Effects of Different Levels of Pressure Support Variability in Experimental Lung Injury

Spieth, Peter M. M.D.*; Carvalho, Alysson R. M.Sc., D.Sc.*; Güldner, Andreas M.D.*; Pelosi, Paolo M.D., Ph.D.‡; Kirichuk, Oleg M.D.*; Koch, Thea M.D., Ph.D.§; de Abreu, Marcelo Gama M.D., M.Sc., Ph.D., D.E.A.A.∥

Free Access
Article Outline
Collapse Box

Author Information

Collapse Box

Abstract

Background: Noisy pressure support ventilation has been reported to improve respiratory function compared to conventional assisted mechanical ventilation. We aimed at determining the optimal level of pressure support variability during noisy pressure support ventilation.
Methods: Twelve pigs were anesthetized and mechanically ventilated. Acute lung injury was induced by surfactant depletion. At four levels of pressure support variability (coefficients of variation of pressure support equal to 7.5, 15, 30, and 45%, 30 min each, crossover design, special Latin squares sequence), we measured respiratory variables, gas exchange, hemodynamics, inspiratory effort, and comfort of breathing. The mean level of tidal volume was constant among variability levels.
Results: Compared to conventional pressure support ventilation, different levels of variability in pressure support improved the elastance of the respiratory system, peak airway pressure, oxygenation, and intrapulmonary shunt. Oxygenation and venous admixture benefited more from intermediate (30%) levels of variability, whereas elastance and peak airway pressure improved linearly with increasing variability. Heart rate as well as mean arterial and pulmonary arterial pressures decreased slightly at intermediate to high (30–45%) levels of variability in pressure support. Inspiratory effort and comfort of breathing were not importantly influenced by increased variability in pressure support.
Conclusion: In a surfactant depletion model of acute lung injury, variability of pressure support improves lung function. The variability level of 30% seems to represent a reasonable compromise to improve lung functional variables during noisy pressure support ventilation.
IN patients suffering from the acute respiratory distress syndrome, mechanical ventilation may be required to treat severe gas exchange impairment and avoid fatigue of respiratory muscles. Volume assist-control is the most common ventilator mode used in this scenario.1 However, controlled ventilation requires deep sedation and sometimes muscle paralysis, which can result in diaphragmatic dysfunction.2–4 Different studies suggest that a certain amount of spontaneous ventilatory effort may be beneficial during mechanical ventilation, not only by avoiding diaphragm dysfunction, but also by improving respiratory mechanics and regional ventilation/perfusion matching.5–8 More recently, the variation of the breathing pattern, i.e., the use of noise, has been reported to improve lung function during both controlled9,10 and pressure support ventilation.11
The degree of variability of tidal volumes (VT) and respiratory frequency (RR) may differently affect lung functional variables. Suki and coworkers postulated that the lungs behave like a stochastic resonance system.12 According to this hypothesis, the level of noise in VT, which represents the input of the system, may influence the amplitude of its output, most notably the oxygenation. Accordingly, excessive as well as lack of variability in VT may worsen lung function. In controlled mechanical ventilation, it has been suggested that variation of VT within 40 to 60% of mean value resulted in improved respiratory system mechanics and oxygenation in endotoxin-induced lung injury.13
In the surfactant depletion model of acute lung injury (ALI), we found that noise in pressure support leading to approximately 20% variation in VT (coefficient of variation [CV], normal distribution) was able to improve oxygenation compared to conventional assisted mechanical ventilation.11 We termed that novel mode of assisted mechanical ventilation “noisy pressure support ventilation (noisy PSV).” However, we did not assess the effects of different degrees of VT variability with noisy PSV.
Basically, noisy PSV differs from other assisted mechanical ventilation modes that may also increase the variability of the respiratory pattern (e.g., proportional assist ventilation) by the fact that the variability does not depend on changes in the patient's inspiratory efforts; rather, it is generated externally by the mechanical ventilator. Thus, noisy PSV is able to guarantee a given level of variability by generating different pressure support values, even if the patient is not able to vary the respiratory pattern due to the underlying disease or sedation.
The aim of this study was to determine the optimal variability for noisy PSV in experimental ALI based on its effects on respiratory mechanics, breathing comfort, gas exchange, and hemodynamics. We hypothesized that noise in pressure support leads to variations in VT that are able to improve lung function and that physiologic variables respond differently to the degree of variability in pressure support.
Back to Top | Article Outline

