Anesthesiology:
Clinical investigations

New Composite Index Based on Midlatency Auditory Evoked Potential and Electroencephalographic Parameters to Optimize Correlation with Propofol Effect Site Concentration: Comparison with Bispectral Index and Solitary Used Fast Extracting Auditory Evoked Potential Index

Vereecke, Hugo E. M. M.D.*; Vasquez, Pablo Martinez M.Sc.†; Jensen, Erik Weber M.Sc., Ph.D.†; Thas, Olivier M.Sc., Ph.D.‡; Vandenbroecke, Rudy M.D., Ph.D.§; Mortier, Eric P. M.D., D.Sc.∥; Struys, Michel M. R. F. M.D., Ph.D.#

Free Access
Article Outline
Collapse Box

Author Information

Collapse Box

Abstract

Background: This study investigates the accuracy of a composite index, the A-Line® auditory evoked potentials index version 1.6 (AAI1.6; Danmeter A/S, Odense, Denmark), as a measure of cerebral anesthetic drug effect in a model for predicting a calculated effect site concentration of propofol (CePROP). The AAI1.6 algorithm extracts information from the midlatency auditory evoked potentials, the spontaneous electroencephalographic activity, and the detection of burst suppression. The former version of this monitor, the A-Line® auditory evoked potential index version 1.5, is only based on fast extracted midlatency auditory evoked potential information.
Methods: After institutional ethics committee approval (University Hospital, Ghent, Belgium), informed consent was obtained from 13 patients (10 women, 3 men) with an American Society of Anesthesiologists physical status of I, aged 18–65 yr, who were scheduled to undergo ambulatory gynecologic or urologic surgery. The authors evaluated for Bispectral Index, A-Line® auditory evoked potential index, version 1.5, AAI1.6 scaled from 0 to 100 and AAI1.6 scaled from 0 to 60, the interpatient stability at baseline, the detection of burst suppression, prediction probability, and correlation with CePROP, during a constant infusion of 1% propofol at 300 ml/h. The authors developed pharmacodynamic models relating the predicted CePROP to each measure of cerebral anesthetic drug effect.
Results: Bispectral Index had the lowest interindividual baseline variability. No significant difference was found with prediction probability analysis for all measures. Comparisons for correlation were performed for all indices. The AAI1.6 scaled to 60 had a significantly higher correlation with CePROP compared with all other measures. The AAI1.6 scaled to 100 had a significant higher correlation with CePROP compared with the A-Line® auditory evoked potential index version 1.5 (P < 0.05)
Conclusions: The authors found that the application of AAI1.6 has a better correlation with a calculated CePROP compared with a solitary fast extracting midlatency auditory evoked potential measure. Whether this improvement in pharmacodynamic tracing is accompanied by an improved clinical performance should be investigated using clinical endpoints.
PROCESSED analysis of the electroencephalogram or midlatency auditory evoked potentials (MLAEPs) is increasingly accepted as surrogate endpoint for quantification of anesthetic cerebral drug effect. For the electroencephalogram, the Bispectral Index (BIS) incorporated in the A-2000 BIS® monitor (Aspect Medical Systems, Newton, MA) has been proven to have a high sensitivity and specificity to measure anesthetic drug effect.1–3 For MLAEPs, Jensen et al.4 developed a new method for extracting the MLAEP from the electroencephalogram by using an autoregressive model with an exogenous input (ARX), allowing fast extraction of the raw MLAEP signal. A new monitoring variable called the A-Line® ARX Index, version 1.5 (AAI1.5), is incorporated in a recently commercialized monitor called the A-Line® (Danmeter A/S, Odense, Denmark). Various investigators have illustrated its clinical usefulness and limitations.2,3,5,6
In previous work, we compared the performance of both systems and found that the correlation between the changes in anesthetic drug effect as measured by BIS or AAI1.5 and the propofol effect site concentration (CePROP) were accurate, except at baseline and during excessive levels of anesthetic drug effect, as detected by the increasing level of burst suppression.7 For BIS, better baseline stability (minimal variability in the absence of drugs between individuals) was observed as compared with AAI1.5.7 Excessive levels of anesthetic drug effect remained detectable by the BIS monitor from a burst suppression ratio of 40%, whereby no changes in AAI1.5 were observed during increasing levels of burst suppression.7 This lack of information is caused by the fact that (1) burst suppression ratio does not participate in the calculation of AAI1.5 and (2) at patient-specific levels of anesthetic drug effect, the amplitude (latency) of the raw MLAEP wave is reduced (increased) in such a way that it becomes nearly a straight line, limiting its discriminating power.8,9
In a recent editorial in Anesthesiology, Kalkman and Drummond10 stated that it might be proven ultimately, if we deem the grail of depth of anesthesia monitoring worth pursuing, that the optimal monitor of depth of anesthesia will be one that integrates parameters extracted from both spontaneous and evoked cerebral electrophysiologic signals. Following this philosophy, the first step is to prove that this monitor has a good correlation with cerebral anesthetic drug effect during the administration of a hypnotic drug.10 Therefore, we hypothesized that the construction of a composite index calculated from the combination of a fast extracted MLAEP, electroencephalogram, and burst suppression might offer a broader range of information on the calculated effect site of propofol—as a pharmacodynamic endpoint of cerebral anesthetic drug effect—compared with an index based on solitary MLAEP or electroencephalographic input. Recently, an upgrade version of the A-Line® has been developed, producing such a composite index, called the A-Line® auditory evoked potential index, version 1.6 (AAI1.6; Danmeter).
This study compares the accuracy of the BIS, AAI1.5, and AAI1.6 in their ability to predict a progressively increasing CePROP. Therefore, we evaluated for all measures the stability at baseline (minimal variability in the absence of drug between individuals),11 accurate detection of burst suppression,12 prediction probability,13,14 and correlation with the CePROP. We also developed pharmacodynamic models relating the predicted CePROP to each measure of cerebral anesthetic drug effect.
Back to Top | Article Outline

