane lung by a peristaltic pump (0–155 ml/min).
The membrane lung and the hemofilter were coupled in series to: (1) increase the pressure inside the membrane lung by adding the downstream resistance exerted by the hemofilter and therefore reduce the risk of air bubble formation20; (2) minimize the need for heparin by diluting the blood entering the membrane lung by recirculating the plasmatic water separated by the hemofilter21; (3) enhance the performance of the extracorporeal device extracting the carbon dioxide dissolved in the plasmatic water separated by the hemofilter and recirculated through the membrane lung22 (fig. 2).
Pressure developed by the roller pump (arterial pressure) was measured and limited to 120–150 mmHg. Reinfusion pressure (venous pressure) and the pressure gradient across the membrane lung and the hemofilter (drop pressure = premembrane lung pressure – venous pressure) were also measured. Detectors of leaks and bubbles were inserted within the circuit. The circuit, including the membrane lung was primed with saline with a volume that ranged between 140 and 160 ml.
A starting dose of heparin (80 IU/kg bolus and 18 IU · kg−1 · h−1 infusion) was delivered by using a syringe pump included in the system. Heparin continuous infusion was hence titrated to maintain the activated partial thromboplastine time ratio to approximately 1.5.
After 72 h of Lower ARDSNet/Carbon Dioxide Removal strategy, the following weaning trial was conducted once a day: flow through the circuit was set to the lowest value (50 ml/min), VT was increased to 6 ml/kg PBW, and PEEP-Fio2 combination was set according to the conventional ARDSNet strategy.2 If PPLAT with these ventilator settings remained for more that 3 h at less than 28 cm H2O, extracorporeal carbon dioxide removal was interrupted, and conventional ARDSNet ventilatory strategy reestablished.
Back to Top | Article Outline
Measurements
Underlying disease responsible for ARDS, clinical status, and laboratory data on admission (the worst value within 24 h after admission) were recorded for calculation of the Simplified Acute Physiology Score II.23
Back to Top | Article Outline
Clinical Variables
Clinical variables (PPLAT, VT, respiratory rate, PEEP, minute ventilation, Pao2/Fio2 ratio, Paco2, pH, continuous cardiac output [Vigileo system; Edwards LifeScience, Irvine, CA], heparin doses, and activated partial thromboplastine time ratio) were prospectively collected at the following scheduled times: after 72 h of ventilation according to the conventional ARDSNet strategy and, in those patients who had 28 ≤ PPLAT ≤ 30 cm H2O, after lowering VT and before initiating carbon dioxide removal (baseline) and after 60–90 min (T1.5), 24 h (T24), 48 h (T48), and 72 h (T72) after initiation of carbon dioxide removal. Blood flow during carbon dioxide removal was recorded at T1.5, T24, T48, and T72.
Potential complications during the procedure were prospectively classified as mechanical (cannula problems, membrane lung failure, clots in the circuit, air in the circuit, pump malfunction, tubing rupture, catheter displacement, system leaks) and patient-related (vein perforation, significant bleeding [i.e., any bleeding event that required the administration of 1 unit of packed red cells], hemodynamic instability [i.e., 80–90 mmHg increase or a 30–40 mmHg decrease in systolic blood pressure relative to the baseline value or need for inotropic drugs for at least 2 h to maintain systolic blood pressure higher than 85 mmHg or electrocardiogram evidence of ischemia or significant ventricular arrhythmias] ischemic/gangrenous bowel, pneumothorax, renal complications [i.e., occurrence after initiation of carbon dioxide removal of creatinine greater than 1.5 mg/dl], infectious complications [i.e., occurrence after initiation of carbon dioxide removal of culture proven new infection], metabolic [i.e., occurrence after initiation of carbon dioxide removal of glucose of at least 240 mg/dl or hyperbilirubinemia], thromboembolic complications [i.e., occurrence after initiation of deep venous thrombosis or pulmonary embolus], and neurologic complications [i.e., occurrence after initiation of carbon dioxide removal of cerebral infarction, or clinical seizure, or cerebral hemorrhage or cerebral edema]), and their occurrence was recorded during the period that the Lower ARDSNet/Carbon Dioxide Removal strategy was recorded.
Back to Top | Article Outline
Lung Morphology
A pulmonary computed tomography (CT) scan of the whole lung was performed in all patients after study enrollment. The CT scan was repeated after approximately 72 h of ventilation with the Lower ARDSNet/Carbon Dioxide Removal strategy in those patients who had 28 ≤ PPLAT ≤ 30 cm H2O and after approximately 72 h of ventilation with the conventional ARDSNet strategy in some of the patients who had 25 < PPLAT < 28 cm H2O (fig. 1). (See text file, Supplemental Digital Content 1, which is a detailed description of the methods used, http://links.lww.com/ALN/A542).3
Weight and volume of the entire lungs and of the nonaerated, poorly aerated, normally aerated, and hyperinflated compartments were assessed.13,24 Protected tidal inflation and tidal hyperinflation were defined as the volume of the normally aerated and hyperinflated compartment at end-inspiration minus the volume of the normally aerated and hyperinflated compartment at end-expiration, respectively. Tidal recruitment of the nonaerated compartment was defined as the volume of the nonaerated compartment at end-expiration minus the volume at end-inspiration. All were expressed as percent of the total tidal inflation-related change in CT lung volume.3,24
Back to Top | Article Outline
Pulmonary Inflammatory Response
A bronchoalveolar lavage was performed in all patients after study enrollment.25 The bronchoalveolar lavage was repeated after approximately 72 h of ventilation with the Lower ARDSNet/Carbon Dioxide Removal strategy in the patients who had 28 ≤ PPLAT ≤ 30 cm H2O and after approximately further 72 h of ventilation with the conventional ARDSNet strategy in some of the patients who had 25 < PPLAT < 28 cm H2O (fig. 1). Interleukin 6 (IL-6), IL-8, IL-1b, and IL-1 receptor antagonist (IL-1Ra) were measured. (See text file, Supplemental Digital Content 1, which is a detailed description of the methods used, http://links.lww.com/ALN/A542).25
Back to Top | Article Outline
Statistical Analysis
Data are expressed as mean ± SD of the mean or median and interquartile (25th and 75th percentile) range as appropriate. To evaluate differences between the two groups, the Fisher exact test for categorical variables and the t test with unequal variance for continuous variables were used. Data at different times during carbon dioxide removal were compared by analysis of variance (ANOVA) for repeated measures by using a Bonferroni correction. If significant (P ≤ 0.05), the values at baseline, T1.5, T24, T48, and T72 were compared with those obtained after 72 h of ventilation according to the ARDSNet strategy by using a paired t test as modified by Dunnett. All tests were two tailed. Analysis was carried out by using the SPSS software package (SPSS Inc. Chicago, IL).
Back to Top | Article Outline

