Transesophageal Echocardiography During MitraClip® Procedure

Guarracino, Fabio MD*; Baldassarri, Rubia MD*; Ferro, Baldassare MD*; Giannini, Cristina MD, PhD; Bertini, Pietro MD*; Petronio, Anna Sonia MD; Di Bello, Vitantonio MD; Landoni, Giovanni MD; Alfieri, Ottavio MD§

Free Access
Supplemental Author Material
Article Outline
Collapse Box

Author Information

From the *Department of Cardiothoracic Anesthesia and Intensive Care Medicine, Azienda Ospedaliero Universitaria Pisana; Cardiac Thoracic and Vascular Department, University of Pisa, Pisa; and Departments of Anesthesiology and Intensive Care, and §Cardiac Surgery, Università Vita-Salute San Raffaele, Milano, Italy.

Accepted for publication January 24, 2014.

Funding: Department of Cardiothoracic Anesthesia and Intensive Care Medicine, Azienda Ospedaliero Universitaria Pisana, Pisa, Italy. No external funds.

The authors declare no conflicts of interest.

Supplemental digital content is available for this article. Direct URL citations appear in the printed text and are provided in the HTML and PDF versions of this article on the journal’s web site.

Reprints will not be available from the authors.

Address correspondence to Fabio Guarracino, MD, Department of Cardiothoracic Anesthesia and Intensive Care Medicine, Azienda Ospedaliero Universitaria Pisana, Italy, Via Paradisa 2 Cisanello -56123 Pisa. Address e-mail to fabiodoc64@hotmail.com.

Collapse Box


The percutaneous mitral valve (MV) repair procedure performed with the MitraClip® delivery system is increasingly used to treat severe mitral regurgitation in high-risk patients. The treatment involves percutaneous insertion and positioning of a clip between the MV leaflets. Transesophageal echocardiography (TEE) plays a key role in the procedure by providing information regarding clip navigation, clip alignment to the MV coaptation line, transmitral advancement of the system, leaflet grasping, confirmation of valve tissue catching, and assessment of the final result. Real-time 3-dimensional TEE has increasing value in percutaneous MV repair providing high-quality visualization of both the heart and the intravascular devices. Optimal visualization by 3-dimensional TEE is obtained through both the atrial and ventricular aspects. In contrast to MV surgery, where TEE is involved in the prebypass assessment phase and in evaluation of the final repair, TEE is mandatory to guide management during MitraClip® repair. Cardiac anesthesiologists may provide assistance to interventional cardiologists during the procedure itself in addition to their anesthetic-related tasks.

Mitral regurgitation (MR) is one of the most common valvular heart diseases, second only to aortic valve stenosis.1 According to international guidelines, acute MR may occur due to papillary muscle rupture during myocardial infarction and generally requires urgent surgery because of the severity of hemodynamic impairment. Chronic MR requires surgical treatment in symptomatic patients, whereas asymptomatic patients should be evaluated individually for surgery.2,3 Despite the lack of consensus regarding the timing of intervention, asymptomatic patients with chronic MR can be recommended for surgery when it is associated with depressed left ventricular (LV) function, LV enlargement, atrial fibrillation, or pulmonary hypertension.2,4–6 It has been established that, when feasible, mitral valve (MV) repair is preferable to valve replacement due to fewer perioperative complications and better clinical outcome.7–9 Recently, percutaneous MV repair has been increasingly recognized as a valid option for high surgical risk patients with symptomatic MR.10,11 The EVEREST (Endovascular Valve Edge-to-Edge Repair Study) trials have convincingly demonstrated that percutaneous MV repair with the MitraClip® delivery system (Evalve Inc., Menlo Park, CA) effectively reduces MR.12,13 Data from EVEREST I suggested that percutaneous repair with the MitraClip® system significantly reduced MR (≤ 2+) in most patients, thereby reducing rates of recurrent MR, additional surgery, and death.14

In addition, the percutaneous procedure is generally well tolerated by high-risk patients with LV dysfunction and major comorbidities.15 In the EVEREST II Trial, percutaneous MV repair was compared with surgery and showed superior safety but less reduction in MR at 1-year overall (MR moderate to severe at 1-year follow-up: 21% vs 3% percutaneous vs surgery group, respectively).16–19 Recently, the United States Food and Drug Administration has approved the premarket approval application for the MitraClip® Clip Delivery System (MitraClip® CDS) for percutaneous reduction of significant (≥3+) symptomatic degenerative MR in patients for whom a heart team has determined the surgical risk to be prohibitive.20

