Investigation of Electrochemical Carbon Monoxide Sensor Monitoring of Anesthetic Gas Mixtures

Bermudez, Joseph A. M.D.*

Free Access
Article Outline
Collapse Box

Author Information

AN electrochemical sensor has previously been reported to provide useful detection of carbon monoxide while exposed to one minimum alveolar concentration of volatile anesthetics. 1 My independent laboratory study yields further evidence supporting the possible utility of electrochemical carbon monoxide sensor measurements in anesthetic gas mixtures.
Several scientific investigations have elucidated the chemistry of base-catalyzed volatile anesthetic degradation into by-products, including carbon monoxide. 2–4 Multiple case reports corroborate the laboratory studies with clinical documentation of significantly elevated intraoperative carboxyhemoglobin levels as a consequence of anesthetic circuit carbon monoxide contamination. 5–7 Only a single case report exists of the suspected correlation between intraoperative carbon monoxide exposure and actual patient morbidity. 8 However, veterinary anesthetic practice may also utilize volatile anesthetic agents in the presence of alkaline carbon dioxide absorbents. Voluntary reporting via the United States Pharmacopeial Convention Veterinary Practitioners Reporting Program recorded a signal clustering (three dogs) of perioperative animal deaths with routine anesthetics/surgeries associated with intraoperative carbon monoxide exposure involving isoflurane and soda lime/Baralyme® (Chemetron Medical Division, Allied Healthcare Products, St. Louis, MO). 9
Current anesthesia monitors provide only indirect evidence of carbon monoxide production in anesthesia circuits. 10–12 Early specific carbon monoxide identification in anesthetic gases could avert patient exposure to that toxic by-product. My in vitro investigation assesses the capability of an electrochemical carbon monoxide sensor in various anesthetic atmospheres with maximal volatile anesthetic concentrations.
Back to Top | Article Outline

Materials and Methods

All carbon monoxide measurements were made at 18.6–23.2°C and one atmosphere using the Bacharach Snifit Model 50 electrochemical carbon monoxide sensor (Bacharach Inc., New Kensington, PA) with a range of 0–1,999 parts per million (ppm) certified for use in 15 to 90% relative humidity (noncondensing). Gas samples were studied with relative humidity ranging from 0 to 52%. Instrument carbon monoxide measurements on 20 samples were obtained for each study atmosphere before and after carbon monoxide addition.
Each of five volatile anesthetic agents (halothane, enflurane, isoflurane, desflurane, and sevoflurane) was tested at the highest possible standard vaporizer delivery concentration in the presence and absence of nitrous oxide (45–64%). The experimental gas test chamber consisted simply of a hard plastic housing for the electrochemical sensor electrode with chrome-plated connecting pipes and plastic stopcocks for gas removal/insertion fitted to borosilicate glass syringes at either end. Certified standard-grade 1.048% carbon monoxide in balance nitrogen was added to the test chamber in quantities verified by room air control trials to result in carbon monoxide concentrations of 38.5 ppm and 385 ppm.
To test the functional resiliency of this electrochemical sensor to complex by-product mixtures, carbon monoxide measurements were also made on a desflurane atmosphere decomposed by dried Sodasorb® (W.R. Grace, Chicago, IL). The decomposition process created a by-product mixture containing carbon monoxide, and samples of that mixture were diluted with air to result in an electrochemical sensor indication of approximately 500 ppm. By-product interference with sensor function was tested by an enrichment technique wherein an aliquot of each sample was removed and exchanged with a certified standard grade carbon monoxide additive sufficient to approximately double the sensor reading. In addition, atmospheres of carbon dioxide (6.38%), oxygen (96.3% and 99.6%), and nitrous oxide (31.9%, 72.2%, and 96.3%) were evaluated by the sensor.
Two-tailed Student t test statistical analysis was employed using the 0.01 significance level.
Back to Top | Article Outline

Results

Fig. 1
Fig. 1
Image Tools
Several volatile anesthetic atmospheres were associated with a statistically significant difference between the mean sensor measurement and the theoretical carbon monoxide level. In all cases except those involving nitrous oxide, however, the observed measurement values did not deviate more than 2% from theoretical values. Hence, none of the volatile agents created any clinically significant instrument bias (fig. 1).
Fig. 2
Fig. 2
Image Tools
Nitrous oxide atmospheres were associated with a statistically significant negative deflection of the sensor readings from theoretical values (fig. 2). Underestimation of actual carbon monoxide level by approximately 13% was determined for atmospheres containing 60% nitrous oxide using this sensor. Nitrous oxide presence also slightly decreased the electrochemical sensor zero (−1 to −2 ppm).
Oxygen (96.3% and 99.6%) caused negative deflection of the room air zero (−3 to −5 ppm) yet minimal effect (less than 2% instrument measurement bias) on the mean carbon monoxide level determinations. Carbon dioxide (6.38%) did not appreciably affect the room air zero or the sensor carbon monoxide measurements.
Detection of 98% of the carbon monoxide enrichment additive to degraded desflurane by-product mixtures demonstrated that the sensor was virtually free of by-product interference as measured by that technique.
Back to Top | Article Outline