Results

Table 1
Table 1
Image Tools
Of the 32 patients who matched study criteria, 22 patients had 25 < PPLAT < 28 cm H2O and 10 patients had a 28 ≤ PPLAT ≤ 30 cm H2O. Patient characteristics are shown in table 1. Age, gender, Simplified Acute Physiology Score II, and underlying diseases responsible for ARDS did not differ between the two groups; Paco2:Fio2 ratio was lower in patients with 28 ≤ PPLAT ≤ 30 cm H2O than in patients with 25 < PPLAT < 28 cm H2O (P < 0.01).
Table 2
Table 2
Image Tools
Lungs were heavier, and the extent of the hyperinflated, nonaerated, and poorly aerated CT lung compartments was larger and the extent of the normally aerated compartment was smaller in patients with 28 ≤ PPLAT ≤ 30 cm H2O than in patients with 25 < PPLAT < 28 cm H2O (all P < 0.001) (table 2). Protected tidal inflation was smaller and tidal hyperinflation was larger in patients who had 28 ≤ PPLAT ≤ 30 cm H2O than in the patients who had 25 < PPLAT < 28 cm H2O (19 ± 6 vs. 81 ± 6 and 67 ± 5 vs. 11 ± 4% of the total tidal inflation-associated change in CT lung compartments, respectively; P < 0.01). Tidal recruitment of the poorly aerated compartment did not differ between the two groups of patients.
Fig. 3
Fig. 3
Image Tools
In the patients who had 28 ≤ PPLAT ≤ 30 cm H2O, VT was decreased from 6.3 ± 0.2 to 4.2 ± 0.3 ml/kg PBW and PPLAT decreased from 29.1 ± 1.2 to 25.0 ± 1.2 cm H2O (all P < 0.001). Despite the increase in respiratory rate (from 31.2 ± 2.3 to 37.0 ± 1.9 breaths/min; P < 0.001) and bicarbonate infusion (20.2 ± 0.8 mEq/h), the reduction in minute ventilation (from 12.03 ± 2.77 to 9.03 ± 1.18 l/min; P < 0.001) increased Paco2 (from 48.4 ± 8.7 to 73.6 ± 11.1 mmHg; P < 0.001) and decreased pH (from 7.36 ± 0.03 to 7.20 ± 0.02; P < 0.001). The increase in PEEP (from 12.1 ± 2.5 cm H2O to 15.2 ± 0.8 cm H2O, P < 0.001) attenuated the reduction in Pao2/Fio2 (from 135 ± 30 to 124 ± 29; P < 0.01) associated to the lowering of VT (fig. 3). All patients met pH criteria for carbon dioxide removal and were therefore connected to the veno-venous bypass.
Fig. 4
Fig. 4
Image Tools
Table 3
Table 3
Image Tools
Sixty to ninety minutes of veno-venous bypass decreased Paco2 to 50.4 ± 8.2 mmHg and increased arterial pH to 7.32 ± 0.03 (P < 0.001). After 72 h of extracorporeal support, Paco2 and arterial pH were 47.2 ± 8.6 mmHg and 7.38 ± 0.04, respectively (P < 0.001; fig. 4). The extracorporeal veno-venous carbon dioxide removal device was used for 144 (84, 168) h. The pump-driven blood flow through the circuit ranged between 191 and 422 ml/min (5–10% of cardiac output; table 3). Heparin infusion ranging between 3 and 19 IU/kg was needed to maintain activated partial thromboplastine time ratio between 1.1 and 1.7 (table 3).
Table 4
Table 4
Image Tools
No patient-related complications were observed. Mechanical complications are reported in table 4. In three cases, the 14-French double-lumen catheter had to be replaced by two 8-French simple-lumen catheters (one for each femoral vein) due to recirculation issues (two cases) and catheter kinking (one case). None of the reported malfunctions impaired patient status during the procedure. The membrane clotting observed in three patients did not result in additional transfusion. The Lower ARDSNet/Carbon Dioxide Removal strategy did not require any increase in nursing resources. A technician with expertise in the Decap® system was present (9 am–5 pm) during the treatment of the first five patients.
Fig. 5
Fig. 5
Image Tools
Average lung density histograms of tidal changes in CT lung compartments in patients who had 28 ≤ PPLAT ≤ 30 cm H2O during the ARDSNet strategy are shown at study entry (fig. 5A, left) and after 72 h of Lower ARDSNet/Carbon Dioxide Removal strategy (fig. 5A, right). Ventilation with the Lower ARDSNet/Carbon Dioxide Removal was associated with: (1) the reduction of lung weight and of the extent of hyperinflated, nonaerated, and poorly aerated CT lung compartments and an increase of the extent of normally aerated lung compartments (P < 0.001; table 2); (2) increase of protected tidal inflation (from 19 ± 6% to 86 ± 8% of the total tidal inflation-associated change in CT lung compartments; P < 0.01); (3) the reduction of tidal hyperinflation (from 67 ± 5% to 5 ± 4% of the total tidal inflation-associated change in CT lung compartments; P < 0.01); (4) a significant improvement of Pao2/Fio2 (from 136 ± 30 to 221 ± 56; P < 0.001) (fig. 3). Same data for the patients who during the ARDSNet strategy had 25 < PPLAT < 28 cm H2O are shown at study entry (fig. 5B, left). In 12 of the 22 patients, CT scans were obtained after further 72 h of conventional ARDSNet strategy (fig. 5B, right); these data were retrospectively obtained by using CT scan performed for other studies or for clinical reasons. Relative to study entry, further 72 h of conventional ARDSNet strategy were also associated to the reduction of lung weight and of the extent of the hyperinflated, nonaerated, and poorly aerated CT lung compartments and the increase of the extent of the normally aerated lung compartments (P < 0.05; table 2) with no changes in amount of protected tidal inflation and tidal hyperinflation (fig. 5B, left); a significant improvement of Pao2/Fio2 (from 185 ± 60 to 301 ± 42; P < 0.001) was also observed.
Fig. 6
Fig. 6
Image Tools
Pulmonary concentration of inflammatory cytokines at study entry were lower (P = 0.001) in patients who during the ARDSNet strategy had 25 < PPLAT < 28 cm H2O than in patients who had 28 ≤ PPLAT ≤ 30 cm H2O. In the former, further 72 h of conventional ARDSNet strategy did not modify concentration of pulmonary inflammatory cytokines (data were retrospectively obtained in 15 of the 22 patients by using bronchoalveolar lavage performed for other studies). In the latter, the use of the Lower ARDSNet/Carbon Dioxide Removal strategy was associated with a significantly reduced concentration of pulmonary inflammatory cytokines (P = 0.001; fig. 6). Bronchoalveolar lavage procedure did not cause complications in any of the studied patients.
Back to Top | Article Outline