Materials and Methods

Fig. 1
Fig. 1
Image Tools
The protocol was approved by the Institutional Animal Care Committee and the Government of the State of Saxony, Germany. Figure 1 shows the sequence of interventions performed, which are described in detail in this section. Throughout this work, variability is used as a synonym for CV, unless stated otherwise.
Back to Top | Article Outline
Anesthesia and Mechanical Ventilation
Twelve juvenile female pigs with a mean bodyweight of 29.0 kg (range 26.6–31.0 kg) were premedicated with 10 mg/kg ketamine (ketamin-ratiopharm; Ratiopharm, Ulm, Germany) and 1 mg/kg midazolam (midazolam-ratiopharm, Ratiopharm). Animals had their trachea intubated with a cuffed 8.0-mm ID endotracheal tube (Malinckrodt, Athlone, Ireland). Anesthesia was deepened and maintained by means of continuous intravenous application of midazolam (initial bolus of 0.5–1.0 mg · kg−1; maintenance with 1.5–6 mg · kg−1 · h−1) and ketamine (initial bolus of 3–4 mg · kg−1; maintenance with 5–30 mg · kg−1 · h−1). Animals were kept in the supine position during the whole experiment. Paralysis was achieved by administration of 4 mg of pancuronium (pancuronium-ratiopharm, Ratiopharm) before baseline and injury measurements. Volume-controlled mechanical ventilation was performed using an intensive care respirator (EVITA XL 4Lab; Dräger Medical, Lübeck, Germany). The ventilator settings during baseline and injury were as follows: fraction of inspired oxygen (Fio2) = 1.0; VT = 10 ml/kg; positive end-expiratory pressure (PEEP) = 5 cm H2O; ratio of inspiratory to expiratory time (I:E) = 1:1 to minimize inspiratory pressures. RR was adjusted to achieve Paco2 in the range of 35–45 mmHg. Volume status was maintained with infusion of a crystalloid solution (E153: osmolarity = 303 mOsm/L, Na = 140 mm, K = 5 mm, Ca = 2.5 mm, Mg = 1.5 mm, Cl = 104.5 mm, acetate = 50 mm; Serumwerke Bernburg, Bernburg, Germany) at 20–40 ml · kg−1 · h−1 to keep pulmonary capillary wedge pressure constant below 14 mmHg.
Back to Top | Article Outline
Instrumentation and Sensor Placement
After surgical preparation of the right internal carotid artery and the right external jugular vein, an indwelling catheter was inserted into the carotid artery to measure the arterial blood pressure (BP) continuously and obtain blood samples. A pulmonary artery catheter (Abbott, Abbott Park, IL) was advanced through the external jugular vein until typical pulmonary artery pressure waveforms could be observed.
The signals of airway pressure, esophageal pressure and airway flow were acquired continuously as described elsewhere.11,14 Briefly, a heated pneumotachograph (Fleisch No. 2; Fleisch, Lausanne, Switzerland) connected to a differential pressure transducer (PXL12X5DN; Sensortechnics, Troy, NY) was placed between the Y-piece of the mechanical ventilator tubing and endotracheal tube to determine V̇. Airway pressure was monitored by a second pressure transducer (SCX01DNC; SenSym ICT, Milpitas, CA.) placed at the proximal end of the endotracheal tube. An esophageal balloon catheter (Erich Jaeger, Höchberg, Germany) was advanced into the mid chest and connected to a pressure transducer (SCX01DNC, SenSym ICT). The signals of airway pressure, esophageal pressure, and airway flow were acquired by a LabVIEW-based data acquisition system (National Instruments, Austin, TX).
Back to Top | Article Outline
Blood Gases and Hemodynamics
Arterial and mixed venous blood samples were analyzed using a standard blood gas analyzer (ABL 505; Radiometer, Copenhagen, Denmark). Oxygen saturation and hemoglobin concentration were measured using an OSM 3 Hemoximeter (Radiometer) calibrated for swine blood. Heart rate, mean arterial BP, and mean pulmonary arterial BP were measured using a commercial monitor (CMS; Agilent, Böblingen, Germany). Cardiac output was determined by the conventional bolus thermodilution method as described elsewhere.14 Venous admixture (Q̇va/Q̇t) was calculated using standard formulae.
Back to Top | Article Outline
Respiratory Mechanics and Derived Parameters
Dynamic respiratory mechanics and derived parameters were calculated offline from continuous recordings (2 min during controlled ventilation and 5 min during assisted ventilation) of airway pressure, esophageal pressure, and airway flow, as described in detail before.14 The product of esophageal pressure over time was calculated during inspiration, taking the first value at the beginning of the respiratory cycle as offset. Airway pressure at 100 ms after beginning of inspiration (P0.1) was determined and used as surrogate of the central respiratory drive. Inspiratory pressure time product and P0.1 values were averaged throughout the whole acquisition periods. Comfort of breathing was evaluated with the Aachen Breathing Comfort Score as proposed by Henzler et al.15
Back to Top | Article Outline
Noisy PSV
