The Combitube in Elective Surgery: A Report of 200 Cases

Gaitini, Luis A. M.D.*; Vaida, Sonia J. M.D.†; Mostafa, Somri M.D.‡; Yanovski, Boris M.D.§; Croitoru, Milian M.D.∥; Capdevila, Marisa D. M.D.#; Sabo, Edmond M.D.**; Ben-David, Bruce M.D.††; Benumof, Jonathan M.D.‡‡

Free Access
Article Outline
Collapse Box

Author Information

Collapse Box

Abstract

Background: The Combitube has proved to be a valuable device for securing the airway in cases of difficult intubation. This study investigated the effectiveness of the Combitube in elective surgery during both mechanical and spontaneous ventilation.
Methods: Two hundred patients classified as American Society of Anesthesiologists physical status I and II, with normal airways, scheduled for elective surgery were randomly allocated into two groups: nonparalyzed, spontaneously breathing (n = 100); or paralyzed, mechanically ventilated (n = 100). After induction of general anesthesia and insertion of the Combitube, oxygen saturation, end-tidal carbon dioxide and isoflurane concentration, systolic and diastolic blood pressure and heart rate, as well as breath-by-breath spirometry data were obtained every 5 min.
Results: In 97% of patients, it was possible to maintain oxygenation, ventilation, and respiratory mechanics, as well as hemodynamic stability during either mechanical or spontaneous ventilation for the entire duration of surgery. The duration of surgery was between 15 and 155 min.
Conclusions: The results of this study suggest that the Combitube is an effective and safe airway device for continued management of the airway in 97% of elective surgery cases.
THE Combitube (Tyco-Kendall-Sheridan, Mansfield, MA) is a new device for difficult airway management. 1,2 The American Society of Anesthesiologists’ Task Force on Management of the Difficult Airway suggests considering the use of the Combitube when intubation problems occur in patients with a previously unrecognized difficult airway, especially in a “cannot ventilate, cannot intubate” situation. 3 The advantages of the Combitube include rapid airway control without the need for neck or head movement, minimized risk for aspiration, firm fixation of the device after inflation of the oropharyngeal balloon, and the fact that it works equally well in either the tracheal or esophageal position. 2,4 It has been demonstrated that the Combitube can be placed by anesthesiologists with relatively little formal training. 5
The Combitube combines the functions of an esophageal obturator airway and a conventional endotracheal tube. Therefore, ventilation of the patient can be achieved with the Combitube positioned in either the esophagus or the trachea. Although the role of the Combitube has been established in the emergency management of the difficult airway and has been previously used for mechanical ventilation in intensive care unit patients, 6 its use for continued airway management in the surgical setting is uncertain. The aim of this study was to determine the safety and effectiveness of the Combitube in primary airway management for routine surgery with both mechanical and spontaneous ventilation of patients.
Back to Top | Article Outline