Materials and Methods

Clinical Protocol
After institutional ethics committee approval (University Hospital, Ghent, Belgium), informed consent was obtained from 13 patients (10 women, 3 men) with an American Society of Anesthesiologists physical status of I, aged 18–65 yr, who were scheduled to undergo ambulatory gynecologic or urologic surgery. Exclusion criteria included weight less than 70% or more than 130% of ideal body weight, neurologic disorder, and recent use of psychoactive medication, including alcohol.
An 18-gauge intravenous line was positioned at a large left forearm vein. Every patient received approximately 100 ml crystalloid fluid during the study period. No fluid load was given before induction. No patient received preoperative medication. No other drugs, including opioids, were given during the study period. All patients maintained spontaneously ventilating (even at high levels of burst suppression) via a facemask delivering 100% oxygen at 6 l/min.
Heart rate and noninvasive blood pressure, arterial oxygen saturation, and capnography were recorded at 1-min intervals using an S-5 monitor (Datex, Helsinki, Finland). Capnography was monitored by putting the side stream sample line in the facemask of the patient. This implies the occurrence of an error for quantification of the partial pressure of carbon dioxide, but it enables monitoring of respiratory rate and free airway. A silent operation room was obtained for all patients.
At the start of the study period, 1% propofol was administered at 300 ml/h via a computer-assisted continuous infusion device (RUGLOOP**). This system captures all monitored data while driving a Fresenius Modular Infusion Pump connected to a Fresenius Base A (Fresenius Vial Infusion Systems, Brézins, France) via an RS-232 interface. During the propofol infusion, the corresponding CePROP is calculated in a time-synchronized way by RUGLOOP using a three-compartment model enlarged with an effect site compartment, previously published by Schnider et al.15,16 The calculated CePROP was computed to yield a time to peak effect of 1.6 min after bolus injection,17 as also published by Schnider et al.15,16 and clinically confirmed by Struys et al. 18 The propofol infusion was continued at the same speed until burst suppression levels of 80% or higher were achieved. However, infusion was stopped earlier if the mean arterial blood pressure became lower than 50 mmHg.
Before the drug administration was started, all patients were asked to close their eyes and relax for 2 min. During that time, signal quality, impedance of the electrodes, and the adequate detection of all parameters by RUGLOOP were verified. Baseline measures were performed during the first 5 s after starting the pumps, when CePROP was 0 in all patients. Averaging of the data was performed using a 10-s interval.
Back to Top | Article Outline
Electroencephalographic and MLAEP Data Collection
BIS-XP (version 4.0) was derived from the frontal electroencephalogram (At-Fpzt) and calculated by the A-2000 BIS® monitor using four BIS® sensor electrodes (Aspect Medical Systems, Inc.). The BIS value ranges from 0 to 100. The smoothening time of the BIS monitor was set at 15 s.
We recorded raw electroencephalographic data including the MLAEP data with a prototype version of the A-Line® monitor. Three electrodes (A-Line® AEP electrodes; Danmeter A/S) were positioned at mid forehead (+), left forehead (reference), and left mastoid (−). Impedance was always lower than 5 kOhm. In this protocol, MLAEP was elicited by a bilateral click at 9 Hz, with a 2-ms duration and an adaptable click intensity set automatically by the monitor according to the measured signal-to-noise ratio (SNR) of the raw MLAEP signal.19 Artifact rejection and 25- to 65-Hz band-pass filtering was conducted previously to the extraction of the MLAEP.
All raw MLAEP data were stored on a compact flash card (SanDisk, Sunnyvale, CA) connected to the prototype A-Line®. Post hoc, both AAI1.5 and AAI1.6 were extracted from these data on a time synchronized basis.
The AAI1.5 is calculated using a fast extracting method called ARX, enabling us to extract information from the MLAEP within 2–6 s. This protocol has been published elsewhere.2 The AAI1.5 value ranges from 0 to 100.
Equation (Uncited)
Equation (Uncited)
Image Tools
The AAI1.6 is a composite index using three sources of information resulting in an index value ranging between 0 and 100. First, an ARX is used to extract information from the raw MLAEP wave during periods with high SNR. The method used is based on the AAI1.5 protocol and is described elsewhere.4 When the MLAEP is reduced to a flat line, the SNR becomes too low to extract a useful AAI calculation. At that time, the electroencephalographic-based information will determine the AAI1.6 calculation. Finally, when burst suppression patterns are detected, this information becomes the major factor in the calculation of the composite index. The AAI1.6 is mathematically formulated as
where AAI1.5 is the result of the ARX model. E30–47 Hz and E10–20 Hz are the results of a power analysis in the raw electroencephalographic spectrum using higher and lower frequencies, respectively. BS is the percentage of burst suppression patterns detected during the last 30 s. k0, k1, and k2 are functions of the SNR, the detection of iatrogenic artifacts (translated in a signal quality index), and the auditory stimulus intensity. The SNR is determined during the averaging process of the raw signal. It evaluates the detection quality of the signal under investigation. For MLAEP, the occurrence of the electroencephalogram, electromyogram, and other artifacts will induce a lower SNR. For the electroencephalogram, only electromyographic and iatrogenic artifacts will be taken into account. The signal quality index is calculated based on the number (and type) of artifacts during a certain period of time. If the amount of artifacts per period of time is higher than a percentage of this time, no AAI1.6 is calculated because it is not reliable. If the number of artifacts per period of time is less than this value, an index is calculated, but the signal quality index bar is decreased on screen, indicating a lower accuracy of the calculated index. When this occurs, an artifact code is transferred to the RUGLOOP program automatically. These data were deleted post hoc from further analysis.
The burst suppression parameter is defined as the percentage of time where the power of the electroencephalogram is smaller than 5 μV over a period of time. However, the calculation of burst suppression differs between monitors. For BIS and AAI1.5, the burst suppression algorithm details are published previously.7