Normalized sets of 600 randomly generated, normally distributed pressure values with mean ± SD = 1 ± 0.075, 1 ± 0.15, 1 ± 0.30, and 1 ± 0.45 were created, correspondeding to the levels of variability of 7.5, 15, 30, and 45%, respectively. The minimal level of variability of pressure support was 7.5% because it is only slightly higher than the intrinsic variability observed during traditional PSV in our previous study.11 The highest level of variability of pressure support was 45% to avoid Ppeak incompatible with clinical use.
To obtain the sequence of pressure support levels to be effectively used during the experiments, each set of normalized values was multiplied by the target mean pressure support. The target mean pressure support represented the value needed to obtain a VT of 6 ml/kg. After completion of a cycle of 600 breaths, the system looped itself. For safety reasons, the pressure limit of the ventilator was set at 40 cm H2O throughout the whole experiment.
Back to Top | Article Outline
Protocol of Measurements
Initially, the lungs were recruited with an inspiratory pressure of 30 cm H2O for 30 s, and the animals were allowed to stabilize for 15 min. Then, baseline measurements were obtained under volume-controlled mechanical ventilation (Baseline volume controlled).
Acute lung injury was induced by repetitive lung lavage of surfactant with warmed (37°C) 0.9% saline.16 Injury was considered stable if Pao2/Fio2 was less than 100 mmHg for at least 30 min. Thereafter, measurements of acute lung injury under volume controlled mechanical ventilation (Injury) with the same settings of Baseline volume controlled were performed.
To resume spontaneous breathing, the depth of anesthesia was decreased (midazolam = 1–2 mg · kg−1 · h−1, ketamine = 5–15 mg · kg−1 · h−1). When inspiratory efforts could be observed in the esophageal pressure signal, the mechanical ventilation mode was changed to conventional PSV with the following settings: Fio2 = 0.7, PEEP = 10 cm H2O (stepwise change), inspiratory pressure = adjusted to reach a VT of 6 ml/kg, flow trigger = 2 l/min. PEEP was increased to permit lung recruitment and stabilization, reproducing usual clinical practice. After a stabilization period of 30–60 min, the ventilator was set at the continuous positive airway pressure mode with PEEP of 5 cm H2O for 2 min to reset the pulmonary volume history (derecruitment maneuver). Thereafter, PSV was resumed for 30 min with the same settings as described above, being followed by baseline measurements during acute lung injury and assisted mechanical ventilation (baseline PSV). The derecruitment maneuver was repeated after baseline PSV as well as after every subsequent variability level of noisy PSV in order to restore the pulmonary volume history and minimize possible carryover effects among the different variability levels within the crossover design. Animals were ventilated with different degrees of pressure support variability (noisy PSV 7.5%, 15%, 30%, and 45%, respectively) for 30 min, and measurements were taken at the end of each level of variability. Except to the degree of variability, the settings of noisy PSV were the same as described for baseline PSV.
The sequence of variability modes in this crossover design was determined for each animal according to a special 4 × 4 (therapies × animals) Latin square.11 The following sequences were used: A-B-C-D; B-D-A-C; D-C-B-A; C-A-D-B, where A, B, C, and D represent the degrees of variability tested. Each sequence was repeated 3 times, for a total of 12 animals. According to this design, a given degree of variability is never preceded or followed by the same degree of variability twice within a block of 4 animals, and all animals are treated with all degrees of variability to balance for possible carryover effects.
At the end of the experiment, animals were killed by bolus injection of 2 g of thiopental (Altana, Konstanz, Germany) and 50 ml of KCl 1m (Serumwerke Bernburg).
Back to Top | Article Outline
Statistical Analyses
Values are given as mean ± SD or median and 25–75% interquartiles for the CV of selected variables. Comparisons of selected CVs at baseline PSV, and different levels of variability were tested nonparametrically with the Wilcoxon test. The response of the respiratory system to different levels of noisy PSV variability as compared to conventional PSV was assessed by paired t tests. Multiple comparisons in univariate tests were adjusted according the Bonferroni procedure. General linear model statistics were used to determine the effects of the four levels of variability on functional variables of the respiratory system (within-subjects factor = four degrees of variability; planned contrasts for the degree of variability = linear and quadratic). Multiple measurements were adjusted according to Sidak.17 All statistical tests were performed using the Software SPSS (Vers. 15.0, Chicago, IL). Statistical significance was accepted at P < 0.05 (two-tailed for all tests).
Back to Top | Article Outline