Back to Top | Article Outline

Etiology of Mitral Valve Regurgitation

The MV complex consists of the leaflets, mitral annulus, the subvalvular apparatus (chordae tendineae and papillary muscles), and their connections to the LV. The mitral annulus is a complex saddle-shaped structure with peaks anteriorly and posteriorly, and nadirs medially and laterally. Valvular competency depends on the correct interaction among the different components of the MV apparatus, and for this reason, MR can result from the impairment of any single element of the system.21,22 Thus, MR can be distinguished as primary or secondary. Primary MR is related to structural alterations of any of the MV apparatus components. Primary MR is most commonly caused by rheumatic and degenerative diseases. Due to the progressive reduction in the prevalence of rheumatic disease in developed countries, degenerative MR is now the most common type of primary MR. Secondary MR results from functional disturbance of the MV apparatus due to the changes in the LV geometry such as dilated cardiomyopathy or acute myocardial infarction.

The recommendations for surgery in patients with symptoms of either primary or secondary chronic MR are presented in Table 1.2

Table 1
Table 1
Image Tools
Back to Top | Article Outline

MitraClip® System

The MitraClip® System is a percutaneous approach that has been introduced into clinical practice.23,24 This approach results in permanent apposition of the 2 MV leaflets by a clip that is placed between the anterior middle scallop (A2) and posterior middle segment (P2) of the anterior mitral leaflet (AML) and the posterior mitral leaflet (PML), respectively.12,25,26 The iatrogenic coaptation of the AML and PML leads to the creation of a double-orifice MV that mimics the surgical edge-to-edge procedure introduced by Alfieri27 in 2001.

The clip is deployed between the 2 MV leaflets by a transeptal approach. The system comprises: (1) a 24-F steerable guide catheter that is advanced to the left atrium (LA) by a previously performed transeptal puncture; (2) a CDS, with the clip located on its distal tip, introduced through the catheter to the MV orifice; (3) the MitraClip® device comprises a mechanical clip with 2 arms that can be opened or closed by a handle mechanism positioned at the proximal end of the delivery catheter.12,28,29

A “grasping” system is located on the internal side of the clip’s 2 arms that stabilizes the clip after its deployment and attaches it to the LV side of the AML and PML. The system is steered by the 2-knob coaxial mechanism that allows medial–lateral and anteroposterior movements of the delivery system’s distal tip. The external handle navigates the CDS through the cardiac chambers: opening, orienting, deploying, and eventually detaching the clip from the MV14 (Fig. 1).

Figure 1
Figure 1
Image Tools
Back to Top | Article Outline

Echocardiography in MitraClip® Candidates

Echocardiography is an important diagnostic tool for the evaluation of MV in patients undergoing percutaneous MV repair. Transesophageal echocardiography (TEE) evaluation of MV anatomy and LV function provide useful information regarding the feasibility of percutaneous MV repair (Table 2).30–36 In EVEREST I, key anatomic inclusion criteria included a regurgitant jet origin associated with the A2 to P2 segments of the MV12. For patients with functional MR, to have sufficient tissue to grasp with the clip arms, a coaptation length of at least 2 mm and a coaptation depth <11 mm are needed. The coaptation length is measured as the vertical length of leaflet tissue in contact during midsystole in the atrial to ventricular direction (Fig. 2A). Coaptation depth is defined as the shortest distance between the coaptation and the annular planes (Fig. 2B). The 2 TEE views for measurement of coaptation length and coaptation depth are the 4-chamber long-axis (LAX) view and the midesophageal (ME) LAX view.

Table 2
Table 2
Image Tools
Figure 2
Figure 2
Image Tools

For patients with a flail mitral leaflet, a flail gap <10 mm and a flail width <15 mm are also important anatomic features to allow proper grasping. Flail gap is defined as the greatest distance between the ventricular side of the flail segment to the atrial side of the opposing leaflet edge in either the 4-chamber LAX or ME LAX view (Fig. 2C). Flail width is defined as the width of the flail leaflet segment as measured along the coaptation line in the transgastric short-axis (SAX) view (Fig. 2D).

Real-time 3-dimensional (D) TEE may have increasing value, compared with that of 2D TEE, providing higher quality visualization of MV anatomy.37 En face visualization of the MV segments, both from the atrial and ventricular orientations, allows detailed detection of MV pathology38–41 with particular emphasis on tissue defects such as clefts that can impact the feasibility of a Mitraclip® and are more difficult to recognize by 2D TEE.

Back to Top | Article Outline

MitraClip® Procedure and Intraprocedural Echocardiography

The MitraClip® procedure is usually performed in anesthestized patients to avoid movement and to facilitate TEE evaluation. Invasive hemodynamic monitoring is generally implemented to continuously evaluate hemodynamic status of these high-risk patients.