Methods

The study was approved by the Human Ethics Committee (B’nai Zion Medical Center, Haifa, Israel), and written consent was obtained from all patients. Subjects were American Society of Anesthesiologists physical status I and II, aged 18–75 yr, who where scheduled for elective surgery with general anesthesia. The Samsoon and Young modification 7 of the Mallampati airway classification 8 was used to evaluate the airway preoperatively, and only patients with a score of I or II were included in the study. Exclusion criteria included known esophageal disease, pulmonary diseases, cardiovascular diseases, patients with a height less than 4 feet, 20% deviation from the ideal body weight, and blind tracheal introduction of the Combitube. The patients were randomized (blocked randomization) into two groups: mechanical ventilation (n = 100) and spontaneous ventilation (n = 100).
The anesthesiologists included in the study were experienced in the use of the Combitube. All patients received a premedication of oral diazepam 10–20 mg. Anesthesia was induced with up to 3 μg/kg fentanyl and 2–3 mg/kg propofol and maintained with 70% N2O in 30% (remainder) oxygen and isoflurane. No muscle relaxants were used in the patients who were spontaneously breathing.
A no. 3 Macintosh laryngoscope blade was used in all patients for introducing the Combitube. In the mechanically ventilated group, neuromuscular blockade was obtained with vecuronium 0.1 mg/kg and assessed using a peripheral nerve stimulator with electrodes placed over the median nerve. After confirmation of neuromuscular blockade, a Combitube was inserted until the printed ringmarks were lying between teeth or alveolar ridges, in accordance with the manufacturer’s recommendations. The lungs were ventilated with volume controlled mechanical ventilation using a ventilator of an AS/3 anesthesia delivery unit (Datex-Ohmeda, Helsinki, Finland) with a semiclosed circuit incorporating a carbon dioxide absorber. Ventilatory settings remained unchanged throughout the procedure, with an inspiratory/expiratory time of 1:2 and an average tidal volume of 12 ml/kg.
In the spontaneous ventilation group, the same anesthesia protocol was used, excluding the neuromuscular relaxants, and the Combitube was inserted after the eyelash reflex was lost. The oropharyngeal balloon of the 37- and 41-French Combitube was inflated with 85 and 100 ml of air, respectively, and the distal balloon was inflated with 10–12 ml and 10–15 ml of air, respectively. Proper positioning of the Combitube was confirmed by bilateral chest movement, bilateral chest auscultation, absence of gastric insufflation, pulse oximetry, and partial pressure of end-tidal carbon dioxide (ETCO2). The oropharyngeal balloon was deflated every 20 min for 1 min to avoid excessive pressure on the pharyngeal mucosa and tongue swelling.
Blood pressure, heart rate, oxygen saturation (SpO2), ETCO2, and end-tidal isoflurane concentration were measured using the AS-3 Datex-Ohmeda monitor. Breath-by-breath spirometry data was obtained using a side-stream spirometry device (D-lite flow sensor; Datex-Ohmeda) attached between the proximal end of the Combitube and the Y piece, which continuously computes flow and pressure readings derived via a pressure system. Data measured included airway pressures (peak, plateau, and positive end-expiratory pressure), lung volumes (minute and tidal volume), graphically displayed loops (pressure–volume and flow–volume) and curves (pressure and flow), the ratio of passively exhaled volume during the first second to the total expiratory tidal volume (V1.0%), and dynamic compliance. Gas leak was observed as a nonclosing flow–volume loop or by auscultation with a stethoscope placed on the neck region. Data were checked at 5-min intervals, beginning after the introduction of the Combitube. Upper airway trauma was assessed by checking for the presence of blood on the Combitube after its removal and 24 h after surgery the patients were examined for sore throat and hoarseness.
The study was terminated immediately if ventilation of the patient’s lungs was clinically unacceptable, peak airway pressure exceeded 40 cm H2O, or SpO2 decreased below 90%.
Back to Top | Article Outline
Statistical Analysis
Statistical process control (x control charts) was used to determine the stability of each measured data. The data was considered stable if it was within the limits of the mean value ± 3 SD and in the absence of a “run” or a “trend.” A run was defined as at least seven successive points on the same side of the mean value. A trend was defined as a succession of seven values that are increasing or decreasing. Data were also analyzed using the Friedman analysis of variance test followed by Wilcoxon matched-pairs signed ranks tests with an appropriate correction analysis of variance for repeated measurements. A P value less than 0.05 was considered significant.
Back to Top | Article Outline