Discussion

Previous research and reports affirm that inadvertent carbon monoxide generation by volatile anesthetic degradation is indeed a real albeit rare problem. 5–9 The findings of high carboxyhemoglobin levels in patients and significant carbon monoxide contamination in anesthesia circuits are startling realities that should heighten anesthesia practitioner awareness and vigilance for this insidious poison.
The results of this study further validate the ability of an electrochemical carbon monoxide sensor to detect carbon monoxide appropriately within a clinically useful range while exposed to the highest volatile anesthetic concentrations attainable in an anesthesia circuit. In addition, the electrochemical sensor properly measured carbon monoxide concentrations during exposure to a complex gas mixture of base-decomposed desflurane and attendant by-products. A concentration-dependent negative instrument bias effect of nitrous oxide on the electrochemical sensor carbon monoxide readings is described, which, although interesting, still allows sensitive detection of clinically significant carbon monoxide in the anesthesia circuit. Because carbon monoxide possesses manifold toxicity mechanisms and may be quickly poisonous even to healthy adults, protection of patients from such a contaminant is an obvious imperative.
The author gratefully acknowledges the incomparable expertise of Edmond I Eger, II, M.D., University of California, San Francisco, San Francisco, California, for helpful critique and Dennis M. Fisher, M.D., “P Less Than” Company, San Francisco, California, for assistance in statistical analysis.
Back to Top | Article Outline

References

1. Dunning M, Woehlck HJ: Performance of an electrochemical carbon monoxide monitor in the presence of anesthetic gases. J Clin Monit 1997; 13: 357–62

2. Baxter PJ, Garton K, Kharasch ED: Mechanistic aspects of carbon monoxide formation from volatile anesthetics. A nesthesiology 1998; 89: 929–41

3. Fang ZX, Eger EI, Laster MJ, Chortkoff BS, Kandel L, Ionescu P: Carbon monoxide production from degradation of desflurane, enflurane, isoflurane, halothane, and sevoflurane by soda lime and Baralyme®. Anesth Analg 1995; 80: 1187–93

4. Woehlck HJ, Dunning M, Raza T, Ruiz F, Bolla B, Zink W: Physical factors affecting the production of carbon monoxide from anesthetic breakdown. A nesthesiology 2001; 94: 453–6

5. Centers for Disease Control and Prevention Morbidity and Mortality Weekly Report: Epidemiologic notes and reports [of] elevated intraoperative blood carboxyhemoglobin levels in surgical patients: Georgia, Illinois, and North Carolina. MMWR Morb Mortal Wkly Rep 1991; 40: 248–9

6. Berry PD, Sessler DI, Larson MD: Severe carbon monoxide poisoning during desflurane anesthesia. A nesthesiology 1999; 90: 613–6

7. Pearson ML, Levine WC, Finton RJ, Ingram CT, Gay KB, Tapelband G, Smith D, Jarvis WR: Anesthesia-associated carbon monoxide exposures among surgical patients. Infect Hosp Epidemiol 2001; 22: 352–6

8. ECRI: Carbon monoxide exposures during inhalation anesthesia: the interaction between halogenated anesthetic agents and carbon dioxide absorbents. Health Devices 2000; 29: 435–8

9. U.S. Pharmacopeia: United States Pharmacopeial Convention: Veterinary Practitioners Reporting Program, Report No. 002608, Rockville, Maryland, U.S. Pharmacopeia, 1998

10. Woehlck HJ, Dunning MB, Kulier AH, Sasse FJ, Nithipataikom K, Henry DW: The response of anesthetic agent monitors to trifluoromethane warns of the presence of carbon monoxide from anesthetic breakdown. J Clin Monit 1997; 13: 149–55

11. Woehlck HJ, Dunning M, Gandhi S, Chang D, Milosavljevic D: Indirect detection of intraoperative carbon monoxide exposure by mass spectrometry during isoflurane anesthesia. A nesthesiology 1995; 83: 213–7

12. Woehlck HJ, Dunning M, Nithipatikom K, Kulier AH, Henry DW: Mass spectrometry provides warning of carbon monoxide exposure via trifluoromethane. A nesthesiology 1996; 84: 1489–93

Cited By:

This article has been cited 2 time(s).

Anesthesia and Analgesia
Detection of Carbon Monoxide During Routine Anesthetics in Infants and Children
Levy, RJ; Nasr, VG; Rivera, O; Roberts, R; Slack, M; Kanter, JP; Ratnayaka, K; Kaplan, RF; McGowan, FX
Anesthesia and Analgesia, 110(3): 747-753.
10.1213/ANE.0b013e3181cc4b9f
CrossRef
Anesthesiology
Sleeping with Uncertainty:: Anesthetics and Desiccated Absorbent
Woehlck, HJ
Anesthesiology, 101(2): 276-278.

PDF (712)
Back to Top | Article Outline

© 2003 American Society of Anesthesiologists, Inc.

Publication of an advertisement in Anesthesiology Online does not constitute endorsement by the American Society of Anesthesiologists, Inc. or Lippincott Williams & Wilkins, Inc. of the product or service being advertised.
Login

Article Tools

Images

Share