Discussion

Use of VT lower than 6 ml/kg PBW in patients who during ARDSNet ventilation had 28 ≤ PPLAT ≤ 30 cm H2O was associated to a significant reduction of inflammatory and morphological markers of ventilator-induced lung injury. Respiratory acidosis consequent to further VT reduction was effectively and safely corrected by using a modified renal replacement device incorporating a membrane lung.
These results support the use of extracorporeal carbon dioxide removal as a tool that, integrated with conventional ventilation, allows more protective ventilator settings.9 However, these data can only be used as proof of principle for the following reasons. (1) The confounding effect of time on the observed improvement of physiologic, radiographic, and inflammatory parameters before and after 72 h of Lower ARDSNet/Carbon Dioxide Removal strategy cannot be ruled out because we don’t have a control group of patients with 28 ≤ PPLAT ≤ 30 cm H2O who received usual care without Lower ARDSNet/Carbon Dioxide Removal. Moreover, the study design prevents us from knowing what the respective effect of decreasing VT, increasing PEEP, and application of carbon dioxide removal. (2) We could not assess weather the observed substantial decrease in Paco2 with the limited blood flow was exclusively related to the device used in the study; measurements of extracorporeal carbon dioxide transfer and of total body carbon dioxide production were not performed. (3) Studies in experimental models suggest that unbuffered respiratory acidosis may reduce ventilator-induced lung injury.26 It is therefore possible that physiologic, morphological, and inflammatory variables would have improved as much or more if we reduced VT and PPLAT without carbon dioxide removal. (4) None of the patients who had 28 ≤ PPLAT ≤ 30 cm H2O during “ARDSNet” ventilation were able to reach the target PPLAT values (25–28 cm H2O) with the increase in respiratory rate and bicarbonate infusion. This may be because VT in three patients was reduced more than required; PPLAT in these patients before initiating extracorporeal carbon dioxide removal was 24.2, 23.3, and 24.1 cm H2O, values that are lower than the target PPLAT. Moreover tris-hydroxymethyl aminomethane, a non-carbon dioxide-generating buffer that was recently shown to efficiently manage respiratory acidosis,27,28 was not used in the current study.
Gattinoni et al. hypothesized that, “to rest the lung,” oxygenation via mechanical ventilation could be dissociated from decarboxylation via extracorporeal carbon dioxide removal.5 In that study, mechanical ventilation was limited to apneic oxygenation and to 3–5 sighs every minute with peak inspiratory pressure not higher than 35–45 cm H2O; PEEP ranged between 15 and 25 cm H2O. carbon dioxide removal was performed by using a pump-driven veno-venous bypass, allowing blood flow to pass through two membrane lungs (9 m2 total membrane surface area). Extracorporeal blood flow was progressively raised from 200 to 300 ml/min to the selected maintenance flow (20–30% of cardiac output). Although the observed mortality rate was lower than expected, several episodes of severe bleeding were reported.5 A subsequent randomized clinical trial did not confirm these data.6 Concerns have been raised regarding the standard use of extracorporeal support6,29 because of the high incidence of serious complications such as hemorrhage, hemolysis, and neurologic impairments.8 As a result, extracorporeal carbon dioxide removal has been restricted to the sickest patients in whom all other treatments have failed30 and to centers with large expertise.31
Retrospective evaluation of the ARDSnet database suggested that VT reduction would have improved outcome, even in patients who already had PPLAT < 30 cm H2O.4 Moreover, physiologic32 and morphological3 evidences of tidal hyperinflation have been described in patients in whom the ARDSNet strategy resulted in 28 ≤ PPLAT ≤ 30 cm H2O. These data have therefore challenged the view of extracorporeal support only as rescue therapy33,34,35 and have generated the hypothesis that extracorporeal carbon dioxide removal may be incorporated in a lung protective strategy to allow further reduction of VT and PPLAT.9
The concept originally developed by Pesenti et al. of removing “only a portion of carbon dioxide production”10 has been recently implemented in new devices that may reduce side effects, complexity, and costs of extracorporeal carbon dioxide removal.19,33 Bein et al. recently reported the use of a pump-less extracorporeal device in ARDS patients.34 Retrospective analysis of 90 patients demonstrated that using this device was possible to obtain physiologic values of Paco2 (31–42 mmHg) and pH (7.38–7.50) despite ventilation with low VT (320–470 ml). However, the authors reported a complication rate of 24%, including limb ischemia, compartment syndrome, and intracranial hemorrhage. In addition, continuous intravenous infusion of norepinephrine was needed to maintain the pressure gradient between arterial and venous blood.
In the current study, carbon dioxide removal was performed through a dedicated pump-driven extracorporeal veno-venous circuit with a neonatal membrane lung and a hemofilter coupled in series (fig. 2). The main elements that characterize and differentiate this system are: (1) a blood flow lower than the one used in standard carbon dioxide removal (191–422 ml/min [5–10% of cardiac output] versus 1.5–2.0 l/min [20–30% of cardiac output])5,6,36; (2) a small neonatal membrane lung (0.33 m2) instead of two large adult membrane lung (3–4.5 m2 each)5,6,36; (3) the use of 14-French double-lumen catheters instead of large 21- to 28-French double- or single-lumen catheters9; (4) a priming volume smaller than currently used (140–160 ml vs. 1,500–1,800 ml)37; (5) a relatively small infusion rate of heparin (3–19 IU/kg) and less hypocoagulation than used in previous studies (activated partial thromboplastine time ratio of 1.1–1.7 vs. 2.0–2.5).5,6
Seventy-two hours of use of this extracorporeal bypass was associated to a 33.6 ± 6.3% reduction of Paco2 (from 73.6 ± 11.1 to 48.5 ± 6.3 mmHg, P < 0.001) sufficient to normalize arterial pH (from 7.20 ± 0.02 to 7.38 ± 0.04, P < 0.001) while ventilating patients with VT ranging between 167 and 340 ml (3.7–4.6 ml/Kg PBW) and minute ventilation ranging between 8.1 and 11.9 l/min. No adverse events in terms of vein perforation, significant bleeding, hemodynamic instability, ischemic/gangrenous bowel, pneumothorax, and renal, infectious, metabolic, thromboembolic, and neurologic complications were observed during the 141 ± 69 h in which the device was in clinical use. However, the level of blood flow required to normalize arterial pH (up to 422 ml/h) was relatively high for a circuit like the one used in the current study38; therefore, the 14-French double-lumen catheter had to be replaced in three cases by two 8-French simple-lumen catheters (one for each femoral vein).
Reabsorption atelectasis may occur during mechanical ventilation with low VT, depending on the Fio2, the regional ventilation/perfusion ratios and the end-expiratory lung volume.17 In a recent study, Dembinsky et al. randomized animals with ARDS to be ventilated for 24 h with a VT of 3 ml/kg or with a VT of 6 ml/kg; consequent respiratory acidosis was managed in the former group removing carbon dioxide via a pumpless system.18 The study showed that, despite a significant reduction of PPLAT, organ function and organ injury assessment did not reveal significant improvements when compared with conventional strategy. On the contrary, pulmonary gas exchange was impaired because of increased pulmonary ventilation/perfusion mismatch. In the current study, after 72 h of Lower ARDSNet/Carbon Dioxide Removal strategy, we observed (1) the reduction of lung weight and of the extent of the nonaerated and poorly aerated; (2) the increase of the extent of the normally aerated CT lung compartments lung (table 2); (3) a significant improvement of Pao2/Fio2 (from 136 ± 30 to 221 ± 56; P < 0.001) (fig. 5). These differences may be explained by the fact that, whereas PEEP was set to 5 cm H2O in both groups in the Dembinsky et al. study, PEEP in our study was increased from 12.1 ± 2.5 cm H2O to 15.2 ± 0.8 cm H2O (P < 0.001).12,39
In conclusion, this study suggests that further reduction of VT minimizes tidal hyperinflation and attenuates pulmonary inflammation in ARDS patients who have a PPLAT ranging between 28 and 30 cm H2O when ventilated with the ARDSNet ventilation. Extracorporeal carbon dioxide removal effectively and safely managed the respiratory acidosis consequent to VT lower than 6 l/kg PBW and reestablished a normal arterial pH. This proof of concept study provides preliminary clinical evidence that extracorporeal lung support may integrate conventional care and allow the use of more protective ventilator settings. Additional clinical studies are required to further confirm these results.
Back to Top | Article Outline