Results

Table 1
Table 1
Image Tools
Patients’ characteristics and duration of surgery are shown in table 1. The type of surgeries for mechanical ventilation were general surgery extraabdominal (n = 42), general surgery intraabdominal (n = 13), gynecology (n = 26), urology (n = 9), and orthopedic (n = 10), and for spontaneous ventilation were general surgery extraabdominal (n = 49), gynecology (n = 6), urology (n = 12), and orthopedic (n = 33).
The Combitube was not used successfully in six patients: four in the mechanical ventilation group (three patients because of airway pressures greater than 40 cm H2O and one because of clinically unacceptable ventilation) and two in the spontaneous ventilation group because of difficulty in ventilation. In one patient the Combitube happened to be placed into the trachea. In the six patients who could not be treated with the Combitube, a standard endotracheal intubation was performed.
Table 2
Table 2
Image Tools
In 97% of patients, it was possible to maintain respiratory mechanics during either mechanical or spontaneous ventilation for the entire duration of surgery as reflected by the spirometry data (table 2).
Table 3
Table 3
Image Tools
Hemodynamic stability was maintained throughout the surgery (table 3). Oxygen saturation, ETCO2, and end-tidal isoflurane concentrations were 97.2 ± 4%, 36.5 ± 5 mmHg, and 0.7 ± 3%, respectively, in the mechanical ventilation group and 96.3 ± 3%, 51.6 ± 4 mmHg, and 0.6 ± 1%, respectively, in the spontaneous ventilation group.
Table 4
Table 4
Image Tools
The scoring of upper airway trauma is presented in table 4. Thirty-five percent of all patients complained of sore throat, and 10% complained of hoarseness 24 h after surgery. These patients required no special treatment and recovered forthwith.
In 15 patients, gas leak was detected at a ventilation pressure of 25 cm H2O, which disappeared after overinflation of the oropharyngeal balloon with an additional 15—20 ml of air. Statistical process control charts showed stable systems for all data. No significant statistical differences were observed by Friedman analysis of variance within the groups for all the data.
Back to Top | Article Outline

Discussion

The difficult airway algorithm of the American Society of Anesthesiologists Task Force on Management of the Difficult Airway 3 suggests using the Combitube when intubation problems occur in patients with a previously unrecognized difficult airway, especially in a “cannot ventilate, cannot intubate” situation. It has been proven to be successful in the airway management in patients with an anticipated difficult airway; however, after its insertion it was replaced by a standard endotracheal tube by the aid of an fiberotic bronchoscope. 9 No published study has assessed the possibility of leaving the Combitube in place after having secured the airway and using it as the definitive airway for the duration of anesthesia. The results of this study suggest that the Combitube appears suitable for anesthesia of at least moderate duration with the use of either spontaneous or mechanical ventilation in patients with near ideal body weight.
One of the limitations of this study is that only patients with anticipated normal airways were included, and thus the performance of the Combitube as a definitive airway in patients with expected difficult intubation remains undefined. Further study is required to examine the question of whether these findings apply to the abnormal or difficult airway.
It is noted that among the ventilatory parameters, a slightly increased positive end-expiratory pressure in both groups and respiratory peak pressure, as well as decreased compliance as compared with normal values, were found in the mechanical ventilation group. Frass et al.10 reported a prolonged expiratory flow time with the Combitube as compared with an endotracheal tube, caused by an increase in expiratory resistance caused by the double-lumen design of the Combitube and the small holes of the esophageal lumen. This effect could be responsible for the slightly decreased compliance value (32.1 ml/cm H20) in the mechanical ventilation group and the slightly increased positive end-expiratory pressure observed in both groups.
Despite a relatively high peak pressure, we did not observe air leakage past the oropharyngeal balloon as checked by observation of the flow–volume loop and by auscultation. Furthermore, the 10–15-ml inflation volume of the esophageal balloon is probably ideal for preventing air leakage to the esophagus and consequent gastric dilatation. Unlike the laryngeal mask airway, the Combitube provides the advantage of airway protection from gastric aspiration, 11 as well as the possibility of emptying the stomach contents by introducing a gastric tube.
The exact incidence of complications associated with the use of Combitube is unknown. In addition to upper airway trauma, 12,13 major complications, including subcutaneous emphysema, pneumomediastinum, and pneumoperitoneum, have been reported. 14
In contrast to Mercer and Gabbott, 12 who reported a 45% incidence of upper airway trauma, we observed a lower incidence (27%) evaluated by checking for the presence of blood on the Combitube after its removal. Our findings of a lower degree of trauma may be explained by the routine use of a laryngoscope in all cases when introducing the Combitube. Previous reports have shown that the laryngoscope is often helpful in Combitube placement, encouraging the manufacturer to include this recommendation on the package insert. 2,5
It is likely that a high inflation volume of the oropharyngeal balloon of the Combitube causes mucosal tears by stretching the pharyngeal mucosa. 13 Decreasing the inflation volume of the oropharyngeal balloon of the Combitube to a point just above that which allows air leakage may reduce the incidence of this injury.
In a recent report, Oczenski et al.15 found a significantly higher hemodynamic and catecholamine stress response after insertion of the Combitube as compared with the laryngeal mask airway and endotracheal intubation. Unlike their findings, we did not observe tachycardia or elevated blood pressures in our patients. However, we did not measure catecholamine levels in the present study.
The use of the Combitube in the management of the difficult airway has been well documented. Heretofore, however, the continued airway management of the surgical patient who has been intubated with a Combitube has been unclear. We believe that the Combitube can be suitable for continued airway mamagement with an already-placed Combitube for anesthesia of a least moderate duration with the use of either spontaneous or mechanical ventilation.
Back to Top | Article Outline