Results

Fig. 2
Fig. 2
Image Tools
Figure 2 shows typical recordings of airway pressure and V· signals obtained with the different levels of variability during PSV. While conventional PSV resulted in an almost monotonic respiratory pattern, noisy PSV led to a polymorph pattern which heterogeneity increased with the variability in pressure support.
Fig. 3
Fig. 3
Image Tools
Figure 3 shows typical recordings of VT, Ppeak, and RR for the different levels of variability in pressure support in one representative animal. Although mean VT remained unchanged, the dispersion of VT became higher with increasing variability in pressure support. A similar increase in dispersion could be observed with Ppeak and RR. Mean values of Ppeak decreased with increasing variability, whereas RR evidenced a nadir at a variability level of 30%.
Table 1
Table 1
Image Tools
As shown in table 1, the CV of VT, Ppeak, and RR increased with variability of pressure support, although they were lower than the variability of the input signal set at the ventilator. Whereas the CV of Ppeak began to increase at the pressure support variability level of 7.5%, CVs of VT and RR started to increase at 15% and 45% variability levels, respectively.
Table 2
Table 2
Image Tools
Table 2 depicts the effects of variability of PSV on respiratory variables. RR did not show significant differences among variability levels or as compared to baseline PSV. Minute ventilation was lower at 15%, 30%, and 45% PSV variability as compared to baseline PSV and differed significantly among variability levels. VT was comparable among variability levels. Levels of variability in pressure support of 30% and 45% were associated with lower Ppeak compared to baseline PSV. Mean Ppeak differed among variability levels and decreased linearly with increasing variability. Pmean decreased by only 0.3 cm H2O at variability levels of 15 and 45% compared to baseline PSV. No significant differences in Pmean were detected among the different levels of pressure support variability.
Fig. 4
Fig. 4
Image Tools
Figure 4 shows typical recordings of elastance of the respiratory system (Ers) at each level of variability in pressure support in one representative animal. Dispersion of Ers increased, while mean values decreased with variability. As shown in table 2, mean Ers decreased with variability levels of 30 and 45% compared to Baseline PSV. In addition, Ers values differed significantly among variability levels, improving linearly with increasing variability.
Variability in pressure support did not result in significant changes in inspiratory pressure time product and P0.1 compared to baseline PSV (table 2). Also, inspiratory pressure time product and P0.1 did not differ among different pressure support variability levels. The Aachen Breathing Comfort Score indicated a relatively high degree of breathing comfort (≥ 43 of 60) with both conventional and noisy PSV. The variability level of 30% was associated with a slightly higher improvement in breathing comfort compared to Baseline PSV.
Table 3
Table 3
Image Tools
The effects on gas exchange and hemodynamics are shown in table 3. All variability levels improved Pao2/Fio2 and Q̇va/Q̇t, but not Paco2, as compared to baseline PSV. Contrast analysis showed more pronounced effects on Pao2/Fio2 and Q̇va/Q̇t at intermediate levels (15 to 30%) of variability. Heart rate, mean arterial BP, and pulmonary arterial BP decreased linearly with increasing variability levels in pressure support.
Back to Top | Article Outline