References

1.Malhotra A: Low-tidal-volume ventilation in the acute respiratory distress syndrome. N Engl J Med 2007; 357:1113–20

2.ARDS Network: Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000; 342:1301–8

3.Terragni PP, Rosboch G, Tealdi A, Corno E, Menaldo E, Davini O, Gandini G, Herrmann P, Mascia L, Quintel M, Slutsky AS, Gattinoni L, Ranieri VM: Tidal hyperinflation during low tidal volume ventilation in acute respiratory distress syndrome. Am J Respir Crit Care Med 2007; 175:160–6

4.Hager DN, Krishnan JA, Hayden DL, Brower RG: Tidal volume reduction in patients with acute lung injury when plateau pressures are not high. Am J Respir Crit Care Med 2005; 172:1241–5

5.Gattinoni L, Pesenti A, Mascheroni D, Marcolin R, Fumagalli R, Rossi F, Iapichino G, Romagnoli G, Uziel L, Agostoni A, Kolobow T, Damia G: Low-frequency positive-pressure ventilation with extracorporeal CO2 removal in severe acute respiratory failure. JAMA 1986; 256:881–6

6.Morris AH, Wallace CJ, Menlove RL, Clemmer TP, Orme JF Jr, Weaver LK, Dean NC, Thomas F, East TD, Pace NL, Suchyta MR, Beck E, Bombino M, Sitting DF, Bohm S, Hoffmann B, Becks H, Butler S, Pearl J, Rasmusson B: Randomized clinical trial of pressure-controlled inverse ratio ventilation and extracorporeal CO2 removal for adult respiratory distress syndrome. Am J Respir Crit Care Med 1994; 149:295–305

7.Deslauriers J, Awad JA: Is extracorporeal CO2 removal an option in the treatment of adult respiratory distress syndrome? Ann Thorac Surg 1997; 64:1581–2

8.Conrad SA, Rycus PT, Dalton H: Extracorporeal Life Support Registry Report 2004. ASAIO J 2005; 51:4–10

9.Mielck F, Quintel M: Extracorporeal membrane oxygenation. Curr Opin Crit Care 2005; 11:87–93

10.Pesenti A, Rossi GP, Pelosi P, Brazzi L, Gattinoni L: Percutaneous extracorporeal CO2 removal in a patient with bullous emphysema with recurrent bilateral pneumothoraces and respiratory failure. Anesthesiology 1990; 72:571–3

11.Bernard GR, Artigas A, Brigham KL, Carlet J, Falke K, Hudson L, Lamy M, Legall JR, Morris A, Spragg R: The American-European Consensus Conference on ARDS. Definitions, mechanisms, relevant outcomes, and clinical trial coordination. Am J Respir Crit Care Med 1994; 149:818–24

12.Brower RG, Lanken PN, MacIntyre N, Matthay MA, Morris A, Ancukiewicz M, Schoenfeld D, Thompson BT: Higher versus lower positive end-expiratory pressures in patients with the acute respiratory distress syndrome. N Engl J Med 2004; 351:327–36

13.Gattinoni L, Caironi P, Cressoni M, Chiumello D, Ranieri VM, Quintel M, Russo S, Patroniti N, Cornejo R, Bugedo G: Lung recruitment in patients with the acute respiratory distress syndrome. N Engl J Med 2006; 354:1775–86

14.Ramsay MA, Savege TM, Simpson BR, Goodwin R: Controlled sedation with alphaxalone-alphadolone. BMJ 1974; 2:656–9

15.Decailliot FDA, Maggiore SM, Jonson B, Duvaldestin P, Brochard L: Pressure-volume curves with and without muscle paralysis in acute respiratory distress syndrome. Intensive Care Med 2006; 32:1322–8

