Skip Navigation LinksHome > January 2009 - Volume 110 - Issue 1 > Acute Airway Obstruction Caused by the New Single Use Laryng...
Anesthesiology:
doi: 10.1097/ALN.0b013e318190bdc9
Case Reports

Acute Airway Obstruction Caused by the New Single Use Laryngeal Mask Airway Supreme™

Kleine-Brueggeney, Maren M.D.*; Theiler, Lorenz G. M.D.†; Luyet, Cedric M.D.†; Greif, Robert M.D., M.M.E.‡

Free Access
Article Outline
Collapse Box

Author Information

THE Laryngeal Mask Airway Supreme™ (LMA-S™; Laryngeal Mask Company Limited, Henley-on-Thames, United Kingdom) is a single-use supraglottic airway device developed as an alternative to the reusable laryngeal mask airway ProSeal™ (Laryngeal Mask Company). Featuring an additional drain tube to suction gastric content, the LMA-S™ may be used as a backup device in emergency situations. Indeed, its use has been described in difficult emergency intubation1 and in cardiopulmonary resuscitation.2 However, even backup devices have their limitations to consider. We report an unforeseen acute airway obstruction directly caused by the LMA-S™.
Back to Top | Article Outline

Case Report

A 62-yr-old man with a body mass index of 30.2 kg/m2 was scheduled for a resection of a malignant melanoma and sentinel lymph node of the left arm under general anesthesia at the Department of Plastic Surgery at the University Hospital Bern. After preoxigenation of more than 5 min, general anesthesia was induced IV (210 mg of propofol and 0.15 mg of fentanyl) and maintained IV with propofol and remifentanil to keep bispectral index between 40 and 60. Face mask ventilation was successful, and a LMA-S™ (size 5) was introduced easily. The LMA-S™ was cuffed, but ventilation was not possible. This was confirmed by missing end tidal CO2, no visible thorax movement, and no end expiratory tidal volumes. Fiberoptic bronchoscopy showed that the LMA-S™ pushed the epiglottis down, resulting in obstruction of the glottis. Fiberoptic-controlled retraction of the LMA-S™ reversed this situation, and a partial view of the glottic opening was noted, corresponding to a grade of 2 according to a fiberscopic rating suggested by Cook3 and earlier described by Kapila.4 Nevertheless, ventilation was still insufficient.
Although Spo2 continued to be stable between 98 and 99% for the entire time, we decided to remove the LMA-S™ and ventilate the patient by face mask, which was easily done. We then introduced the LMA-S™ a second time. After cuff inflation, only tidal volumes below 300 ml were achieved. Auscultation revealed expiratory and inspiratory stridor over the trachea, and ventilation worsened. Fiberoptic bronchoscopy showed severe narrowing of the laryngeal inlet with barely visible vocal cords, which was interpreted as a supraglottic laryngeal edema.
We removed the LMA-S™ again and intubated the patient (100 mg of succinylcholin) easily (Cormack Lehane grade 1). Surprisingly, no laryngeal swelling was revealed by direct laryngoscopy. Ventilation was sufficient, and surgery was performed without further events. The patient was extubated uneventfully 15 minutes postoperatively and some 2 h and 20 min after induction of anesthesia, and he was discharged to the recovery room without any further respiratory sequels.
Back to Top | Article Outline

Discussion

Laryngeal masks are used broadly for elective and emergency airway management and are an essential part of the American and European difficult airway management algorithm.5,6 Due to their wide use, noticeable complications and side effects have been reported over the last years. The most common side effects are hoarseness and dysphagia. More threatening situations like impossible ventilation with desaturation are much less common. The rare reports of airway obstruction directly triggered by the laryngeal mask are swelling of the pharyngeal soft tissues caused by the leakage of irrigation fluid,7 herniation of the laryngeal mask airway cuff,8–10 foreign bodies (Ascaris lumbricoides11), and intermittent obstruction related to a vagal nerve stimulator.12
Similar to our case, Chin reported a case of increased airway pressure in a ProSeal™ laryngeal mask airway13 and related it to laryngeal edema. However, in a reply by Stix et al., this case was associated with mechanical obstruction of the laryngeal inlet by the cuff and drain tube of the ProSeal™.14
We primarily attributed the rapidly deteriorating ventilation to soft tissue swelling and edema caused by airway manipulation and the mechanical stimulus of the LMA-S™ pushing down the epiglottis. This thesis was not confirmed; direct laryngoscopy after removal of the LMA-S™ showed normal laryngeal anatomy without edema. Laryngeal spasm was also excluded as an underlying cause because a propofol bolus did not improve ventilation.
We therefore attribute this case of airway obstruction to mechanical obstruction of the laryngeal inlet by the cuff as described by Stix.14 The inflated LMA-S™ cuff displaced the cuneiform and corniculate cartilages medially, thereby narrowing the laryngeal inlet as observed during fiberoptic control, obstructing ventilation substantially.
In summary, laryngeal masks are often used as backup devices for the management of possible airway difficulties, and laryngeal masks have been life-saving in daily clinical practice over the years. Supraglottic airway devices are easy to introduce, and they provide a patent airway in most cases. Nevertheless, even newly developed types of laryngeal masks (as the LMA-S™) may be cause for the obstruction of the laryngeal inlet when the mask displaces laryngeal cartilages medially, narrowing the laryngeal inlet. This is the first report of this phenomenon of medialization in a LMA-S™.
Because laryngeal masks are important backup devices in the management of difficult airways, this infrequent cause of a laryngeal mask failure must be made known. Strategies to handle laryngeal mask failures might be included in the algorithms for difficult airway management, and backup strategies for the “backup device” laryngeal mask should be considered.
Back to Top | Article Outline