The delivery catheter is inserted through a femoral vein, using TEE and fluoroscopic guidance in the cardiac catheterization laboratory, as previously described.12,42 Intraprocedural TEE,14,42–44 preferably using real-time 3D to better evaluate the spatial interaction between the cardiac structures and the intracardiac devices by visualizing them from both atrial and ventricular perspectives, is instituted.45–47 Fluoroscopy is used primarily for assessing the opening angle of the clip arms, whereas it has a limited role in the steering and positioning of the clip. The latter are appropriately accomplished by TEE, so it is used as the primary imaging modality to guide the procedure and is reported to be essential to its success.48 TEE guidance is needed to implement the phases of the percutaneous procedure as follows: transseptal catheterization, axial alignment of the clip delivery system, leaflet grasping, assessment of leaflet capture, clip deployment, post–clip deployment assessment, and withdrawal of the clip delivery system.

These are the steps for the procedure:

Back to Top | Article Outline
Step 1: Transseptal Puncture

Proper localization of the septal puncture site is important.49 The septal puncture must be performed in the posterior–superior portion of the interatrial septum to assure sufficient movement of the CDS into the LA and to provide optimal orientation of the clip with respect to the MV annular plane.

The TEE views that help guide the transseptal catheterization follow (Fig. 3):

Figure 3
Figure 3
Image Tools

The ME SAX view (approximately 40°) shows the anterior–posterior portion of the interatrial septum and the junction of septum primum with septum secundum at the edge of the fossa ovalis. This view allows visualization of the aorta in a transverse orientation and is used for anterior–posterior (anterior towards the aorta) adjustments of the needle position and direction during transeptal puncture.

The ME bicaval view (approximately 90°) shows the superior–inferior portion of the interatrial septum. This view is used for inferior–superior orientation with tenting of the fossa (superior toward superior vena cava).

The ME 4-chamber view (approximately at 0°–30°) allows for the evaluation of the distance from the point where the septum is punctured to the mitral annular plane (Fig. 3).

Avoidance of accidental puncture of the aortic valve and/or the aortic root during this phase of the procedure is important.29,43 The puncture site is determined by creating tenting on the interatrial septum. The ideal site of transseptal puncture is localized across the fossa ovalis in a posterior trajectory toward the line of mitral leaflet coaptation at least 30 to 35 mm away from the valve, to allow the guide and device catheter a perpendicular approach to the MV and facilitate optimal movement and orientation through the distal portion of the guide catheter in the LA.

The linear distance between the tenting point on the septum and the MV annular plane can be easily measured at the beginning of the procedure using 2D TEE views.

Back to Top | Article Outline
Step 2: Axial Alignment of the Clip Delivery System

Once the transseptal puncture has been performed, the 24-F guide catheter is advanced into the LA and oriented perpendicular to the coaptation line and near the A2–P2 segments. The acquisition of the following views allows for navigation of the tip of the device across the MV, guidance of the clip orthogonal axial alignment with the plane of the MV annulus, and navigation parallel to the direction of anterograde mitral flow:

ME commissural view (approximately 60°) is used to display the medial–lateral plane of the MV.

ME LAX view (approximately 120°) identifies the anteroposterior plane of the MV.

Three-dimensional TEE is particularly helpful in guiding the medial, lateral, anterior, and posterior adjustments of the clip delivery system (Fig. 4). Once the clip’s arms are opened, 3D TEE is also useful for the evaluation of the adequacy of the clip arms alignment with respect to the MV’s line of coaptation. The atrial perspective of the MV and the accurate leaflets visualization provided by 3D TEE images suggest the proper rotation of the clip to achieve a perpendicular position to the mitral leaflets at the origin of the regurgitation jet (Video 1, see Supplemental Digital Content 1, http://links.lww.com/AA/A891).