16.Conti GVV, Rocco M, DeBlasi RA, Lappa A, Bufi M, Antonelli M, Gasparetto A: Paralysis has no effect on chest wall and respiratory system mechanics of mechanically ventilated, sedated patients. Intensive Care Med 1995; 21:808–12

17.Grasso S, Mascia L, Del Turco M, Malacarne P, Giunta F, Brochard L, Slutsky AS, Ranieri VM: Effects of recruiting maneuvers in patients with acute respiratory distress syndrome ventilated with protective ventilatory strategy. Anesthesiology 2002; 96:795–802

18.Dembinski R, Hochhausen N, Terbeck S, Uhlig S, Dassow C, Schneider M, Schachtrupp A, Henzler D, Rossaint R, Kuhlen R: Pumpless extracorporeal lung assist for protective mechanical ventilation in experimental lung injury. Crit Care Med 2007; 35:2359–66

19.Livigni S, Maio M, Ferretti E, Longobardo A, Potenza R, Rivalta L, Selvaggi P, Vergano M, Bertolini G: Efficacy and safety of a low-flow veno-venous carbon dioxide removal device: Results of an experimental study in adult sheep. Crit Care 2006; 10:R151

20.Lynch JE, Riley JB: Microemboli detection on extracorporeal bypass circuits. Perfusion 2008; 23:23–32

21.Shulman RI, Singer M, Rock J: Continuous renal replacement therapy. Keeping the circuit open: Lessons from the lab. Blood Purif 2002; 20:275–81

22.Jones NL: An obsession with CO2. Appl Physiol Nutr Metab 2008; 33:641–50

23.Le Gall S Jr, Saulnier F: A new Simplified Acute Physiology Score (SAPS II) based on a European/North American multicenter study. JAMA 1993; 270:2957–63

24.Grasso S, Terragni P, Mascia L, Fanelli V, Quintel M, Herrmann P, Hedenstierna G, Slutsky AS, Ranieri VM: Airway pressure-time curve profile (stress index) detects tidal recruitment/hyperinflation in experimental acute lung injury. Crit Care Med 2004; 32:1018–27

25.Ranieri VM, Suter PM, Tortorella C, De Tullio R, Dayer JM, Brienza A, Bruno F, Slutsky AS: Effect of mechanical ventilation on inflammatory mediators in patients with acute respiratory distress syndrome: A randomized controlled trial. JAMA 1999; 282:54–61

26.Laffey JG, O’Croinin D, McLoughlin P, Kavanagh BP: Permissive hypercapnia–role in protective lung ventilatory strategies. Intensive Care Med 2004; 30:347–56

27.Kallet RH, Jasmer RM, Luce JM, Lin LH, Marks JD: The treatment of acidosis in acute lung injury with tris-hydroxymethyl aminomethane (THAM). Am J Respir Crit Care Med 2000; 161:1149–53

28.Weber T, Tschernich H, Sitzwohl C, Ullrich R, Germann P, Zimpfer M, Sladen RN, Huemer G: Tromethamine buffer modifies the depressant effect of permissive hypercapnia on myocardial contractility in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med 2000; 162:1361–5

29.Morris AH: Extracorporeal support and patient outcome: Credible causality remains elusive. Crit Care Med 2006; 34:1551–2

30.Lewandowski K, Rossaint R, Pappert D, Gerlach H, Slama KJ, Weidemann H, Frey DJ, Hoffmann O, Keske U, Falke KJ: High survival rate in 122 ARDS patients managed according to a clinical algorithm including extracorporeal membrane oxygenation. Intensive Care Med 1997; 23:819–35

31.Bartlett RH, Roloff DW, Custer JR, Younger JG, Hirschl RB: Extracorporeal life support: The University of Michigan experience. JAMA 2000; 283:904–8

32.Grasso S, Stripoli T, De Michele M, Bruno F, Moschetta M, Angelelli G, Munno I, Ruggiero V, Anaclerio R, Cafarelli A, Driessen B, Fiore T: ARDSnet ventilatory protocol and alveolar hyperinflation: Role of positive end-expiratory pressure. Am J Respir Crit Care Med 2007; 176:761–7

33.Reng M, Philipp A, Kaiser M, Pfeifer M, Gruene S, Schoelmerich J: Pumpless extracorporeal lung assist and adult respiratory distress syndrome. Lancet 2000; 356:219–20

34.Bein T, Weber F, Philipp A, Prasser C, Pfeifer M, Schmid FX, Butz B, Birnbaum D, Taeger K, Schlitt HJ: A new pumpless extracorporeal interventional lung assist in critical hypoxemia/hypercapnia. Crit Care Med 2006; 34:1372–7

35.Gattinoni L, Caironi P, Carlesso E: How to ventilate patients with acute lung injury and acute respiratory distress syndrome. Curr Opin Crit Care 2005; 11:69–76

36.Pesenti A, Pelizzola A, Mascheroni D, Uziel L, Pirovano E, Fox U, Gattinoni L, Kolobow T: Low frequency positive pressure ventilation with extracorporeal CO2 removal (LEPPV-ECCO2R) in acute respiratory failure (ARF): Technique. Trans Am Soc Artif Intern Organs 1981; 27:263–6