References

1. Benumof JL: Management of the difficult airway. Anesthesiology 1991; 75: 1087–110

2. Frass M: The combitube: Esophageal/tracheal double lumen airway, Airway Management: Principles and Practice.Edited by Benumof JL. St. Louis, Mosby, 1996, pp 444–55

3. American Society of Anesthesiologists Task Force on Management of the Difficult Airway: Practice guidelines for management of the difficult airway. Anesthesiology 1993; 78: 597–602

4. Frass M, Frenzer R, Zdrahal F, Hofleher G, Porges P, Lackner F: The esophageal tracheal combitube preliminary results with a new airway for CPR. Ann Emerg Med 1987; 16: 768–72

5. Bishop M, Kharasch ED: Is the Combitube a useful emergency device for anesthesiologists? Anesth Analg 1998; 86: 1141–2

6. Frass M, Frenzer R, Mayer G, Popovic R, Leithner C: Mechanical ventilation with the esophageal tracheal combitube (ETC) in the intensive care unit. Arch Emerg Med 1987; 4: 219–25

7. Samsoon GLT, Young JRB: Difficult tracheal intubation: A retrospective study. Anaesthesia 1987; 42: 487–90

8. Mallampati SR, Gatt SP, Gugino LD, Desai SP, Waraksa B, Freiberger D, Liu PL: A clinical sign to predict difficult tracheal intubation: A prospective study. Can Anaesth Soc J 1985; 32: 429–34

9. Gaitini LA, Vaida SJ, Somri M, Fradis M, Ben-David B: Fiberoptic-guided airway exchange of the esophageal-tracheal Combitube in spontaneously breathing vs mechanically ventilated patients. Anesth Analg 1999; 88: 193–6

10. Frass M, Rodler S, Frenzer R, Ilias W, Leithner C, Lackner F: Esophageal tracheal combitube, endotracheal airway, and mask: Comparison of ventilatory pressure curves. J Trauma 1989; 29: 1476–9

11. Urtubia RM, Aguila CM, Cumsille MA: Combitube: A study of proper use. Anesth Analg 2000; 90: 958–62

12. Mercer MH, Gabbott DA: The influence of neck position on ventilation using the combitube airway. Anaesthesia 1998; 53: 146–50

13. Oczenski W, Krenn H, Dahaba AA, Binder M, El-Schahawi-Kienzl I, Kohout S, Schwartz S, Fitzgerald RD: Complications following the use of the Combitube, tracheal tube and laryngeal mask airway. Anaesthesia 1999; 54: 1161–5

14. Vézina D, Lessard MR, Bussières J, Topping C, Trepanier CA: Complications associated with the use of the esophageal-tracheal Combitube. Can J Anesth 1998; 45: 76–80

15. Oczenski W, Krenn H, Dahaba AA, Binder M, El-Schhawi-Kienzl I, Jellinek H, Schwatz S, Fitzgerald RD: Hemodynamic and catecholamine stress response to insertion of the combitube, laryngeal mask airway or tracheal intubation. Anesth Analg 1999; 88: 1389–94

Cited By:

This article has been cited 40 time(s).