Figure 4
Figure 4
Image Tools
Back to Top | Article Outline
Step 3: Leaflet Grasping

The clip’s arms are now opened into the grasping position and oriented perpendicularly with respect to the MV leaflets’ coaptation line (Fig. 5). The device is now advanced into the LV through the MV and the arms of the clip grasp the mitral leaflets (Video 2, see Supplemental Digital Content 2, http://links.lww.com/AA/A892). The ME LAX view is generally used to guide this crucial phase of the procedure (Fig. 6). Three-dimensional TEE can adequately evaluate the positioning and orientation of the clip by providing images of the MV and the clip from both the atrial and ventricular site (Fig. 7). The residual MR jet estimate, the diastolic transmitral peak velocity measurement, and mean mitral gradient should be performed to confirm adequacy of clip implantation and ensure the lack of MV stenosis.36 When the grasp is not optimal or a partial clip detachment occurs, the clip can be opened, removed from the initial position (Video 3, see Supplemental Digital Content 3, http://links.lww.com/AA/A893), and redeployed (Video 4, see Supplemental Digital Content 4, http://links.lww.com/AA/A894). The location of the MitraClip® device can be changed to achieve the best apposition of the mitral leaflets to optimize reduction of MR. The echocardiographer’s acquisition of real-time information regarding the position of the clip and the severity of the MR can successfully guide the procedure.

Figure 5
Figure 5
Image Tools
Figure 6
Figure 6
Image Tools
Figure 7
Figure 7
Image Tools
Back to Top | Article Outline
Step 4: Assessment of Leaflet Capture and Clip Deployment

The adequacy of leaflet capture and clip deployment is confirmed by TEE assessment of combined grasped tissue stability and MR reduction, in the presence of a double-orifice MV. Color flow Doppler is used to analyze the origin, vena contracta, size, and direction of the residual jets.50 Although both 2D and 3D TEE views can adequately visualize the double-orifice image (Figs. 8–9), 3D TEE, in live 3D or wide zoom modality, better evaluates the efficacy of the MV repair by visualizing the MV orifices either from the atrial or the ventricular site (Video 5, see Supplemental Digital Content 5, http://links.lww.com/AA/A895).27 TEE provides information about the need for a second clip when the first clip is not positioned optimally and when severe residual MR is detected. Once efficacy of the percutaneous MV repair is demonstrated on TEE, the clip is closed and released.

Figure 8
Figure 8
Image Tools
Figure 9
Figure 9
Image Tools
Back to Top | Article Outline
Step 5: Post–Clip Deployment Assessment and Withdrawal of the Clip Delivery System

A comprehensive TEE examination is required after clip deployment to confirm reduction of MR, adequate position of the clip, evaluation of gradients (in particular after 2 clip implantation), and to detect any new findings (Video 6, see Supplemental Digital Content 6, http://links.lww.com/AA/A896).

The MV area should be measured by planimetry in the basal transgastric SAX. After percutaneous MV repair, each of the 2 orifices is planimetred at the level of the clip and summed (Fig. 10). Neither pulsed or continuous wave Doppler mitral velocity have been validated for use with a double-orifice valve after percutaneous MV repair.

Figure 10
Figure 10
Image Tools

Once the positioning of the clip is deemed adequate, the system is removed. TEE guidance is used to carefully withdraw the clip delivery system into the right atrium and out of the patient.

Back to Top | Article Outline

Detection of Complications

TEE provides a method for early detection of potential complications of clip placement including perforation of the atrial wall that may cause pericardial effusion/tamponade or partial detachment of the clip defined as detachment of a single leaflet from the clip (Video 7, see Supplemental Digital Content 7, http://links.lww.com/AA/A897). In the EVEREST I study, partial clip detachment occurred in 10 patients (9%) without the need for urgent intervention.12 TEE is used to check for iatrogenic atrial septal defect after the procedure51 (Fig. 11).

Figure 11
Figure 11
Image Tools
Back to Top | Article Outline


Real-time 3D TEE significantly contributes to performance adequacy and to evaluating the individual steps of Mitraclip® device implantation.52

In addition, 3D TEE allows for complete visualization of the MV apparatus by providing images from both the atrial and the ventricular perspective.37–41 The 3D views of the intracardiac devices’ position with respect to the heart structures help the operator to guide the clip delivery system through the heart and to deliver the clip correctly. With the recent introduction of real-time 3D TEE, we should consider the real advantage of this modality in the operative management of percutaneous MV repair. Clinical experience and evidence from the literature suggest that real-time 3D TEE improves visualization of the MV through the display of the whole valve from a single image. This means that consulting off-line sources or relying on one’s memory is not needed and that even physicians who are not expert in ultrasound imaging, such as surgeons and interventional cardiologists, can easily interpret the echo view. As a consequence, 3D TEE ameliorates communication between operator and echocardiographer, possibly reducing the duration of the procedure. However, 3D TEE has limitations that need to be considered. This modality is strongly dependent on spatial resolution and frame rate; a good 3D image is always the result of good 2D image quality.

Back to Top | Article Outline


TEE not only plays a major role in the MitraClip® procedure, but also it is an indispensable tool. Real-time 3D TEE can improve echocardiographic performance during the MitraClip® procedure. Cardiac anesthesiologists, who combine expertise in intraoperative echocardiography with skill in hemodynamic management, may play a pivotal role in the new MitraClip® procedure.