Discussion

Our major findings were that: (1) compared to conventional PSV, different levels of variability in pressure support improved Ers, Ppeak, Pao2/Fio2, and Q̇va/Q̇t without influencing Paco2; (2) Pao2/Fio2 and Q̇va/Q̇t benefitted more from intermediate (30%) levels of variability, but Ers and Ppeak improved linearly with increasing variability; (3) heart rate, mean arterial BP, and pulmonary arterial BP decreased slightly at intermediate (30%) to high (45%) levels of variability in pressure support; and (4) inspiratory effort and comfort of breathing were not importantly influenced by increased variability in pressure support.
Biologic systems seem to benefit from noise, and the loss of variability is usually associated with organ dysfunction.12,18 During controlled mechanical ventilation, the use of biologic variable respiratory patterns are associated with improved lung function under conditions as different as ALI,10 asthma,19 atelectasis,20,21 and one-lung ventilation.22 Suki et al.12 suggested that the respiratory system may function as a stochastic resonance system, where variability of the input signal (e.g., breathing pattern) influences the amplitude of the output signal (e.g., gas exchange and lung mechanics). According to this hypothesis, the variability of VT and RR can be tuned to improve the respiratory variables.
PSV represents the most used form of assisted ventilation.1 During PSV, the subject is able to initiate the inspiratory cycle by triggering the ventilator, partially controlling the inspiratory time through the inspiratory and expiratory muscular activity. However, being the pressure support level and the inspiratory effort constant due to use of sedation and/or underlying disease, the variability of the respiratory pattern may be reduced. Recently, we showed that noisy PSV is superior to conventional PSV to improve respiratory function.11 The present study confirms that finding. Moreover, we found that the level of variability in pressure support affects the breathing pattern, Ers, and oxygenation.
Noisy PSV cannot be considered simply as a combination of sighs with conventional PSV. Patroniti et al.23 showed that intermittent sighs during PSV may improve respiratory function in clinical ALI. However, in our previous work, noisy PSV was found to improve oxygenation compared to PSV+sighs.11 Theoretically, variability could also be applied to other settings of PSV to yield a variable respiratory pattern. For example, the cycling-off criteria could be modulated to obtain variable VT. We opted for variation of pressure support because of its direct association with VT and ease of remote control.
Back to Top | Article Outline
Effects on Breathing Pattern
The increased variability of pressure support was associated with a progressive increase in the variability of VT. The variability of Ppeak also increased, but absolute values were lower than those set for pressure support due to the fact that the maximal inspiratory pressure was limited to 40 cm H2O to protect the lungs against excessive inflation. The fact that CV of VT was higher than CV of Ppeak is most probably explained by the lung pressure-volume curve, which may make the distribution curve of VT flatter than that of Ppeak values. Surprisingly, we found that the variability levels of pressure support of 7.5, 15, and 30% were not associated with higher variability in RR than conventional PSV. This suggests that the respiratory center triggered the inspiration at variable time intervals, whereas VT and Ppeak depended more importantly on the levels of pressure support. This hypothesis is supported by our previous finding that the inspiratory effort does not correlate with VT during noisy PSV.11
Back to Top | Article Outline
Effects on Respiratory Variables
Noisy PSV markedly reduced Ers and Ppeak, as compared to conventional PSV. Different hypotheses can explain this observation: (1) recruitment of previously collapsed lung regions; (2) different distribution of alveolar inflation; (3) structural changes in the mechanical properties of the lung tissue. In our previous study, we did not find evidence of recruitment during noisy PSV.11 However, PEEP levels used in that study were lower than in the present one (5 vs. 10 cm H2O). It is possible that the level of PEEP as used in the present study (10 cm H2O) was enough to keep the lungs open after recruitment induced by noisy PSV. We cannot exclude that more homogeneous redistribution of ventilation and/or structural changes in the mechanical properties of the lung tissues induced by noisy PSV could have contributed to this finding. During controlled mechanical ventilation, Arold et al. also found a progressive decrease in tissue elastance with increased variability.13
Minute ventilation decreased with increasing variability of pressure support as compared to conventional PSV mainly due to a reduction in mean RR with constant mean VT. The reduction in RR during noisy PSV could be explained by the Hering-Breuer reflex; higher end-inspiratory volumes in the lungs occurred more frequently at higher variability levels of pressure support.
The fact that noisy PSV did not lead to clinically relevant effects on inspiratory effort and breathing comfort is in agreement with our previous data showing that variability in pressure support of 30% does not change inspiratory pressure time product or P0.1 compared to conventional PSV.11
Back to Top | Article Outline
Effects on Gas Exchange and Hemodynamics
We confirmed our previous finding that the use of noisy PSV improves Pao2/Fio2 and Q̇va/Q̇t without affecting Paco2 compared to conventional PSV.11 In addition, we observed that the variability of pressure support of 30% optimized oxygenation, although absolute Pao2/Fio2 values were relatively high. This is likely the result of recruitment due to increase of PEEP. Since one important mechanism of improvement of oxygenation during noisy PSV seems to be redistribution of perfusion towards the better aerated nondependent areas of lungs,11 our data suggest that the variability of pressure support does not have a relevant effect on regional perfusion. In addition, we used higher PEEP levels than in our previous evaluation of noisy PSV (10 vs. 5 cm H2O); it is therefore likely that lung recruitment did play a role in the improvements observed in the present study, with increased variability of pressure support, as supported by the improvement in Ers. Unfortunately, we cannot distinguish between recruitment and perfusion-distribution effects. However, it was beyond the scope of this study to the address the mechanisms of noisy PSV.
At highest (45%) levels of pressure support variability, inspiratory pressures could have been high enough to squeeze out regional intrathoracic blood volume, contributing to mismatch of ventilation-perfusion ratio24 and explaining the decrease in Pao2/Fio2 at that level. However, the fact that Paco2 did not change despite decreased minute ventilation suggests that ventilation-perfusion matching improved and dead space decreased at higher pressure support variability. Thus, it is likely that even at the lowest (7.5%) level of pressure support variability, inspiratory pressures in isolated breath cycles were higher than local opening pressures in some lung areas.25
The decrease in mean pulmonary arterial BP with increased variability could be explained by redistribution of pulmonary blood flow towards vascular areas with lower impedance11 and also increased cross-sectional lung capillary area.
Back to Top | Article Outline
Possible Implications in Clinical Practice
The intrinsic variability of the respiratory drive may be reduced due to the underlying disease and use of sedation; therefore, noisy PSV could prove useful to increase the variability of the respiratory pattern as a means to improve lung function during assisted spontaneous breathing. Obviously, noisy PSV should not replace judicious dosing of sedative drugs.
Back to Top | Article Outline
Limitations
Our study has several limitations. First, the lung injury model used does not reproduce all complex clinical features of ALI and, therefore, precludes direct extrapolation of our results to other ALI models and the clinical scenario. Second, we limited our observational period to 30 min for each level of variability in pressure support. Experimental models of ALI may be unstable, and we focused on functional variables, so we tried to keep the observational time as short as possible to allow comparability among the different levels of variability. Third, baseline PSV was not performed in randomized sequence; therefore, we cannot exclude that improvement of gas exchange in noisy PSV was biased by certain instability of the lung injury model over time. However, we compensated for that by using a Latin square design and periodically derecruitment maneuvers before each level of variability. Fourth, the range of variability in VT was no higher than approximately 40%; however, the variability of VT in normal subjects is situated in the range of 20–30%.18,26 Moreover, higher variability in VT could promote lung injury by excessive stretching, which could limit the clinical applicability of our results. Fifth, although the use of a crossover design increased the power of the analysis of functional variables, it precluded the measurement of inflammatory response. Thus, before noisy PSV can be considered for clinical use, its impact on lung inflammation must be determined.
Back to Top | Article Outline