References

1. Pearson DM, Young PJ: Use of the LMA-Supreme for airway rescue. Anesthesiology 2008; 109:356–7

2. Murdoch H, Cook TM: Effective ventilation during CPR via an LMA-Supreme. Anaesthesia 2008; 63:326

3. Cook TM, Cranshaw J: Randomized crossover comparison of ProSeal Laryngeal Mask Airway with Laryngeal Tube Sonda during anaesthesia with controlled ventilation. Br J Anaesth 2005; 95:261–6

4. Kapila A, Addy EV, Verghese C, Brain AI: The intubating laryngeal mask airway: an initial assessment of performance. Br J Anaesth 1997; 79:710–3

5. Practice guidelines for management of the difficult airway: an updated report by the American Society of Anesthesiologists Task Force on Management of the Difficult Airway. Anesthesiology 2003; 98:1269–77

6. Henderson JJ, Popat MT, Latto IP, Pearce AC: Difficult Airway Society guidelines for management of the unanticipated difficult intubation. Anaesthesia 2004; 59:675–94

7. Yoshimura E, Yano T, Ichinose K, Ushijima K: Airway obstruction involving a laryngeal mask airway during arthroscopic shoulder surgery. J Anesth 2005; 19:325–7

8. Solaidhanasekaran S, Bharamgoudar M: Airway obstruction secondary to herniation of the paediatric laryngeal mask airway. Anaesthesia 2008; 63:785–6

9. Christelis ND, Doolan GK: Complete airway obstruction caused by cuff herniation of an overused laryngeal mask airway. Anaesth Intensive Care 2008; 36:274–5

10. Wrobel M, Ziegeler S, Grundmann U: Airway obstruction due to cuff herniation of a classic reusable laryngeal mask airway. Anesthesiology 2007; 106:1255–6

11. Roy K, Kundra P, Ravishankar M: Unusual foreign body airway obstruction after laryngeal mask airway insertion. Anesth Analg 2005; 101:294–5

12. Bernards CM: An unusual cause of airway obstruction during general anesthesia with a laryngeal mask airway. Anesthesiology 2004; 100:1017–8

13. Chin KJ, Chee VW: Laryngeal edema associated with the ProSeal laryngeal mask airway in upper respiratory tract infection. Can J Anaesth 2006; 53:389–92

14. Stix MS, O’Connor CJ, Jr., Valade DR: The ProSeal LMA does not cause laryngeal edema. Can J Anaesth 2006; 53:961–3

Cited By:

This article has been cited 2 time(s).

Anesthesiology
Detecting the Etiologies of Acute Airway Obstruction Associated with the Laryngeal Mask Airway Supreme™
Osborn, IP; Behringer, EC; Cooper, RM; Verghese, C
Anesthesiology, 111(2): 451-452.
10.1097/ALN.0b013e3181adf285
PDF (86) | CrossRef
Anesthesiology
Detecting the Etiologies of Acute Airway Obstruction Associated with the Laryngeal Mask Airway Supreme™
Theiler, LG; Kleine-Brueggeney, M; Greif, R
Anesthesiology, 111(2): 452-453.
10.1097/ALN.0b013e3181adffb3
PDF (85) | CrossRef
Back to Top | Article Outline

© 2009 American Society of Anesthesiologists, Inc.

Publication of an advertisement in Anesthesiology Online does not constitute endorsement by the American Society of Anesthesiologists, Inc. or Lippincott Williams & Wilkins, Inc. of the product or service being advertised.
Login

Article Tools

Share