Back to Top | Article Outline


Name: Fabio Guarracino, MD.

Contribution: This author prepared the manuscript.

Attestation: Fabio Guarracino approved the final manuscript.

Name: Rubia Baldassarri, MD.

Contribution: This author helped preparing the manuscript.

Attestation: Rubia Baldassarri approved the final manuscript.

Name: Baldassare Ferro, MD.

Contribution: This author helped preparing the manuscript.

Attestation: Baldassare Ferro approved the final manuscript.

Name: Cristina Giannini, MD, PhD.

Contribution: This author helped preparing the manuscript and collecting the figures.

Attestation: Cristina Giannini approved the final manuscript.

Name: Pietro Bertini, MD.

Contribution: This author helped preparing and reviewing the manuscript and editing the movies.

Attestation: Pietro Bertini approved the final manuscript.

Name: Anna Sonia Petronio, MD.

Contribution: This author helped preparing and reviewed the manuscript.

Attestation: Anna Sonia Petronio approved the final manuscript.

Name: Vitantonio Di Bello, MD.

Contribution: This author helped preparing and reviewed the manuscript.

Attestation: Vitantonio Di Bello approved the final manuscript.

Name: Giovanni Landoni, MD.

Contribution: This author reviewed the manuscript.

Attestation: Giovanni Landoni approved the final manuscript.

Name: Ottavio Alfieri, MD.

Contribution: This author reviewed the manuscript and the figures.

Attestation: Ottavio Alfieri approved the final manuscript.

This manuscript was handled by: Martin J. London, MD.

Back to Top | Article Outline


1. Iung B, Baron G, Butchart EG, Delahaye F, Gohlke-Bärwolf C, Levang OW, Tornos P, Vanoverschelde JL, Vermeer F, Boersma E, Ravaud P, Vahanian A. A prospective survey of patients with valvular heart disease in Europe: The Euro Heart Survey on Valvular Heart Disease. Eur Heart J. 2003;24:1231–43

2. Vahanian A, Alfieri O, Andreotti F, Antunes MJ, Baro` n-Esquivias G, Baumgartner H, Borger MA, Carrel TP, DeBonis M, Evangelista A, Falk V, Iung B,, Lancellotti P,, Pierard L, Price S,, Scha¨fers HJ, Schuler G,, Stepinska J,, Swedberg K,, Takkenberg J, Von Oppell UO, Windecker S,, Zamorano JL, Zembala M. Guidelines on the management of valvular heart disease (version 2012) The Joint Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J. 2012;33:2451–96

3. Bonow RO, Carabello BA, Kanu C, de Leon AC Jr, Faxon DP, Freed MD, Gaasch WH, Lytle BW, Nishimura RA, O’Gara PT, O’Rourke RA, Otto CM, Shah PM, Shanewise JS, Smith SC Jr, Jacobs AK, Adams CD, Anderson JL, Antman EM, Faxon DP, Fuster V, Halperin JL, Hiratzka LF, Hunt SA, Lytle BW, Nishimura R, Page RL, Riegel B. American College of Cardiology/American Heart Association Task Force on Practice Guidelines, Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Interventions, Society of Thoracic Surgeons. ACC/AHA 2006 Guidelines for the Management of Patients with Valvular Heart Disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): developed in collaboration with the Society of Cardiovascular Anesthesiologists: endorsed by the Society for Cardiovascular Angiography and Interventions and the Society of Thoracic Surgeons Circulation. 2006;114:e84–231

4. Rosenhek R, Rader F, Klaar U, Gabriel H, Krejc M, Kalbeck D, Schemper M, Maurer G, Baumgartner H. Outcome of watchful waiting in asymptomatic severe mitral regurgitation. Circulation. 2006;113:2238–44

5. Kang DH, Kim JH, Rim JH, Kim MJ, Yun SC, Song JM, Song H, Choi KJ, Song JK, Lee JW. Comparison of early surgery versus conventional treatment in asymptomatic severe mitral regurgitation Circulation. 2009;119:797–804

6. Sparano DM, Ward RP. Management of asymptomatic, severe mitral regurgitation. Curr Treat Options Cardiovasc Med. 2012;14:575–83

7. Enriquez-Sarano M, Schaff HV, Orszulak TA, Tajik AJ, Bailey KR, Frye RL. Valve repair improves the outcome of surgery for mitral regurgitation: a multivariate analysis Circulation. 1995;91:1022––8