Conclusion

In an experimental surfactant depletion model of acute lung injury, variability of pressure support improved the respiratory function. High variability (45%) levels of pressure support improved Ppeak and Ers, and moderate variability (30%) levels improved Pao2/Fio2 and Q̇va/Q̇t. In addition, variability of pressure support had no clinical relevant influence on inspiratory effort or comfort of breathing. Our findings suggest that a variability level of 30% in pressure support represent the best compromise to improve pulmonary function during noisy PSV.
The authors thank. Thomas Handzsuj, Eng., Dräger Medical, Lübeck, Germany, for his support during the development and implementation of noisy PSV. The authors also thank the students of the Pulmonary Engineering Group, Dresden, Germany, for their support during the experiments.
Back to Top | Article Outline

References

1. Esteban A, Ferguson ND, Meade MO, Frutos-Vivar F, Apezteguia C, Brochard L, Raymondos K, Nin N, Hurtado J, Tomicic V, Gonzalez M, Elizalde J, Nightingale P, Abroug F, Pelosi P, Arabi Y, Moreno R, Jibaja M, D'Empaire G, Sandi F, Matamis D, Montanez AM, Anzueto A: Evolution of mechanical ventilation in response to clinical research. Am J Respir Crit Care Med 2008; 177:170–7

2. Froese AB, Bryan AC: Effects of anesthesia and paralysis on diaphragmatic mechanics in man. Anesthesiology 1974; 41:242–55

3. Falk DJ, DeRuisseau KC, Van Gammeren DL, Deering MA, Kavazis AN, Powers SK: Mechanical ventilation promotes redox status alterations in the diaphragm. J Appl Physiol 2006; 101:1017–24

4. Levine S, Nguyen T, Taylor N, Friscia ME, Budak MT, Rothenberg P, Zhu J, Sachdeva R, Sonnad S, Kaiser LR, Rubinstein NA, Powers SK, Shrager JB: Rapid disuse atrophy of diaphragm fibers in mechanically ventilated humans. N Engl J Med 2008; 358:1327–35

5. Putensen C, Zech S, Wrigge H, Zinserling J, Stuber F, von Spiegel T, Mutz N: Long-term effects of spontaneous breathing during ventilatory support in patients with acute lung injury. Am J Respir Crit Care Med 2001; 164:43–9

6. Putensen C, Hering R, Wrigge H: Controlled versus assisted mechanical ventilation. Curr Opin Crit Care 2002; 8:51–7

7. Putensen C, Mutz NJ, Putensen-Himmer G, Zinserling J: Spontaneous breathing during ventilatory support improves ventilation-perfusion distributions in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med 1999; 159:1241–8

