Home Current Issue Previous Issues Podcasts Online First ASA Practice Parameters CME For Authors Journal Info
Skip Navigation LinksHome > December 2013 - Volume 119 - Issue 6 > Failure of the Laryngeal Mask Airway Unique™ and Classic™ in...
Anesthesiology:
doi: 10.1097/ALN.0000000000000015
Perioperative Medicine

Failure of the Laryngeal Mask Airway Unique™ and Classic™ in the Pediatric Surgical Patient: A Study of Clinical Predictors and Outcomes

Mathis, Michael R. M.D.*; Haydar, Bishr M.D.; Taylor, Emma L. M.B.Ch.B., F.R.C.A.; Morris, Michelle M.S.§; Malviya, Shobha V. M.D.; Christensen, Robert E. M.D.; Ramachandran, Satya-Krishna M.B.B.S., F.R.C.A.; Kheterpal, Sachin M.D., M.B.A.

University of Michigan
Collapse Box

Abstract

Background: Although predictors of laryngeal mask airway failure in adults have been elucidated, there remains a paucity of data regarding laryngeal mask airway failure in children.
Methods: The authors performed a retrospective database review of all pediatric patients who received a laryngeal mask anesthetic at their institution from 2006 to 2010. Device brands were restricted to LMA Unique™ (Cardinal Health, Dublin, OH) and LMA Classic™ (LMA North America, San Diego, CA), and primary outcome was laryngeal mask failure, defined as any airway event requiring device removal and tracheal intubation. Potential risk factors were analyzed with both univariate and multivariate techniques and included medical history, physical examination, surgical, and anesthetic characteristics.
Results: Of the 11,910 anesthesia cases performed in the study, 102 cases (0.86%) experienced laryngeal mask failure. Common presenting features of laryngeal mask failures included leak (25%), obstruction (48%), and patient intolerance such as intractable coughing/bucking (11%). Failures occurred before incision in 57% of cases and after incision in 43%. Independent clinical associations included ear/nose/throat surgical procedure, nonoutpatient admission status, prolonged surgical duration, congenital/acquired airway abnormality, and patient transport.
Conclusions: The findings of the study support the use of the LMA Unique™ and LMA Classic™ as reliable pediatric supraglottic airway devices, demonstrating relatively low failure rates. Predictors of laryngeal mask airway failure in the pediatric surgical population do not overlap with those in the adult population and should therefore be independently considered.

© 2013 American Society of Anesthesiologists, Inc.

Publication of an advertisement in Anesthesiology Online does not constitute endorsement by the American Society of Anesthesiologists, Inc. or Lippincott Williams & Wilkins, Inc. of the product or service being advertised.