In the AAI1.6, the burst suppression detection algorithm is based on a maximum likelihood cumsum algorithm, with variable probability functions, using a preprocessed signal after artifact detection and filtering. The main property of this method is that low-amplitude signal criteria are based on probabilities and not on a fixed number. Second, the signal has no fixed segmentation, allowing a higher time resolution for detecting changes in comparison with other methods as applied in AAI1.5 and BIS. Finally, burst suppression is evaluated over a window of 30 s in AAI1.6, instead of 22 s in the former version.
Because it might be hypothesized that no additional information on loss of consciousness is revealed for AAI1.6 values between 60 and 100,3 the A-Line® monitor offers the possibility to adjust the range of the index toward a 0–60 scale (defined as AAI1.660). However, by setting all values higher than 60 to 60, it has to be proven that no information on cerebral anesthetic drug effect is lost, which is crucial when using the monitor in a pharmacodynamic protocol or in a clinical setting. Therefore, applicable performance measurements (baseline variability, individualized Spearman rank correlation, prediction probability analysis) were calculated on both scales.
Back to Top | Article Outline
Performance Measures
The significance level was set at 5% unless otherwise reported.
Back to Top | Article Outline
Baseline Variability.
The baseline variability is calculated by computing the coefficient of variance on the electroencephalographic data points obtained during the first 5 s of the protocol, before any drug has been delivered. In this setting, the baseline variability will mainly reflect interpatient variability because of the short duration of measurement.
Back to Top | Article Outline
Comparison between AAI1.5 and AAI1.6.
A Pearson correlation was calculated between AAI1.5 and AAI1.6.
Back to Top | Article Outline
Burst Suppression Detection and Parameter Correlation.
The relations between the burst suppression and each measure of anesthetic drug effect are plotted. A Spearman rank correlation was calculated. For each scatterplot, a model was fitted to the data using curve estimation function from Sigmaplot2000® for Windows® (SPSS Inc., Chicago, IL). The curve estimation function produces curve estimation regression statistics and related plots for different curve estimation regression models including linear, logarithmic, inverse, quadratic, cubic, power, compound, S-curve, logistic, growth, and exponential. A separate model is produced for each dependent variable, together with its regression coefficients, predicted values, residuals, and predictive intervals. After this, the most appropriate regression model can be selected.
Back to Top | Article Outline
Prediction Probability.
The ability of the different indicators to describe the anesthetic drug effect was evaluated using prediction probability (PK), which compares the performance of indicators having different units of measurements. PK was calculated using a custom spreadsheet macro PKMACRO, developed by Smith et al.13,14 The PK value was calculated for every individual patient and for all parameters studied. We evaluated the predictive capacity of BIS, AAI1.5, and AAI1.6 for detecting the calculated CePROP, which has a known correlation with the anesthetic drug effect.18 The calculated CePROP reached at every 5 s was used as an endpoint for PK calculation. CePROP was calculated with a precision of two decimals. A PK of 1 for BIS, AAI1.5, or AAI1.6 would mean that BIS, AAI1.5, or AAI1.6 always decreases (increases) as the patient reaches higher (lower) drug concentrations according to the CePROP. Such an indicator can perfectly predict the anesthetic drug concentration. Alternatively, a PK value of 0.5 would mean that the indicator is useless for predicting anesthetic drug concentration. The jackknife method was used to compute the SE of the estimate, based on the assumption that all assessments were independent.13,14 After having evaluated normal distribution, a Student t test with Bonferroni correction was used to evaluate significant difference between PK means.
Back to Top | Article Outline
Individualized Spearman Rank Correlation.
As an additional nonparametric approach, the Spearman rank correlations between CePROP and BIS, AAI1.5, or AAI1.6 (both scales) were individualized by computing the correlation first for every patient i separately, defined as Ri. The reported Spearman rank correlation, R, is a weighted average of the Ri (weighted according to the number of observations for each patient). In this way, R retained its usual interpretation. The confidence intervals on R were obtained by the bootstrap method in which the hierarchical nature of the data were incorporated by resampling within patients. Equality of two correlation coefficients was tested at the 5% level of significance by constructing the 95% confidence intervals of the difference (confidence intervals were also computed with the bootstrap technique). All bootstrap calculations were based on 10,000 simulation runs.
Back to Top | Article Outline
Pharmacodynamic Modeling
Equation (Uncited)
Equation (Uncited)
Image Tools
The relation between CePROP and the electroencephalographic measures of anesthetic drug effect were analyzed using a sigmoid Emax model,
Equation (Uncited)
Equation (Uncited)
Image Tools
where Effect is the electroencephalographic effect being measured (BIS, AAI1.5, or AAI1.6 [both scales]), E0 is the baseline measurement when no drug is present, Emax is the maximum possible drug effect, Ce is the calculated effect site concentration of propofol, Ce50 is the effect site concentration associated with 50% maximal drug effect, and γ is the steepness of the concentration–response relation curve. The model parameters were estimated using NONMEM V (Globomax LLC, Hanover, MD). Interindividual variability was modeled using a log-normal distribution,
where Pi is the parameter value (E0, Emax, y or Ce50) in the ith patient, Ptv is the typical value of the parameter in the population, and η is a random variable with a mean of 0 and a variance of ω.2 Individual variability is reported as ω, the SD of η in the log domain, which is approximately the coefficient of variance in the standard domain. Residual intraindividual variability was modeled using a standard additive error model.20
Back to Top | Article Outline