8. Dembinski R, Max M, Bensberg R, Rossaint R, Kuhlen R: Pressure support compared with controlled mechanical ventilation in experimental lung injury. Anesth Analg 2002; 94:1570–6

9. Mutch WA, Eschun GM, Kowalski SE, Graham MR, Girling LG, Lefevre GR: Biologically variable ventilation prevents deterioration of gas exchange during prolonged anaesthesia. Br J Anaesth 2000; 84:197–203

10. Bellardine CL, Hoffman AM, Tsai L, Ingenito EP, Arold SP, Lutchen KR, Suki B: Comparison of variable and conventional ventilation in a sheep saline lavage lung injury model. Crit Care Med 2006; 34:439–45

11. Gama de Abreu M, Spieth PM, Pelosi P, Carvalho AR, Walter C, Schreiber-Ferstl A, Aikele P, Neykova B, Hübler M, Koch T: Noisy pressure support ventilation: A pilot study on a new assisted ventilation mode in experimental lung injury. Crit Care Med 2008; 36:818–27

12. Suki B, Alencar AM, Sujeer MK, Lutchen KR, Collins JJ, Andrade JS, Ingenito EP, Zapperi S, Stanley HE: Life-support system benefits from noise. Nature 1998; 393:127–8

13. Arold SP, Mora R, Lutchen KR, Ingenito EP, Suki B: Variable tidal volume ventilation improves lung mechanics and gas exchange in a rodent model of acute lung injury. Am J Respir Crit Care Med 2002; 165:366–71

14. Gama de Abreu M, Quelhas AD, Spieth P, Brauer G, Knels L, Kasper M, Pino AV, Bleyl JU, Hubler M, Bozza F, Salluh J, Kuhlisch E, Giannella-Neto A, Koch T: Comparative effects of vaporized perfluorohexane and partial liquid ventilation in oleic acid- induced lung injury. Anesthesiology 2006; 104:278–89

15. Henzler D, Pelosi P, Bensberg R, Dembinski R, Quintel M, Pielen V, Rossaint R, Kuhlen R: Effects of partial ventilatory support modalities on respiratory function in severe hypoxemic lung injury. Crit Care Med 2006; 34:1738–45

16. Lachmann B, Robertson B, Vogel J: In vivo lung lavage as an experimental model of the respiratory distress syndrome. Acta Anaesthesiol Scand 1980; 24:231–6

17. Leon AC, Heo M: A comparison of multiplicity adjustment strategies for correlated binary endpoints. J Biopharm Stat 2005; 15:839–55

18. Tobin MJ, Mador MJ, Guenther SM, Lodato RF, Sackner MA: Variability of resting respiratory drive and timing in healthy subjects. J Appl Physiol 1988; 65:309–17

19. Mutch WA, Buchman TG, Girling LG, Walker EK, McManus BM, Graham MR: Biologically variable ventilation improves gas exchange and respiratory mechanics in a model of severe bronchospasm. Crit Care Med 2007; 35:1749–55

20. Boker A, Haberman CJ, Girling L, Guzman RP, Louridas G, Tanner JR, Cheang M, Maycher BW, Bell DD, Doak GJ: Variable ventilation improves perioperative lung function in patients undergoing abdominal aortic aneurysmectomy. Anesthesiology 2004; 100:608–16

21. Mutch WA, Harms S, Ruth GM, Kowalski SE, Girling LG, Lefevre GR: Biologically variable or naturally noisy mechanical ventilation recruits atelectatic lung. Am J Respir Crit Care Med 2000; 162:319–23

22. McMullen MC, Girling LG, Graham MR, Mutch WA: Biologically variable ventilation improves oxygenation and respiratory mechanics during one-lung ventilation. Anesthesiology 2006; 105:91–7

23. Patroniti N, Foti G, Cortinovis B, Maggioni E, Bigatello LM, Cereda M, Pesenti A: Sigh improves gas exchange and lung volume in patients with acute respiratory distress syndrome undergoing pressure support ventilation. Anesthesiology 2002; 96:788–94

24. Musch G, Harris RS, Vidal Melo MF, O'Neill KR, Layfield JD, Winkler T, Venegas JG: Mechanism by which a sustained inflation can worsen oxygenation in acute lung injury. Anesthesiology 2004; 100:323–30

25. Borges JB, Okamoto VN, Matos GF, Caramez MP, Arantes PR, Barros F, Souza CE, Victorino JA, Kacmarek RM, Barbas CS, Carvalho CR, Amato MB: Reversibility of lung collapse and hypoxemia in early acute respiratory distress syndrome. Am J Respir Crit Care Med 2006; 174:268–78

26. Kuratomi Y, Okazaki N, Ishihara T, Arai T, Kira S: Variability of breath-by-breath tidal volume and its characteristics in normal and diseased subjects. Ventilatory monitoring with electrical impedance pneumography. Jpn J Med 1985; 24:141–9

Cited By:

This article has been cited 11 time(s).