Anaesthesia
Consequences of cuff pressure
Mandal, NG
Anaesthesia, 57(7): 717-718.

Acta Anaesthesiologica Scandinavica
Tracheal esophageal combitube: a useful airway for morbidly obese patients who cannot intubate or ventilate
Della Puppa, A; Pittoni, G; Frass, M
Acta Anaesthesiologica Scandinavica, 46(7): 911-913.

Anasthesiologie Intensivmedizin Notfallmedizin Schmerztherapie
The significance of the Combitube in the clinical routine and in emergencies - Replies
Hrska, F
Anasthesiologie Intensivmedizin Notfallmedizin Schmerztherapie, 37(): 763-764.

Journal of Clinical Anesthesia
Effective use of the esophageal tracheal Combitube (TM) following severe burn injury
Hagberg, CA; Johnson, S; Pillai, D
Journal of Clinical Anesthesia, 15(6): 463-466.
10.1016/S0952-8180(03)00109-0
CrossRef
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie
Complications associated with the Esophageal-Tracheal Combitube (R) in the pre-hospital setting
Vezina, MC; Trepanier, CA; Nicole, PC; Lessard, MR
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie, 54(2): 124-128.

Wiener Klinische Wochenschrift
Safeguarding of the air way in emergency situation
Frass, M
Wiener Klinische Wochenschrift, 120(): 191-196.
10.1007/s00508-008-0963-z
CrossRef
Wiener Klinische Wochenschrift
Comparison of mouth-to-mouth resuscitation and Combitube ventilation in a bench model
Sherif, C; Erdos, J; Sohm, M; Schonbauer, R; Schellongowski, P; Kostler, W; Boltuch, J; Rabitsch, W; Schuster, E; Frass, M
Wiener Klinische Wochenschrift, 117(): 769-775.

Minerva Anestesiologica
New devices and techniques for airway management
Agro, FE; Cataldo, R; Mattei, A
Minerva Anestesiologica, 75(3): 141-149.

Journal of Clinical Anesthesia
Current status of the Combitube (TM): A review of the literature
Agro, F; Frass, M; Benumof, JL; Krafft, P
Journal of Clinical Anesthesia, 14(4): 307-314.
PII S0952-8180(02)00356-2
CrossRef
Military Medicine
Airway and ventilatory equipment in field anesthesia: What's new?
Tang, KC; Chiu, JW; Low, E
Military Medicine, 169(5): 342-348.

Acta Anaesthesiologica Scandinavica
'Tricks of the trade' with the Esophageal-Tracheal Combitube (R)
Urtubia, RM
Acta Anaesthesiologica Scandinavica, 46(3): 340.

Anasthesiologie Intensivmedizin Notfallmedizin Schmerztherapie
The importance of the esophageal-tracheal-double lumen tube for clinical use and emergencies
Hrska, F; Hofbauer, R; Staudinger, T; Schellongowski, P; Frass, M
Anasthesiologie Intensivmedizin Notfallmedizin Schmerztherapie, 37(7): 378-385.

Anesthesia and Analgesia
An evaluation of the laryngeal Tube (R) during general anesthesia using mechanical ventilation
Gaitini, LA; Vaida, SJ; Somri, M; Kaplan, V; Yanovski, B; Markovits, R; Hagberg, CA
Anesthesia and Analgesia, 96(6): 1750-1755.
10.1213/01.ANE.0000066252.60734.4E
CrossRef
Anaesthesia
Difficult Airway Society guidelines for management of the unanticipated difficult intubation
Henderson, JJ; Popat, MT; Latto, IP; Pearce, AC
Anaesthesia, 59(7): 675-694.