37.Gattinoni L, Mascheroni D, Torresin A, Marcolin R, Fumagalli R, Vesconi S, Rossi GP, Rossi F, Baglioni S, Bassi F, Nastri G, Pesenti A: Morphological response to positive end expiratory pressure in acute respiratory failure. Computerized tomography study. Intensive Care Med 1986; 12:137–42

38.Ronco C: Continuous renal replacement therapies in the treatment of acute renal failure in intensive care patients. Part 2. Clinical indications and prescription. Nephrol Dial Transplant 1994; 9(Suppl 4):201–9

39.Corbridge TC, Wood LD, Crawford GP, Chudoba MJ, Yanos J, Sznajder JI: Adverse effects of large tidal volume and low PEEP in canine acid aspiration. Am Rev Respir Dis 1990; 142:311–5

Cited By:

This article has been cited 27 time(s).

Critical Care
Bench to bedside review: Extracorporeal carbon dioxide removal, past present and future
Cove, ME; MacLaren, G; Federspiel, WJ; Kellum, JA
Critical Care, 16(5): -.
ARTN 232
CrossRef
Anaesthesia and Intensive Care
Carbon dioxide clearance in critical care
Tiruvoipati, R; Botha, JA; Pilcher, D; Bailey, M
Anaesthesia and Intensive Care, 41(2): 157-162.