8. Braunberger E, Deloche A, Berrebi A, Abdallah F, Celestin JA, Meimoun P, Chatellier G, Chauvaud S, Fabiani JN, Carpentier A. Very long-term results (more than 20 years) of valve repair with carpentier’s techniques in nonrheumatic mitral valve insufficiency. Circulation. 2001;104:I8–11

9. Shuhaiber J, Anderson RJ. Meta-analysis of clinical outcomes following surgical mitral valve repair or replacement. Eur J Cardiothorac Surg. 2007;31:267–75

10. Block PC. Percutaneous transcatheter repair for mitral regurgitation. J Interv Cardiol. 2006;19:547–51

11. Franzen O, Baldus S, Rudolph V, Meyer S, Knap M, Koschyk D, Treede H, Barmeyer A, Schofer J, Costard-Jäckle A, Schlüter M, Reichenspurner H, Meinertz T. Acute outcomes of MitraClip therapy for mitral regurgitation in high-surgical-risk patients: emphasis on adverse valve morphology and severe left ventricular dysfunction. Eur Heart J. 2010;31:1373–81

12. Feldman T, Kar S, Rinaldi M, Fail P, Hermiller J, Smalling R, Whitlow PL, Gray W, Low R, Herrmann HC, Lim S, Foster E, Glower DEVEREST Investigators. . Percutaneous mitral repair with the MitraClip system: safety and midterm durability in the initial EVEREST (Endovascular Valve Edge-to-Edge REpair Study) cohort. J Am Coll Cardiol. 2009;54:686–94

13. Rogers JH, Franzen O. Percutaneous edge-to-edge MitraClip therapy in the management of mitral regurgitation. Eur Heart J. 2011;32:2350–7

14. Feldman T, Wasserman HS, Herrmann HC, Gray W, Block PC, Whitlow P, St Goar F, Rodriguez L, Silvestry F, Schwartz A, Sanborn TA, Condado JA, Foster E. Percutaneous mitral valve repair using the edge-to-edge technique: six-month results of the EVEREST Phase I Clinical Trial. J Am Coll Cardiol. 2005;46:2134–40

15. Maisano F, Godino C, Giacomini A, Denti P, Buzzatti N, Arendar I, Colombo A, Alfieri O, La Canna G. Patient selection for MitraClip therapy impaired left ventricular systolic function. Minerva Cardioangiol. 2011;59:455–71

16. Feldman T, Foster E, Glower DD, Kar S, Rinaldi MJ, Fail PS, Smalling RW, Siegel R, Rose GA, Engeron E, Loghin C, Trento A, Skipper ER, Fudge T, Letsou GV, Massaro JM, Mauri LEVEREST II Investigators. . Percutaneous repair or surgery for mitral regurgitation N Engl J Med. 2011;364:1395–406

17. Mauri L, Garg P, Massaro JM, Foster E, Glower D, Mehoudar P, Powell F, Komtebedde J, McDermott E, Feldman T. The EVEREST II Trial: design and rationale for a randomized study of the evalve mitraclip system compared with mitral valve surgery for mitral regurgitation. Am Heart J. 2010;160:23–9

18. Herrmann HC, Rohatgi S, Wasserman HS, Block P, Gray W, Hamilton A, Zunamon A, Homma S, Di Tullio MR, Kraybill K, Merlino J, Martin R, Rodriguez L, Stewart WJ, Whitlow P, Wiegers SE, Silvestry FE, Foster E, Feldman T. Mitral valve hemodynamic effects of percutaneous edge-to-edge repair with the MitraClip device for mitral regurgitation. Catheter Cardiovasc Interv. 2006;68:821–8

19. Whitlow PL, Feldman T, Pedersen WR, Lim DS, Kipperman R, Smalling R, Bajwa T, Herrmann HC, Lasala J, Maddux JT, Tuzcu M, Kapadia S, Trento A, Siegel RJ, Foster E, Glower D, Mauri L, Kar SEVEREST II Investigators. . Acute and 12-month results with catheter-based mitral valve leaflet repair: the EVEREST II (Endovascular Valve Edge-to-Edge Repair) High Risk Study. J Am Coll Cardiol. 2012;59:130–9

21. Mathew JP, Ayoub C Clinical Manual and Review of Transesophageal Echocardiography.. 2005 New York, NY McGraw-Hill, Medical Pub. Division

22. Debonnaire P, Palmen M, Marsan NA, Delgado V. Contemporary imaging of normal mitral valve anatomy and function. Curr Opin Cardiol. 2012;27:455–64