Anaesthesist
Comparison of algorithms for management of the difficult airway
Heidegger, T; Gerig, HJ; Keller, C
Anaesthesist, 52(5): 381-392.
10.1007/s00101-003-0501-3
CrossRef
Acta Anaesthesiologica Scandinavica
Airway management practices at German university and university-affiliated teaching hospitals - equipment, techniques and training: results of a nationwide survey
Goldmann, K; Braun, U
Acta Anaesthesiologica Scandinavica, 50(3): 298-305.
10.1111/j.1399-6576.2005.00853.x
CrossRef
Wiener Klinische Wochenschrift
Evaluation of the oesophageal-tracheal double-lumen tube (Combitube) during general anaesthesia
Rabitsch, W; Krafft, P; Lackner, FX; Frenzer, R; Hofbauer, R; Sherif, C; Frass, M
Wiener Klinische Wochenschrift, 116(3): 90-93.

Annals of Emergency Medicine
The Combitube as rescue device: Recommended use of the small adult size for all patients six feet tall or shorter
Levitan, RM; Frass, M
Annals of Emergency Medicine, 44(1): 92.
10.1016/j.annemergmed.2003.12.034
CrossRef
Anaesthesist
Intubation anesthesia and fixed dental appliances
Thierbach, A
Anaesthesist, 50(8): 615-616.

Acta Anaesthesiologica Scandinavica
New insertion technique of the Esophageal-Tracheal Combitube (R)
Urtubia, RM; Frass, M; Agro, F
Acta Anaesthesiologica Scandinavica, 46(3): 340-341.

Resuscitation
Comparison of a conventional tracheal airway with the Combitube in an urban emergency medical services system run by physicians
Rabitsch, W; Schellongowski, P; Staudinger, T; Hofbauer, R; Dufek, V; Eder, B; Raab, H; Thell, R; Schuster, E; Frass, M
Resuscitation, 57(1): 27-32.
10.1016/S0300-9572(02)00435-5
CrossRef
Acta Anaesthesiologica Scandinavica
Less tongue engorgement with lateral placement of the Esophageal Tracheal Combitube
Ahmed, SM; Rizvi, KA; Khan, RM; uz Zafar, M; Nadeem, A
Acta Anaesthesiologica Scandinavica, 52(6): 834-837.
10.1111/j.1399-6576.2007.01546.x
CrossRef
Anasthesiologie Intensivmedizin Notfallmedizin Schmerztherapie
The significance of the Combitube in the clinical routine and in emergencies
Oczenski, W
Anasthesiologie Intensivmedizin Notfallmedizin Schmerztherapie, 37(): 762.

Neurocritical Care
Airway management in neurological emergencies
Roppolo, LP; Walters, K
Neurocritical Care, 1(4): 405-414.

Annals of Emergency Medicine
Recommendation of the minimal volume technique to avoid tongue engorgement with prolonged use of the esophageal-tracheal combitube
Rabitsch, W; Kostler, WJ; Burgmann, H; Krafft, P; Frass, M
Annals of Emergency Medicine, 45(5): 566.
10.1016/j.annemergmed.2004.11.031
CrossRef
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie
Supraglottic airways and pulmonary aspiration: the role of the drain tube
Drolet, P
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie, 56(): 715-720.
10.1007/s12630-009-9156-0
CrossRef
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie
Combitube((TM)) rescue for Cesarean delivery followed by ninth and twelfth cranial nerve dysfunction
Zamora, JE; Saha, TK
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie, 55(): 779-784.