Intensive Care Medicine
Neurally adjusted ventilatory assist: letting the respiratory center take over control of ventilation
de Abreu, MG; Belda, FJ
Intensive Care Medicine, 39(8): 1481-1483.
10.1007/s00134-013-2953-5
CrossRef
Journal of Clinical Monitoring and Computing
NAVA enhances tidal volume and diaphragmatic electro-myographic activity matching: a Range90 analysis of supply and demand
Moorhead, KT; Piquilloud, L; Lambermont, B; Roeseler, J; Chiew, YS; Chase, JG; Revelly, JP; Bialais, E; Tassaux, D; Laterre, PF; Jolliet, P; Sottiaux, T; Desaive, T
Journal of Clinical Monitoring and Computing, 27(1): 61-70.
10.1007/s10877-012-9398-1
CrossRef
Comprehensive Physiology
Ventilatory Failure, Ventilator Support, and Ventilator Weaning
Tobin, MJ; Laghi, F; Jubran, A
Comprehensive Physiology, 2(4): 2871-2921.
10.1002/cphy.c110030
CrossRef
Advances in Biomedical Research, Proceedings
Studies on the natural course of RSV-, HMPV, and RSV/HMPV-double-infections in children in an optimized animal model
Schildgen, V; Lusebrink, J; Ditt, V; Tillmann, R; Simon, A; Muller, A; Schildgen, O
Advances in Biomedical Research, Proceedings, (): 106-121.

Intensive Care Medicine
A novel adaptive control system for noisy pressure-controlled ventilation: a numerical simulation and bench test study
Beda, A; Spieth, PM; Handzsuj, T; Pelosi, P; Carvalho, NC; Koch, E; Koch, T; de Abreu, MG
Intensive Care Medicine, 36(1): 164-168.
10.1007/s00134-009-1665-3
CrossRef
Anesthesiology
Mild Endotoxemia during Mechanical Ventilation Produces Spatially Heterogeneous Pulmonary Neutrophilic Inflammation in Sheep
Costa, EL; Musch, G; Winkler, T; Schroeder, T; Harris, RS; Jones, HA; Venegas, JG; Vidal Melo, MF
Anesthesiology, 112(3): 658-669.
10.1097/ALN.0b013e3181cbd1d4
PDF (1226) | CrossRef
Anesthesiology
Neurally Adjusted Ventilatory Assist Increases Respiratory Variability and Complexity in Acute Respiratory Failure
Schmidt, M; Demoule, A; Cracco, C; Gharbi, A; Fiamma, M; Straus, C; Duguet, A; Gottfried, SB; Similowski, T
Anesthesiology, 112(3): 670-681.
10.1097/ALN.0b013e3181cea375
PDF (1174) | CrossRef
Anesthesiology
A Slightly Different Tune: The Multipart Harmony of Noisy Life Support
Mutch, WA
Anesthesiology, 111(3): 681.
10.1097/ALN.0b013e3181b27c37
PDF (62) | CrossRef
Anesthesiology
Mechanical Ventilation Induces a Toll/Interleukin-1 Receptor Domain-containing Adapter-inducing Interferon β–dependent Inflammatory Response in Healthy Mice
Netea, MG; van der Hoeven, JG; Scheffer, GJ; Vaneker, M; Heunks, LM; Joosten, LA; van Hees, HW; Snijdelaar, DG; Halbertsma, FJ; van Egmond, J
Anesthesiology, 111(4): 836-843.
10.1097/ALN.0b013e3181b76499
PDF (989) | CrossRef
Anesthesiology
Noisy Mechanical Ventilation: Listen to the Melody
Shimabukuro, DW; Gropper, MA
Anesthesiology, 110(2): 214-215.
10.1097/ALN.0b013e3181945b63
PDF (83) | CrossRef
Current Opinion in Critical Care
Technological development in mechanical ventilation
Conti, G; Costa, R
Current Opinion in Critical Care, 16(1): 26-33.
10.1097/MCC.0b013e328334b1e3
PDF (481) | CrossRef
Back to Top | Article Outline

© 2009 American Society of Anesthesiologists, Inc.

Publication of an advertisement in Anesthesiology Online does not constitute endorsement by the American Society of Anesthesiologists, Inc. or Lippincott Williams & Wilkins, Inc. of the product or service being advertised.
Login

Article Tools

Images

Share