23. Fann JI, St Goar FG, Komtebedde J, Oz MC, Block PC, Foster E, Butany J, Feldman T, Burdon TA. Beating heart catheter-based edge-to-edge mitral valve procedure in a porcine model: efficacy and healing response. Circulation. 2004;110:988–93

24. St Goar FG, Fann JI, Komtebedde J, Foster E, Oz MC, Fogarty TJ, Feldman T, Block PC. Endovascular edge-to-edge mitral valve repair: short-term results in a porcine model Circulation. 2003;108:1990–3

25. Altiok E, Paetsch I, Jahnke C, Brehmer K, Reith S, Becker M, Marx N, Hoffmann R. Percutaneous edge-to-edge mitral valve repair: assessment of immediate post-procedural treatment effect using color 3-dimensional transesophageal echocardiography and cardiac magnetic resonance imaging J Am Coll Cardiol. 2011;58:e21

26. Ussia GP, Cammalleri V, Scandura S, Immè S, Pistritto AM, Ministeri M, Chiarandà M, Caggegi A, Barbanti M, Aruta P, Tamburino C. Update on percutaneous mitral valve therapy: clinical results and real life experience. Minerva Cardioangiol. 2012;60:57–70

27. Alfieri O, Maisano F, De Bonis M, Stefano PL, Torracca L, Oppizzi M, La Canna G. The double-orifice technique in mitral valve repair: a simple solution for complex problems. J Thorac Cardiovasc Surg. 2001;122:674–81

28. Tamburino C, Ussia GP, Maisano F, Capodanno D, La Canna G, Scandura S, Colombo A, Giacomini A, Michev I, Mangiafico S, Cammalleri V, Barbanti M, Alfieri O. Percutaneous mitral valve repair with the MitraClip system: acute results from a real world setting. Eur Heart J. 2010;31:1382–9

29. Alegria-Barrero E, Chan PH, Paulo M, Duncan A, Price S, Moat N, Di Mario C, Serra A, García E, Franzen O. Edge-to-edge percutaneous repair of severe mitral regurgitation–state-of-the-art for Mitraclip® implantation. Circ J. 2012;76:801–8

30. Topilsky Y, Grigioni F, Enriquez-Sarano M. Quantitation of mitral regurgitation. Semin Thorac Cardiovasc Surg. 2011;23:106–14

31. Cavalcante JL, Rodriguez LL, Kapadia S, Tuzcu EM, Stewart WJ. Role of echocardiography in percutaneous mitral valve interventions. JACC Cardiovasc Imaging. 2012;5:733–46

32. Armstrong EJ, Rogers JH, Swan CH, Upadhyaya D, Viloria E, McCulloch C, Slater J, Qureshi M, Williams J, Whisenant B, Feldman T, Foster E. Echocardiographic predictors of single versus dual MitraClip device implantation and long-term reduction of mitral regurgitation after percutaneous repair Catheter Cardiovasc Interv. 2012;82:673–9

33. Kneeshaw JD. Transoesophageal echocardiography (TOE) in the operating room. Br J Anesth. 2006;97:77–84

34. Cheitlin MD, Armstrong WF, Aurigemma GP, Beller GA, Bierman FZ, Davis JL, Douglas PS, Faxon DP, Gillam LD, Kimball TR, Kussmaul WG, Pearlman AS, Philbrick JT, Rakowski H, Thys DM, Antman EM, Smith SC Jr, Alpert JS, Gregoratos G, Anderson JL, Hiratzka LF, Hunt SA, Fuster V, Jacobs AK, Gibbons RJ, Russell ROAmerican College of Cardiology; American Heart Association; American Society of Echocardiography. . ACC/AHA/ASE 2003 guideline update for the clinical application of echocardiography: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASE Committee to Update the 1997 Guidelines for the Clinical Application of Echocardiography). Circulation. 2003;108:1146–62

35. Iglesias I. Intraoperative TEE assessment during mitral valve repair for degenerative and ischemic mitral valve regurgitation. Semin Cardiothorac Vasc Anesth. 2007;11:301–5

36. Silvestry FE, Rodriguez LL, Herrmann HC, Rohatgi S, Weiss SJ, Stewart WJ, Homma S, Goyal N, Pulerwitz T, Zunamon A, Hamilton A, Merlino J, Martin R, Krabill K, Block PC, Whitlow P, Tuzcu EM, Kapadia S, Gray WA, Reisman M, Wasserman H, Schwartz A, Foster E, Feldman T, Wiegers SE. Echocardiographic guidance and assessment of percutaneous repair for mitral regurgitation with the Evalve MitraClip: lessons learned from EVEREST I. J Am Soc Echocardiogr. 2007;20:1131–40