Results

The population characteristics were as follows: weight, 64.0 ± 3.8 kg; age, 38 ± 6.0; height, 168 ± 7.8 cm; sex, 10 women and 3 men. All patients remained within a safe hemodynamic and respiratory clinical state.
Fig. 1
Fig. 1
Image Tools
All measured data were included in the analysis. Figure 1 shows the raw data over time for the four electroencephalographic measures (BIS, AAI1.5, AAI1.6, AAI1.660) and CePROP. All patients remained in a hemodynamic and respiratory safe condition during the study period.
Back to Top | Article Outline
Performance Measures
Table 1
Table 1
Image Tools
The interpatient baseline variability was calculated for all studied electroencephalographic measures (BIS, AAI1.5, AAI1.6, AAI1.660). The coefficient of variance on the first 5 s of measurement before administration of propofol is shown in table 1.
Fig. 2
Fig. 2
Image Tools
AAI1.5 and AAI1.6 (both scaled between 0 and 100) were compared as shown in figure 2. This shows a strong correlation between both AAI versions (Pearson correlation R = 0.91209833, P < 0.001). We notice a large variability at high values of both parameters and a tendency for AAI1.5 to remain on a higher value at higher levels of CePROP in comparison with AAI1.6. For AAI1.5, no data lower than 7 are observed in contrast to AAI1.6.
Fig. 3
Fig. 3
Image Tools
The relations between the percentage of burst suppression and BIS, AAI1.5, and AAI1.6 are plotted in figure 3 together with the most appropriate curve estimation. For BIS, a cubic polynomial regression curve was considered to be the best fit. For AAI1.5 and AAI1.6, a sigmoid regression curve with three parameters was selected as the best fit. For AAI1.660, identical results to AAI1.6 were found. Therefore, no graph is presented. The Spearman rank correlations between burst suppression and BIS, AAI1.5, and AAI1.6 are −0.728, −0.551, and −0.871, respectively.
Table 2
Table 2
Image Tools
The PK values for BIS, AAI1.5, AAI1.6, and AAI1.660 are shown in table 2. The individualized Spearman rank correlations between CePROP and BIS, AAI1.5, AAI1.6, and AAI1.660 are also included in table 2.
Back to Top | Article Outline
Pharmacodynamic Modeling
Fig. 4
Fig. 4
Image Tools
Fig. 5
Fig. 5
Image Tools
Table 3
Table 3
Image Tools
Figure 4 shows the behavior of the four electroencephalographic measures versus CePROP for all patients. With increasing CePROP, all measures decreased. The relations of CePROP to BIS, AAI1.5, AAI1.6, and AAI1.660 are modeled and shown in figure 5. The typical parameter values for each population model including the coefficient of variance (as a measure for interindividual variability in the standard domain) are found in table 3. The SD for each model (as a measure of the intraindividual variability in the log domain) was 5.63 for BIS, 13.67 for AAI1.5, 7.24 for AAI1.6, and 4.02 for AAI1.660.
Back to Top | Article Outline