American Journal of Respiratory and Critical Care Medicine
Extracorporeal Membrane Oxygenation for Pandemic Influenza A(H1N1)-induced Acute Respiratory Distress Syndrome A Cohort Study and Propensity-matched Analysis
Pham, T; Combes, A; Roze, H; Chevret, S; Mercat, M; Roch, A; Mourvillier, B; Ara-Somohano, C; Bastien, O; Zogheib, E; Clavel, M; Constan, A; Richard, JCM; Brun-Buisson, C; Brochard, L
American Journal of Respiratory and Critical Care Medicine, 187(3): 276-285.
10.1164/rccm.201205-0815OC
CrossRef
Critical Care
Permissive hypofiltration: an alternative view
Ricci, Z; Romagnoli, S; Emma, F
Critical Care, 16(6): -.
ARTN 458
CrossRef
Acta Anaesthesiologica Scandinavica
Intensive buffering can keep pH above 7.2 for over 4 h during apnea: an experimental porcine study
Hostman, S; Engstrom, J; Hedenstierna, G; Larsson, A
Acta Anaesthesiologica Scandinavica, 57(1): 63-70.
10.1111/aas.12012
CrossRef
Intensive Care Medicine
Blood oxygenation and decarboxylation determinants during venovenous ECMO for respiratory failure in adults
Schmidt, M; Tachon, G; Devilliers, C; Muller, G; Hekimian, G; Brechot, N; Merceron, S; Luyt, CE; Trouillet, JL; Chastre, J; Leprince, P; Combes, A
Intensive Care Medicine, 39(5): 838-846.
10.1007/s00134-012-2785-8
CrossRef
Intensive Care Medicine
Lower tidal volume strategy (approximate to 3 ml/kg) combined with extracorporeal CO2 removal versus 'conventional' protective ventilation (6 ml/kg) in severe ARDS
Bein, T; Weber-Carstens, S; Goldmann, A; Muller, T; Staudinger, T; Brederlau, J; Muellenbach, R; Dembinski, R; Graf, BM; Wewalka, M; Philipp, A; Wernecke, KD; Lubnow, M; Slutsky, AS
Intensive Care Medicine, 39(5): 847-856.
10.1007/s00134-012-2787-6
CrossRef
Intensive Care Medicine
From protective ventilation to super-protective ventilation for acute respiratory distress syndrome
Schultz, MJ; Juffermans, NP; Matthay, MA
Intensive Care Medicine, 39(5): 963-965.
10.1007/s00134-012-2805-8
CrossRef
Deutsche Medizinische Wochenschrift
ARDS - an update
Braune, S; Kluge, S
Deutsche Medizinische Wochenschrift, 138(): 1019-1022.
10.1055/s-0032-1333051
CrossRef
Bmc Medicine
Novel approaches to minimize ventilator-induced lung injury
Fan, E; Villar, J; Slutsky, AS
Bmc Medicine, 11(): -.
ARTN 85
CrossRef
Annals of Thoracic Medicine
Extracorporeal life support for acute respiratory distress syndromes
Hayes, D; Tobias, JD; Kukreja, J; Preston, TJ; Yates, AR; Kirkby, S; Whitson, BA
Annals of Thoracic Medicine, 8(3): 133-141.
10.4103/1817-1737.114290
CrossRef
Chest
A Novel Extracorporeal CO2 Removal System Results of a Pilot Study of Hypercapnic Respiratory Failure in Patients With COPD
Burki, NK; Mani, RK; Herth, FJF; Schmidt, W; Teschler, H; Bonin, F; Becker, H; Randerath, WJ; Stieglitz, S; Hagmeyer, L; Priegnitz, C; Pfeifer, M; Blaas, SH; Putensen, C; Theuerkauf, N; Quintel, M; Moerer, O
Chest, 143(3): 678-686.
10.1378/chest.12-0228
CrossRef
Burns
Strategies to reduce ventilator-associated lung injury (VALI)
Salman, D; Finney, SJ; Griffiths, MJD
Burns, 39(2): 200-211.
10.1016/j.burns.2012.10.013
CrossRef
European Cytokine Network
Inflammatory response in ventilated left and collapsed right lungs, serum and pleural fluid, in transthoracic esophagectomy for cancer
Zingg, U; Forberger, J; Frey, DM; Esterman, AJ; Oertli, D; Beck-Schimmer, B; Zollinger, A
European Cytokine Network, 21(1): 50-57.
10.1684/ecn.2009.0180
CrossRef
Critical Care
Inhaled activated protein C protects mice from ventilator-induced lung injury
