The Accuracy, Precision and Reliability of Measuring Ventilatory Rate and Detecting Ventilatory Pause by Rainbow Acoustic Monitoring and Capnometry

Ramsay, Michael A. E. MD*; Usman, Mohammad PhD; Lagow, Elaine RN; Mendoza, Minerva RN§; Untalan, Emylene RN§; De Vol, Edward PhD

doi: 10.1213/ANE.0b013e318290c798
Technology, Computing, and Simulation: Research Report

BACKGROUND: Current methods for monitoring ventilatory rate have limitations including poor accuracy and precision and low patient tolerance. In this study, we evaluated a new acoustic ventilatory rate monitoring technology for accuracy, precision, reliability, and the ability to detect pauses in ventilation, relative to capnometry and a reference method in postsurgical patients.

METHODS: Adult patients presenting to the postanesthesia care unit were connected to a Pulse CO-Oximeter with acoustic monitoring technology (Rad-87, version 7804, Masimo, Irvine, CA) through an adhesive bioacoustic sensor (RAS-125, rev C) applied to the neck. Each subject also wore a nasal cannula connected to a bedside capnometer (Capnostream20, version 4.5, Oridion, Needham, MA). The acoustic monitor and capnometer were connected to a computer for continuous acoustic and expiratory carbon dioxide waveform recordings. Recordings were retrospectively analyzed by a trained technician in a setting that allowed for the simultaneous viewing of both waveforms while listening to the breathing sounds from the acoustic signal to determine inspiration and expiration reference markers within the ventilatory cycle without using the acoustic monitor- or capnometer-calculated ventilatory rate. This allowed the automatic calculation of a reference ventilatory rate for each device through a software program (TagEditor, Masimo). Accuracy (relative to the respective reference) and precision of each device were estimated and compared with each other. Sensitivity for detection of pauses in ventilation, defined as no inspiration or expiration activity in the reference ventilatory cycle for ≥30 seconds, was also determined. The devices were also evaluated for their reliability, i.e., the percentage of the time when each displayed a value and did not drop a measurement.

RESULTS: Thirty-three adults (73% female) with age of 45 ± 14 years and weight 117 ± 42 kg were enrolled. A total of 3712 minutes of monitoring time (average 112 minutes per subject) were analyzed across the 2 devices, reference ventilatory rates ranged from 1.9 to 49.1 bpm. Acoustic monitoring showed significantly greater accuracy (P = 0.0056) and precision (P- = 0.0024) for respiratory rate as compared with capnometry. On average, both devices displayed data over 97% of the monitored time. The (0.95, 0.95) lower tolerance limits for the acoustic monitor and capnometer were 94% and 84%, respectively. Acoustic monitoring was marginally more sensitive (P = 0.0461) to pauses in ventilation (81% vs 62%) in 21 apneic events.

CONCLUSIONS: In this study of a population of postsurgical patients, the acoustic monitor and capnometer both reliably monitored ventilatory rate. The acoustic monitor was statistically more accurate and more precise than the capnometer, but differences in performance were modest. It is not known whether the observed differences are clinically significant. The acoustic monitor was more sensitive to detecting pauses in ventilation. Acoustic monitoring may provide an effective and convenient means of monitoring ventilatory rate in postsurgical patients.

Published ahead of print April 30, 2013 Supplemental Digital Content is available in the text.

From the *Department Of Anesthesiology and Pain Management, Baylor Research Institute, Baylor University Medical Center, Dallas, Texas; Masimo Corporation, Irvine, California; Baylor Research Institute and §Post Anesthesia Care Unit, Baylor University Medical Center; and Department of Quantitative Sciences, Baylor Health Care System, Dallas, Texas.

Accepted for publication February 12, 2013.

Published ahead of print April 30, 2013

Funding: Masimo provided equipment and sensors for the study and funded the cost of research personnel who collected study data.

Conflict of Interest: See Disclosures at the end of the article.

Reprints will not be available from the authors.

Address correspondence to Michael A. E. Ramsay, MD, Department Of Anesthesiology and Pain Management, Baylor University Medical Center, 3500 Gaston Ave., 2 Roberts, Dallas, Texas 75246. Address e-mail to docram@baylorhealth.edu.

© 2013 International Anesthesia Research Society