Discussion

This study shows that the application of a composite index (AAI1.6) based on MLAEPs and spontaneous electroencephalography optimizes the prediction of CePROP as compared with a solitary fast extracting MLAEP measure (AAI1.5). Comparing several monitors for cerebral hypnotic drug effect demands a systematic approach. The methodology used in this trial is based on a former publication from Vanluchene et al.,7 where spectral entropy is compared with BIS and AAI1.5.
The A-Line® monitor offers the possibility to reduce the upper scale limit to 60, based on the rationale that values between 60 and 100 suffer from a high interpatient variability and offer no additional information on loss of consciousness.2 Although this adjustment might be interesting on a clinical level, it is crucial to prove whether this scale is applicable to pharmacodynamic studies investigating subhypnotic levels of anesthesia (when consciousness is not lost), without losing any information on the propofol drug concentration. Therefore, we also investigated the performance accuracy of the AAI1.660 as an individual index.
High baseline variation, as defined by the coefficient of variance, might decrease the predictive ability of the electroencephalogram-derived measures when used to detect cerebral drug effect, as stated by Bruhn et al. 11 BIS showed the lowest baseline variability. In contrast, AAI1.5 and AAI1.6 showed a high but comparable coefficient of variance. We found a major improvement in baseline variability in AAI1.660 in comparison to AAI1.5 and AAI1.6. Although it is potentially interesting when applying the A-Line® monitor clinically to monitor loss of consciousness, one might question that it does not make sense to reduce baseline variability by cutting of data higher than 60, if this new scale causes a decreased sensitivity for the detection of subhypnotic levels of anesthetic drug effect. This reduced upper limit only makes sense if the other performance measures reveal a beneficial effect.
Burst suppression is a benign electroencephalographic pattern frequently seen in healthy brain at deep levels of anesthesia. Because the detection of burst suppression represents an important component to the level of hypnotic effect, it might be helpful to take into account burst suppression analysis when measuring deep levels of anesthetic drug effect because it might optimize the discriminating power of the investigated index at these levels. For BIS, burst suppression is included in the algorithm. Various authors have found that above a burst suppression ratio of 40%, a linear correlation exists between BIS and burst suppression ratio, indicating that the burst suppression ratio is the only determinant factor for BIS values below 30.7,21 Our results for BIS, as shown in figure 3, are in agreement with these previously published findings. If burst suppression was excluded from the algorithm, BIS would become very resistant at excessive levels of anesthetic drug effect. In contrast, the calculation of the AAI1.5 is only based on the MLAEP. As already observed in previous work, the MLAEP flattens at patient-specific levels of anesthetic drug effect, thereby limiting its discriminating power.7–9 This phenomenon results in a maximum decreased AAI1.5 “plateau” level higher than 0. As shown in figures 2 and 4, the lowest observed AAI1.5 values are ceiled around 7. Beyond this level, no additional information on cerebral drug effect can be obtained. When taking in consideration that the ED95 for loss of consciousness is around 19 for AAI1.5, the discriminating power might become very low after loss of consciousness. For AAI1.6, the electroencephalogram and burst suppression are included as a component of the composite index as described in the Materials and Methods section. This results in an increased discriminating power at deep levels of anesthetic drug effect. As shown in figure 3, the absolute values of AAI1.6 are able to decrease to 0 at increasing levels of burst suppression. This improvement between AAI1.5 and AAI1.6 is also reflected in the statistically significant better individualized Spearman rank correlation for AAI1.6 (table 2).
When studying performance accuracy, the question is how well the observed measure, which is the electroencephalographic response, predicts the unobserved “underlying” state of the patient, which is represented by the CePROP. Therefore, PK was calculated for BIS, AAI1.5, AAI1.6, and AAI1.660. The underlying theoretical background of PK calculations has been published elsewhere.7 In our study, we did not obtain any significant difference between PK values, although AAI1.6 tended to have the highest median value.
As an additional analysis for correlation between parameters and CePROP, we calculated the Individualized Spearman rank correlation coefficients for each measure. This weighted nonparametric statistical approach depicts the nonlinearity in the system much more accurately as compared with PK analyses. Results confirm the significant increase in correlation between AAI1.6 versus CePROP in comparison to AAI1.5, thereby proving that the combination of MLAEP and derived measures from the spontaneous electroencephalogram (power spectrum analysis and burst suppression calculation) increases the accuracy to measure cerebral anesthetic drug effect. Initially, no difference was found between BIS and AAI1.5 or between BIS and AAI1.6 when scaled to 100. However, when the upper limit of the scale was reduced to 60, remarkable results were observed. Figure 4 shows that a high variability and a high amount of nonlinearity are observed in all measures when scaled to 100. For AAI1.6, one can say that the signal between 60 and 100 nearly behaves as “random noise.” By eliminating this part of the AAI1.6 range, the high variability resulting in a high level of nonlinearity is drastically reduced. This results in a very high individualized Spearman rank correlation as shown in table 2. However, one can raise the question of whether a constant number around 60 offers more information compared with the complete scale, both when predicting propofol anesthetic drug effect and when evaluating a clinical level of sedation before loss of consciousness is obtained. This must be studied in further research. We were not able to evaluate the clinical level of sedation in this population because we wanted to avoid any stimulation during induction.
Figure 5 shows the pharmacodynamic modeling for all measures versus CePROP. Previously, the relation between measures of anesthetic drug effect and CePROP was observed following a sigmoid Emax model. Therefore, in our study, the relation between the measures of anesthetic drug effect and CePROP was also modeled using a sigmoid Emax model, and the model parameters were estimated using a population approach. In NONMEM, the parameters in the individual are weighted in a Bayesian manner toward the mean for the population, based on the variance of the individual parameters. Examining our raw data, we found for all parameters a single sigmoid regression to be adequate for describing the range of concentrations used in this study. When looking at figure 5, one can conclude that the smoothest regression is found in AAI1.660. This is reflected in an optimization of both the interindividual and the intraindividual variability in the population. In table 3, the coefficients of variation of the typical values of the cerebral drug effect measures reveal a major decrease in the interindividual variability of the slope (γ) when AAI1.660 is compared to AAI1.6. Simultaneously, a decrease in intraindividual variability at baseline is observed for AAI1.660, reflected by the lower SD compared with AAI1.6. However, we want to stress the fact that the regression analysis for AAI1.660 should be interpreted cautiously because of the effect of rejection of data on the NONMEM analysis. It might cause a distortion on the data input in such a way that the behavior of the population regression is closer to the a priori chosen Emax model. Further research is needed to clarify whether this result is an artificial overestimation of the performance parameters or whether AAI1.660 is indeed able to maintain a high performance as a hypnotic drug effect monitor.
We conclude that the use of the composite index AAI1.6, which combines information from the MLAEP, spontaneous electroencephalogram, and burst suppression, increases the correlation with the cerebral drug effect of propofol as compared with the AAI1.5, which is a solitary fast extracted MLAEP index. By adjusting the upper scale limit to 60, the performance measures are optimized even more. However, the implications of this newly chosen upper scale limit should be further explored on both the clinical and the pharmacodynamic level.
Back to Top | Article Outline