Journal of Clinical Anesthesia
Comparison of the EasyTube (R) and endotracheal tube during general anesthesia in fasted adult patients
Lorenz, V; Rich, JM; Schebesta, K; Taslakian, S; Mullner, M; Frass, M; Schuster, E; Illievich, UM; Kaye, AD; Vaida, S; Krafft, P
Journal of Clinical Anesthesia, 21(5): 341-347.
10.1016/j.jclinane.2008.09.008
CrossRef
Anaesthesist
Methods of airway management in prehospital emergency medicine
Keul, W; Bernhard, M; Volkl, A; Gust, R; Gries, A
Anaesthesist, 53(): 978-992.
10.1007/s00101-004-0734-9
CrossRef
Journal of Clinical Anesthesia
The esophageal-tracheal combitube resistance and ventilatory pressures
Gaitini, LA; Vaida, SJ; Mostafa, S; Yanovski, B; Ben-David, B; Benumof, JL
Journal of Clinical Anesthesia, 17(1): 26-29.
10.1016/j.jclinane.2004.03.010
CrossRef
Military Medicine
The combitube, self-inflating bulb, and colorimetric carbon dioxide detector to advance airway management in the first echelon of the battlefield
Rich, JM; Thierbach, A; Frass, M
Military Medicine, 171(5): 389-395.

Ent-Ear Nose & Throat Journal
Endotracheal intubation with flexible fiberoptic bronchoscopy in patients with abnormal anatomic conditions of the head and neck
Elizondo, E; Navarro, F; Perez-Romo, A; Ortega, C; Munoz, H; Cicero, R
Ent-Ear Nose & Throat Journal, 86(): 682-684.

Anesthesiology
Combitube and Similar Devices Should Undergo Long-term Safety Evaluation before Their Routine Use in Clinical Practice
Mandal, NG
Anesthesiology, 95(4): 1041-1042.

Anesthesiology
Airway Management with Endotracheal Tube versus Combitube® during Parabolic Flights
Rabitsch, W; Moser, D; Inzunza, MR; Niedermayr, M; Köstler, WJ; Staudinger, T; Locker, GJ; Schellongowski, P; Wulkersdorfer, B; Rich, JM; Meyer, B; Benumof, JL; Frass, M
Anesthesiology, 105(4): 696-702.

PDF (444)
Anesthesiology
Comparison of Mucosal Pressures Induced by Cuffs of Different Airway Devices
Schuster, E; Frass, M; Ulrich-Pur, H; Hrska, F; Krafft, P; Friehs, H; Wulkersdorfer, B; Köstler, WJ; Rabitsch, W; Staudinger, T
Anesthesiology, 104(5): 933-938.

PDF (451)
Anesthesiology
Is the Combitube Traumatic?
Urtubia, RM; Gazmuri, RR
Anesthesiology, 98(4): 1021-1022.

PDF (1442)
Anesthesiology
Is It Unethical to Use the Combitube in Elective Surgery Patients?
Krafft, P; Hartmann, T; Agro, F; Gaitini, LA; Vaida, SJ
Anesthesiology, 98(4): 1022.

PDF (1442)
Anesthesiology
Is It Unethical to Use the Combitube in Elective Surgery Patients?: In Reply:
Brimacombe, JR; Keller, C
Anesthesiology, 98(4): 1022-1024.

PDF (1442)
European Journal of Anaesthesiology (EJA)
A comparison of the proseal laryngeal mask airway™, the laryngeal tube S® and the oesophageal‐tracheal combitube™ during routine surgical procedures
Bein, B; Carstensen, S; Gleim, M; Claus, L; Tonner, PH; Steinfath, M; Scholz, J; Dörges, V
European Journal of Anaesthesiology (EJA), 22(5): 341-346.
10.1017/S026502150500058X
PDF (95) | CrossRef
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie
Improving styletted oral tracheal intubation: rational use of the OTSU
Stasiuk, RBP
Canadian Journal of Anaesthesia-Journal Canadien D Anesthesie, 48(9): 911-918.

Back to Top | Article Outline

© 2001 American Society of Anesthesiologists, Inc.

Publication of an advertisement in Anesthesiology Online does not constitute endorsement by the American Society of Anesthesiologists, Inc. or Lippincott Williams & Wilkins, Inc. of the product or service being advertised.
Login

Article Tools

Images

Share