37. Little SH. Three-dimensional echocardiography to quantify mitral valve regurgitation. Curr Opin Cardiol. 2012;27:477–84

38. Shernan SK. Perioperative transesophageal echocardiographic evaluation of the native mitral valve Crit Care Med. 2007;35(8 suppl):S372–83

39. Franke A, Hung J, Perez de Isla L, Kamp O, Kasprzak JD, Lancellotti P, Marwick TH, McCulloch ML, Monaghan MJ, Nihoyannopoulos P, Pandian NG, Pellikka PA, Pepi M, Roberson DA, Shernan SA, Shirali GS, Sugeng L, Ten Cate FJ, Vannan MA, Zamorano JL, Zoghbi WA. EAE/ASE recommendations for image acquisition and display using three-dimensional echocardiography Eur Heart J Cardiovasc Imag. 2012;13:1–46

40. Ciobanu A, Bennett S, Azam M, Clark A, Vinereanu D. Incremental value of three-dimensional transoesophageal echocardiography for guiding double percutaneous MitraClip ® implantation in a ‘no option’ patient. Eur J Echocardiogr. 2011;12:E11

41. Salcedo EE, Quaife RA, Seres T, Carroll JD. A framework for systematic characterization of the mitral valve by real-time three-dimensional transesophageal echocardiography. J Am Soc Echocardiogr. 2009;22:1087–99

42. Zamorano JL, Badano LP, Bruce C, Chan KL, Gonçalves A, Hahn RT, Keane MG, La Canna G, Monaghan MJ, Nihoyannopoulos P, Silvestry FE, Vanoverschelde JL, Gillam LD. EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. J Am Soc Echocardiogr. 2011;24:937–65

43. Naqvi TZ. Echocardiography in percutaneous valve therapy. JACC Cardiovasc Imaging. 2009;2:1226–37

44. Slipczuk L, Siegel RJ, Jilaihawi H, Hussaini A, Kar S. Optimizing procedural outcomes in percutaneous mitral valve therapy using transesophageal imaging: a stepwise analysis. Expert Rev Cardiovasc Ther. 2012;10:901–16

45. Balzer J, Kühl H, Rassaf T, Hoffmann R, Schauerte P, Kelm M, Franke A. Real-time transesophageal three-dimensional echocardiography for guidance of percutaneous cardiac interventions: first experience. Clin Res Cardiol. 2008;97:565–74

46. Altiok E, Becker M, Hamada S, Reith S, Marx N, Hoffmann R. Optimized guidance of percutaneous edge-to edge repair of the mitral valve using real-time 3-D transesophageal echocardiography. Clin Res Cardiol. 2011;100:675–81

47. Altiok E, Hamada S, Brehmer K, Kuhr K, Reith S, Becker M, Schröder J, Almalla M, Lehmacher W, Marx N, Hoffmann R. Analysis of procedural effects of percutaneous edge-to-edge mitral valve repair by 2D and 3D echocardiography. Circ Cardiovasc Imaging. 2012;5:748–55

48. Earley MJ. How to perform a transseptal puncture. Heart. 2009;95:85–92

49. Singh K, Raphael J, Colquhoun D.. A rare case of mitral stenosis after MitraClip placement: transesophageal echocardiography findings and examination Anesth Analg. 2013;117:777–9

50. Lancellotti P, Tribouilloy C, Hagendorff A, Popescu BA, Edvardsen T, Pierard LA, Badano L, Zamorano JLScientific Document Committee of the European Association of Cardiovascular Imaging. . Recommendations for the echocardiographic assessment of native valvular regurgitation: an executive summary from the European Association of Cardiovascular Imaging. Eur Heart J Cardiovasc Imaging. 2013;14:611–44

51. Saitoh T, Izumo M, Furugen A, Tanaka J, Miyata-Fukuoka Y, Gurudevan SV, Tolstrup K, Siegel RJ, Kar S, Shiota T. Echocardiographic evaluation of iatrogenic atrial septal defect after catheter-based mitral valve clip insertion. Am J Cardiol. 2012;109:1787–91

52. Swaans MJ, Van den Branden BJ, Van der Heyden JA, Post MC, Rensing BJ, Eefting FD, Plokker HW, Jaarsma W. Three-dimensional transoesophageal echocardiography in a patient undergoing percutaneous mitral valve repair using the edge-to-edge clip technique. Eur J Echocardiogr. 2009;10:982–3

Supplemental Digital Content

Back to Top | Article Outline

© 2014 International Anesthesia Research Society


Become a Society Member

Article Level Metrics