References

1. Sleigh JW, Donovan J: Comparison of bispectral index, 95% spectral edge frequency and approximate entropy of the electroencephalogram,with changes in heart rate variability during induction of general anaesthesia. BrJ Anaesth 1999; 82:666–71

2. Struys MM, Jensen EW, Smith W, Smith NT, Rampil I, Dumortier FJ, Mestach C, Mortier EP: Performance of the ARX-derived auditory evoked potential index as an indicator of anesthetic depth: A comparison with Bispectral Index and hemodynamic measures during propofol administration. Anesthesiology 2002; 96: 803–16

3. Struys MM, Vereecke H, Moerman A, Jensen EW, Verhaeghen D, De Neve N, Dumortier FJ, Mortier EP: Ability of the Bispectral Index, autoregressive modelling with exogenous input-derived auditory evoked potentials, and predicted propofol concentrations to measure patient responsiveness during anesthesia with propofol and remifentanil. Anesthesiology 2003; 99:802–12

4. Jensen EW, Lindholm P, Henneberg SW: Autoregressive modeling with exogenous input of middle-latency auditory-evoked potentials to measure rapid changes in depth of anesthesia. Methods Inf Med 1996; 35:256–60

5. Litvan H, Jensen EW, Galan J, Lund J, Rodriguez BE, Henneberg SW, Caminal P, Villar Landeira JM: Comparison of conventional averaged and rapid averaged, autoregressive-based extracted auditory evoked potentials for monitoring the hypnotic level during propofol induction. Anesthesiology 2002; 97:351–8

6. Litvan H, Jensen EW, Revuelta M, Henneberg SW, Paniagua P, Campos JM, Martinez P, Caminal P, Villar Landeira JM: Comparison of auditory evoked potentials and the A-line ARX Index for monitoring the hypnotic level during sevoflurane and propofol induction. Acta Anaesthesiol Scand 2002; 46:245–51

7. Vanluchene ALG, Vereecke HEM, Thas O, Mortier EP, Shafer SL, Struys MMRF: Spectral entropy as an electroencephalographic measure of anesthetic drug effect: A comparison with Bispectral Index and processed midlatency auditory evoked response. Anesthesiology 2004; 101:34–42

8. Thornton C, Konieczko KM, Knight AB, Kaul B, Jones JG, Dore CJ, White DC: Effect of propofol on the auditory evoked response and oesophageal contractility. Br J Anaesth 1989; 63:411–7

9. Schwender D, Madler C, Klasing S, Poppel E, Peter K: Monitoring intraoperative processing of acoustic stimuli with auditory evoked potentials [in German]. Infusionsther Transfusionsmed 1993; 20:272–6