Maniatis, NA; Letsiou, E; Orfanos, SE; Kardara, M; Dimopoulou, I; Nakos, G; Lekka, ME; Roussos, C; Armaganidis, A; Kotanidou, A
Critical Care, 14(2): -.
ARTN R70
CrossRef
Nephrology Dialysis Transplantation
Intensive care
Ricci, Z; Cruz, DN; Ronco, C
Nephrology Dialysis Transplantation, 25(3): 678-683.
10.1093/ndt/gfq005
CrossRef
Critical Care
Mechanical ventilation with high tidal volume induces inflammation in patients without lung disease
de Oliveira, RP; Hetzel, MP; Silva, MD; Dallegrave, D; Friedman, G
Critical Care, 14(2): -.
ARTN R39
CrossRef
Transplantation Proceedings
The Use of CO2 Removal Devices in Patients Awaiting Lung Transplantation: An Initial Experience
Ricci, D; Boffini, M; Del Sorbo, L; El Qarra, S; Comoglio, C; Ribezzo, M; Bonato, R; Ranieri, VM; Rinaldi, M
Transplantation Proceedings, 42(4): 1255-1258.
10.1016/j.transproceed.2010.03.117
CrossRef
European Respiratory Journal
Pulmonologists and intensivists: ''two hearts are better than one"
Nava, S; Pelosi, P
European Respiratory Journal, 35(3): 463-466.
10.1183/09031936.00002410
CrossRef
Transplantation Reviews
Intraoperative protective ventilation strategies in lung transplantation
Verbeek, GL; Myles, PS
Transplantation Reviews, 27(1): 30-35.
10.1016/j.trre.2012.11.004
CrossRef
Deutsche Medizinische Wochenschrift
Extracorporeal membrane oxygenation
Karagiannidis, C; Philipp, A; Buchwald, D
Deutsche Medizinische Wochenschrift, 138(5): 188-191.
10.1055/s-0032-1332822
CrossRef
Critical Care
Preliminary study of ventilation with 4 ml/kg tidal volume in acute respiratory distress syndrome: feasibility and effects on cyclic recruitment - derecruitment and hyperinflation
Retamal, J; Libuy, J; Jimenez, M; Delgado, M; Besa, C; Bugedo, G; Bruhn, A
Critical Care, 17(1): -.
10.1186/cc12487
CrossRef
Intensive Care Medicine
What is new in extracorporeal membrane oxygenation for ARDS in adults?
Abrams, D; Brodie, D; Combes, A
Intensive Care Medicine, 39(): 2028-2030.
10.1007/s00134-013-3035-4
CrossRef
New England Journal of Medicine
High-Frequency Oscillation in Early Acute Respiratory Distress Syndrome
Ferguson, ND; Cook, DJ; Guyatt, GH; Mehta, S; Hand, L; Austin, P; Zhou, Q; Matte, A; Walter, SD; Lamontagne, F; Granton, JT; Arabi, YM; Arroliga, AC; Stewart, TE; Slutsky, AS; Meade, MO
New England Journal of Medicine, 368(9): 795-805.
10.1056/NEJMoa1215554
CrossRef
Anesthesiology
Case Scenario: Acute Postoperative Negative Pressure Pulmonary Edema
Krodel, D; Bittner, E; Abdulnour, R; Brown, R; Eikermann, M
Anesthesiology, 113(1): 200-207.
10.1097/ALN.0b013e3181e32e68
PDF (1333) | CrossRef
Anesthesiology
Ventilator-induced Lung Injury: Less Ventilation, Less Injury
Bigatello, LM; Pesenti, A
Anesthesiology, 111(4): 699-700.
10.1097/ALN.0b013e3181b76545
PDF (87) | CrossRef
Current Opinion in Critical Care
Ventilatory support for acute respiratory failure: new and ongoing pathophysiological, diagnostic and therapeutic developments
Del Sorbo, L; Slutsky, AS
Current Opinion in Critical Care, 16(1): 1-7.
10.1097/MCC.0b013e32833500bc
PDF (424) | CrossRef
Back to Top | Article Outline

Supplemental Digital Content

Back to Top | Article Outline

© 2009 American Society of Anesthesiologists, Inc.

Publication of an advertisement in Anesthesiology Online does not constitute endorsement by the American Society of Anesthesiologists, Inc. or Lippincott Williams & Wilkins, Inc. of the product or service being advertised.
Login

Article Tools

Images

Share