10. Kalkman CJ, Drummond JC: Monitors of depth of anesthesia, quo vadis? Anesthesiology 2002; 96:784–7

11. Bruhn J, Bouillon TW, Hoeft A, Shafer SL: Artifact robustness, inter- and intraindividual baseline stability, and rational electroencephalogram parameter selection. Anesthesiology 2002; 96:54–9

12. Bruhn J, Bouillon TW, Shafer SL: Onset of propofol-induced burst suppression may be correctly detected as deepening of anaesthesia by approximate entropy but not by bispectral index. Br J Anaesth 2001; 87:505–7

13. Smith WD, Dutton RC, Smith NT: A measure of association for assessing prediction accuracy that is a generalization of non-parametric ROC area. Stat Med 1996; 15:1199–215

14. Smith WD, Dutton RC, Smith NT: Measuring the performance of anesthetic depth indicators. Anesthesiology 1996; 84:38–51

15. Schnider TW, Minto CF, Gambus PL, Andresen C, Goodale DB, Shafer SL, Youngs EJ: The influence of method of administration and covariates on the pharmacokinetics of propofol in adult volunteers. Anesthesiology 1998; 88:1170–82

16. Schnider TW, Minto CF, Shafer SL, Gambus PL, Andresen C, Goodale DB, Youngs EJ: The influence of age on propofol pharmacodynamics. Anesthesiology 1999; 90:1502–16

17. Shafer SL, Gregg KM: Algorithms to rapidly achieve and maintain stable drug concentrations at the site of drug effect with a computer-controlled infusion pump. J Pharmacokinet Biopharm 1992; 20:147–69

18. Struys MM, De Smet T, Depoorter B, Versichelen LF, Mortier EP, Dumortier FJ, Shafer SL, Rolly G: Comparison of plasma compartment versus two methods for effect compartment–controlled target-controlled infusion for propofol. Anesthesiology 2000; 92:399–406

19. Picton TW, HS, Krausz HI, Galambos R: Human auditory evoked potentials: I. Evaluation of components. Electroenc Clin Neurophysiology 1974; 36:179–90

20. Beal S, Sheiner L: NONMEM User’s Guide. San Francisco, University of California, 1992

21. Bruhn J: Burst suppression ratio is the only determinant for BIS values below 30. Can J Anaesth 2002; 49: 755–6

** RUGLOOP program. Available at: http://www.anesthesia-uzgent.be. Accessed April 11, 2005. Cited Here...

Cited By:

This article has been cited 19 time(s).

Acta Anaesthesiologica Scandinavica
Prospective evaluation of the time to peak effect of propofol to target the effect site in children
Munoz, HR; Leon, PJ; Fuentes, RS; Echevarria, GC; Cortinez, LI
Acta Anaesthesiologica Scandinavica, 53(7): 883-890.
10.1111/j.1399-6576.2009.01995.x
CrossRef
Pediatric Anesthesia
AEP-monitor/2 derived, composite auditory evoked potential index (AAI-1.6) and bispectral index as predictors of sevoflurane concentration in children
Ironfield, CM; Davidson, AJ
Pediatric Anesthesia, 17(5): 452-459.
10.1111/j.1460-9592.2006.02155.x
CrossRef
Anesthesia and Analgesia
Variability comparison of the composite auditory evoked potential index and the bispectral index during propofol-fentanyl anesthesia
Rehberg, B; Ryll, C; Hadzidiakos, D; von Dincklage, F; Baars, JH
Anesthesia and Analgesia, 107(1): 117-124.
10.1213/ane.0b013e31816f1965
CrossRef
British Journal of Anaesthesia
Nervus medianus evoked potentials and bispectral index during repeated transitions from consciousness to unconsciousness
Rundshagen, I; Mast, J; Mueller, N; Pragst, F; Spies, C; Cortina, K
British Journal of Anaesthesia, 101(3): 366-373.
10.1093/bja/aen186
CrossRef
Anesthesia and Analgesia
A Comparison in Adolescents of Composite Auditory Evoked Potential Index and Bispectral Index During Propofol-Remifentanil Anesthesia for Scollosis Surgery with Intraoperative Wake-Up Test
van Oud-Alblas, HJB; Peters, JWB; de Leeuw, TG; Vermeylen, KTA; de Klerk, LWL; Tibboel, D; Klein, J; Weber, F
Anesthesia and Analgesia, 107(5): 1683-1688.
10.1213/ane.0b013e3181852d94
CrossRef
Anesthesia and Analgesia
Estimation of optimal modeling weights for a Bayesian-based closed-loop system for propofol administration using the bispectral index as a controlled variable: A simulation study
De Smet, T; Struys, MMRF; Greenwald, S; Mortier, EP; Shafer, SL
Anesthesia and Analgesia, 105(6): 1629-1638.
10.1213/01.ane.0000287269.06170.Of
CrossRef
British Journal of Anaesthesia
Depth of anaesthesia monitoring: what's available, what's validated and what's next?
Bruhn, J; Myles, PS; Sneyd, R; Struys, MMRF
British Journal of Anaesthesia, 97(1): 85-94.
10.1093/bja/ael120
CrossRef
British Journal of Anaesthesia
Changes in a surgical stress index in response to standardized pain stimuli during propofol-remifentanil infusion
Struys, MMRF; Vanpeteghem, C; Huiku, M; Uutela, K; Blyaert, NBK; Mortier, EP
British Journal of Anaesthesia, 99(3): 359-367.
10.1093/bja/aem173
CrossRef
Veterinary Journal
Effects of sedative and hypnotic drug combinations on transcranial magnetic motor evoked potential, bispectral index and ARX-derived auditory evoked potential index in dogs
Van Soens, I; Struys, MM; Polis, IE; Tshamala, M; Nollet, H; Bhatti, SF; Van Ham, LM
Veterinary Journal, 181(2): 163-170.
10.1016/j.tvjl.2008.02.014
CrossRef
Anesthesia and Analgesia
Performance of the cerebral state index during increasing levels of propofol anesthesia: A comparison with the bispectral index
Cortinez, LI; Delfino, AE; Fuentes, R; Munoz, HR
Anesthesia and Analgesia, 104(3): 605-610.
10.1213/01.ane.0000255152.96354.17
CrossRef
Ieee Transactions on Biomedical Engineering
Detrended fluctuation analysis of EEG as a measure of depth of anesthesia
Jospin, M; Caminal, P; Jensen, EW; Lifvan, H; Vallverdu, M; Strays, MMRF; Vereecke, HEM; Kaplan, DT
Ieee Transactions on Biomedical Engineering, 54(5): 840-846.
10.1109/TBME.2007.893453
CrossRef
British Journal of Anaesthesia
Muscle relaxation and depth of anaesthesia: where is the missing link?
Bonhomme, V; Hans, P
British Journal of Anaesthesia, 99(4): 456-460.
10.1093/bja/aem243
CrossRef
Anesthesiology
Cerebral State Index during Propofol Anesthesia: A Comparison with the Bispectral Index and the A-Line ARX Index
Jensen, EW; Litvan, H; Revuelta, M; Rodriguez, BE; Caminal, P; Martinez, P; Vereecke, H; Struys, MM
Anesthesiology, 105(1): 28-36.

PDF (1347)
Anesthesiology
Comparison of Bispectral Index and Composite Auditory Evoked Potential Index for Monitoring Depth of Hypnosis in Children
Weber, F; van Oud-Alblas, HJ; Peters, JW; de Leeuw, TG; Tibboel, D; Klein, J
Anesthesiology, 108(5): 851-857.
10.1097/ALN.0b013e31816bbd6e
PDF (418) | CrossRef
European Journal of Anaesthesiology (EJA)
Comparison of a new composite index based on midlatency auditory evoked potentials and electroencephalographic parameters with bispectral index (BIS) during moderate propofol sedation
Hadzidiakos, D; Petersen, S; Baars, J; Herold, K; Rehberg, B
European Journal of Anaesthesiology (EJA), 23(11): 931-936.
10.1017/S0265021506001050
PDF (116) | CrossRef
Anesthesiology
Response Surface Modeling of the Interaction between Propofol and Sevoflurane
Schumacher, PM; Dossche, J; Mortier, EP; Luginbuehl, M; Bouillon, TW; Struys, MM
Anesthesiology, 111(4): 790-804.
10.1097/ALN.0b013e3181b799ef
PDF (2382) | CrossRef
European Journal of Anaesthesiology (EJA)
Performance of AEP Monitor/2‐derived composite index as an indicator for depth of sedation with midazolam and alfentanil during gastrointestinal endoscopy
Huang, Y; Chu, Y; Chang, K; Wang, Y; Chan, K; Tsou, M
European Journal of Anaesthesiology (EJA), 24(3): 252&hyhen;257.
10.1017/S0265021506001633
PDF (120) | CrossRef
Current Opinion in Anesthesiology
Processed electroencephalogram in depth of anesthesia monitoring
Palanca, BA; Mashour, GA; Avidan, MS
Current Opinion in Anesthesiology, 22(5): 553-559.
10.1097/ACO.0b013e3283304032
PDF (1968) | CrossRef
Anesthesiology
The Effects of Ketamine and Rocuronium on the A-Line® Auditory Evoked Potential Index, Bispectral Index, and Spectral Entropy Monitor during Steady State Propofol and Remifentanil Anesthesia
Vereecke, HE; Vanluchene, AL; Mortier, EP; Everaert, K; Struys, MM
Anesthesiology, 105(6): 1122-1134.

PDF (1852)
Back to Top | Article Outline

© 2005 American Society of Anesthesiologists, Inc.

Publication of an advertisement in Anesthesiology Online does not constitute endorsement by the American Society of Anesthesiologists, Inc. or Lippincott Williams & Wilkins, Inc. of the product or service being advertised.
Login

Article Tools

Images

Share